N5227B.pdf - 第16页

Find us at www. keysi ght.com Page 16 Table 7 . System Dynamic Range at Test Port (dB ), All LFE Options (LFE Enabled) Description Specificatio n 1 Typical Ports 1, 3 Ports 2, 4 Ports 1, 3 Ports 2, 4 500 Hz to 900 Hz -- …

100%1 / 90
Find us at www.keysight.com Page 15
Table 6. System Dynamic Range at Test Port (dB), Options 220, 420
Description Specification
3
Typical (dB)
Ports
2
1, 3
Ports
2
2, 4
Ports
2
1, 3
Ports
2
2, 4
10 MHz to 50 MHz
1
79
79
88
87
50 MHz to 100 MHz
1
105
104
112
112
100 MHz to 500 MHz
112
112
126
127
500 MHz to 1 GHz
126
126
133
134
1 GHz to 2 GHz
127
130
135
137
2 GHz to 3.2 GHz
123
128
131
135
3.2 GHz to 10 GHz
128
129
135
136
10 GHz to 13.5 GHz
128
127
136
135
13.5 GHz to 16 GHz
128
127
135
135
16 GHz to 19 GHz
126
125
134
134
19 GHz to 20 GHz
126
125
134
134
20 GHz to 24 GHz
125
125
133
133
24 GHz to 26.5 GHz
125
125
133
132
26.5 GHz to 30 GHz
117
117
126
125
30 GHz to 32 GHz
115
115
124
124
32 GHz to 35 GHz
115
114
124
123
35 GHz to 40 GHz
105
105
115
115
40 GHz to 43.5 GHz
110
108
120
117
43.5 GHz to 50 GHz
110
109
119
119
50 GHz to 60 GHz
108
106
117
116
60 GHz to 64 GHz
109
107
118
117
64 GHz to 67 GHz
109
108
118
118
67 GHz to 70 GHz
--
--
115
114
1
With Option 220, 420 installed and LFE disabled, applied to frequencies <= 100 MHz. Above 100 MHz, performance is
the same for both LFE enabled or disabled.
2
Either port can be used as the source port. Any other port can be used as the receiver port.
3
Specification applies to new instruments. Test limit during recertification is defined by Maximum Leveled Power - Noise
Floor.
Find us at www.keysight.com Page 16
Table 7. System Dynamic Range at Test Port (dB), All LFE Options (LFE Enabled)
Description Specification
1
Typical
Ports 1, 3
Ports 2, 4
Ports 1, 3
Ports 2, 4
500 Hz to 900 Hz
--
--
105
105
900 Hz to 1 kHz
100
102
109
110
1 kHz to 10 kHz
103
105
110
111
10 kHz to 100 kHz
113
115
120
121
100 kHz to 1 MHz
120
121
124
125
1 MHz to 5 MHz
121
122
126
127
5 MHz to 10 MHz
112
114
118
119
10 MHz to 50 MHz
110
112
116
117
50 MHz to 100 MHz
110
112
116
117
1
Specification applies to new instruments. Test limit during recertification is defined by Maximum Leveled Power - Noise
Floor.
Find us at www.keysight.com Page 17
Extended Dynamic Range
Table 8. Extended Dynamic Range at Direct Receiver Access Input (dB) - Typical
Description Options 201, 401 Options 219, 419
Ports
1
1, 3 Ports
1
2, 4 Ports
1
1, 3 Ports
1
2, 4
10 MHz to 50 MHz
136
134
135
134
50 MHz to 100 MHz
145
144
144
144
100 MHz to 500 MHz
158
158
158
159
500 MHz to 1 GHz
164
165
164
165
1 GHz to 2 GHz
153
155
153
155
2 GHz to 3.2 GHz
150
154
149
153
3.2 GHz to 10 GHz
152
154
152
153
10 GHz to 13.5 GHz
153
152
152
151
13.5 GHz to 16 GHz
152
152
151
151
16 GHz to 19 GHz
153
152
150
150
19 GHz to 20 GHz
153
153
150
150
20 GHz to 24 GHz
152
152
149
149
24 GHz to 26.5 GHz
151
151
149
148
26.5 GHz to 30 GHz
145
144
141
140
30 GHz to 32 GHz
142
142
139
139
32 GHz to 35 GHz
142
142
139
138
35 GHz to 40 GHz
135
135
130
130
40 GHz to 43.5 GHz
138
136
134
131
43.5 GHz to 50 GHz
138
138
133
133
50 GHz to 60 GHz
136
136
130
129
60 GHz to 64 GHz
136
136
130
129
64 GHz to 67 GHz
138
138
130
130
67 GHz to 70 GHz
135
135
127
126
1