N5227B.pdf - 第14页

Find us at www. keysi ght.com Page 14 Table 5 . System Dynamic Range at Test Port (dB), Options 205, 405 Description Specificatio n 3 Typical (d B) Ports 2 1, 3 Ports 2 2, 4 Ports 2 1, 3 Ports 2 2, 4 10 MHz to 50 M Hz 1 …

100%1 / 90
Find us at www.keysight.com Page 13
Table 4. System Dynamic Range at Test Port (dB), Options 219 or 419
Description Specification
2
Typical (dB)
Ports
1
1, 3
Ports
1
2, 4
Ports
1
1, 3
Ports
1
2, 4
10 MHz to 50 MHz
86
86
95
94
50 MHz to 100 MHz
109
108
116
116
100 MHz to 500 MHz
116
116
130
131
500 MHz to 1 GHz
129
129
136
137
1 GHz to 2 GHz
130
133
138
140
2 GHz to 3.2 GHz
126
131
134
138
3.2 GHz to 10 GHz
130
131
137
138
10 GHz to 13.5 GHz
130
129
138
137
13.5 GHz to 16 GHz
130
129
137
137
16 GHz to 19 GHz
128
127
136
136
19 GHz to 20 GHz
128
127
136
136
20 GHz to 24 GHz
127
127
135
135
24 GHz to 26.5 GHz
127
127
135
134
26.5 GHz to 30 GHz
118
118
127
126
30 GHz to 32 GHz
116
116
125
125
32 GHz to 35 GHz
116
115
125
124
35 GHz to 40 GHz
106
106
116
116
40 GHz to 43.5 GHz
111
109
121
118
43.5 GHz to 50 GHz
111
110
120
120
50 GHz to 60 GHz
109
107
118
117
60 GHz to 64 GHz
110
108
119
118
64 GHz to 67 GHz
110
109
119
119
67 GHz to 70 GHz
--
--
116
115
1
Either port can be used as the source port. Any other port can be used as the receiver port.
2
Specification applies to new instruments. Test limit during recertification is defined by Maximum Leveled Power - Noise
Floor.
Find us at www.keysight.com Page 14
Table 5. System Dynamic Range at Test Port (dB), Options 205, 405
Description Specification
3
Typical (dB)
Ports
2
1, 3
Ports
2
2, 4
Ports
2
1, 3
Ports
2
2, 4
10 MHz to 50 MHz
1
81
76
89
87
50 MHz to 100 MHz
1
103
104
112
113
100 MHz to 500 MHz
110
112
125
126
500 MHz to 1 GHz
123
123
132
133
1 GHz to 2 GHz
130
129
137
139
2 GHz to 3.2 GHz
127
129
134
137
3.2 GHz to 10 GHz
128
128
136
137
10 GHz to 13.5 GHz
128
126
137
136
13.5 GHz to 16 GHz
128
127
137
136
16 GHz to 19 GHz
126
127
136
137
19 GHz to 20 GHz
128
128
137
137
20 GHz to 24 GHz
126
127
135
136
24 GHz to 26.5 GHz
127
127
135
136
26.5 GHz to 30 GHz
118
118
128
128
30 GHz to 32 GHz
116
117
126
126
32 GHz to 35 GHz
117
117
126
127
35 GHz to 40 GHz
107
107
119
119
40 GHz to 43.5 GHz
112
111
123
123
43.5 GHz to 50 GHz
112
113
123
124
50 GHz to 60 GHz
111
111
121
121
60 GHz to 64 GHz
111
112
122
122
64 GHz to 67 GHz
110
114
122
124
67 GHz to 70 GHz
--
--
119
119
1
With Option 205, 405 installed and LFE disabled, applied to frequencies <= 100 MHz. Above 100 MHz, performance is
the same for both LFE enabled or disabled.
2
Either port can be used as the source port. Any other port can be used as the receiver port.
3
Specification applies to new instruments. Test limit during recertification is defined by Maximum Leveled Power - Noise
Floor.
Find us at www.keysight.com Page 15
Table 6. System Dynamic Range at Test Port (dB), Options 220, 420
Description Specification
3
Typical (dB)
Ports
2
1, 3
Ports
2
2, 4
Ports
2
1, 3
Ports
2
2, 4
10 MHz to 50 MHz
1
79
79
88
87
50 MHz to 100 MHz
1
105
104
112
112
100 MHz to 500 MHz
112
112
126
127
500 MHz to 1 GHz
126
126
133
134
1 GHz to 2 GHz
127
130
135
137
2 GHz to 3.2 GHz
123
128
131
135
3.2 GHz to 10 GHz
128
129
135
136
10 GHz to 13.5 GHz
128
127
136
135
13.5 GHz to 16 GHz
128
127
135
135
16 GHz to 19 GHz
126
125
134
134
19 GHz to 20 GHz
126
125
134
134
20 GHz to 24 GHz
125
125
133
133
24 GHz to 26.5 GHz
125
125
133
132
26.5 GHz to 30 GHz
117
117
126
125
30 GHz to 32 GHz
115
115
124
124
32 GHz to 35 GHz
115
114
124
123
35 GHz to 40 GHz
105
105
115
115
40 GHz to 43.5 GHz
110
108
120
117
43.5 GHz to 50 GHz
110
109
119
119
50 GHz to 60 GHz
108
106
117
116
60 GHz to 64 GHz
109
107
118
117
64 GHz to 67 GHz
109
108
118
118
67 GHz to 70 GHz
--
--
115
114
1
With Option 220, 420 installed and LFE disabled, applied to frequencies <= 100 MHz. Above 100 MHz, performance is
the same for both LFE enabled or disabled.
2
Either port can be used as the source port. Any other port can be used as the receiver port.
3
Specification applies to new instruments. Test limit during recertification is defined by Maximum Leveled Power - Noise