N5227B.pdf - 第53页

Find us at www. keysi ght.com Page 53 Table 52 . 0.1 dB Compre ssion, All P o rts, All LFE Opt ions (LFE E nabled) - Typical Description Test Port Po wer (dBm) 500 Hz to 900 Hz 13 900 Hz to 1 kHz 13 1 kHz to 10 kHz 13 10…

100%1 / 90
Find us at www.keysight.com Page 52
Table 50. Direct Receiver Access Input Noise Floor (dBm) @ 10 Hz IFBW
Total average (rms) noise power calculated as the mean value of a linear magnitude trace expressed in
dBm.
May typically be degraded at particular frequencies below 500 MHz due to spurious receiver residuals.
Description Specification Typical
Options 201, 205, 219, 220, 401, 405, 419, 420 Options 201, 205, 219, 220, 401, 405, 419, 420
10 MHz to 50 MHz
--
-116
50 MHz to 100 MHz
-104
-124
100 MHz to 500 MHz
-113
-133
500 MHz to 1 GHz
-122
-142
1 GHz to 2 GHz
-126
-133
2 GHz to 10 GHz
-125
-133
10 GHz to 13.5 GHz
-125
-134
13.5 GHz to 24 GHz
-127
-135
24 GHz to 26.5 GHz
-128
-135
26.5 GHz to 30 GHz
-117
-126
30 GHz to 35 GHz
-116
-125
35 GHz to 40 GHz
-114
-123
40 GHz to 45 GHz
-111
-120
45 GHz to 50 GHz
-110
-120
50 GHz to 67 GHz
-109
-118
67 GHz to 70 GHz
--
-116
Table 51. 0.1 dB Receiver Compression at Test Port (dBm), Option 201, 219, 401, 419, and All LFE Options - Typical
Description Test Port Power (dBm)
10 MHz to 100 MHz
1
15
100 MHz to 30 GHz
12
30 GHz to 67 GHz
11
1
With an LFE option installed and LFE disabled, applied to frequencies 100 MHz. Above 100 MHz, performance is the
same for both LFE enabled or disabled. For LFE enabled performance 100 MHz, see Table Table 52.
Find us at www.keysight.com Page 53
Table 52. 0.1 dB Compression, All Ports, All LFE Options (LFE Enabled) - Typical
Description Test Port Power (dBm)
500 Hz to 900 Hz
13
900 Hz to 1 kHz
13
1 kHz to 10 kHz
13
10 kHz to 100 kHz
13
100 kHz to 1 MHz
13
1 MHz to 5 MHz
11
5 MHz to 10 MHz
13
10 MHz to 50 MHz
14
50 MHz to 100 MHz
14
Table 53. Receiver Compression at Test Power - Specification
Description Test Port Power (dBm) Receiver compression
Option 200, 400
Option 201, 205,
401, 405
Option 219, 220,
419, 420
Magnitude (dB) Phase (degrees)
10 MHz to 500 MHz
1
--
--
--
--
--
500 MHz to 2 GHz
10
13
13
0.15
1.2
2 GHz to 3.2 GHz
7
12
11
0.15
1.2
3.2 GHz to 10 GHz
10
13
12
0.15
1.2
10 GHz to 13.5 GHz
8
11
9
0.15
1.2
13.5 GHz to 16 GHz
8
12
10
0.15
1.2
16 GHz to 20 GHz
6
11
8
0.15
1.2
20 GHz to 24 GHz
6
11
8
0.15
1.2
24 GHz to 30 GHz
6
10
8
0.15
1.2
30 GHz to 35 GHz
5
9
8
0.15
1.2
35 GHz to 40 GHz
0
8
8
0.15
1.2
40 GHz to 67 GHz
4
10
8
0.15
1.2
1
Test port receiver compression at input levels below 500 MHz is negligible due to coupler roll off.
Find us at www.keysight.com Page 54
Table 54. Compression, All Ports, All LFE Options (LFE Enabled)- Specification
Description Test Port Power (dBm) Receiver Compression
All Options Magnitude (dB) Phase (degrees)
500 Hz to 900 Hz
--
--
--
900 Hz to 1 kHz
10
0.2
1
1 kHz to 10 kHz
12
0.2
1
10 kHz to 100 kHz
12
0.2
1
100 kHz to 1 MHz
12
0.2
1
1 MHz to 5 MHz
10
0.2
1
5 MHz to 10 MHz
9
0.2
1
10 MHz to 50 MHz
8
0.2
1
50 MHz to 100 MHz
8
0.2
1
Table 55. Trace Noise
1
Magnitude (dB rms)
Description Specification Typical
1 kHz IFBW 1 kHz IFBW 100 kHz IFBW 600 kHz IFBW
10 MHz to 50 MHz
2
0.05
0.0177
0.173
0.416
50 MHz to 100 MHz
2
0.006
0.0012
0.012
0.029
100 MHz to 500 MHz
0.002
0.0006
0.006
0.014
500 MHz to 1 GHz
0.002
0.0004
0.003
0.006
1 GHz to 26.5 GHz
0.002
0.0005
0.002
0.005
26.5 GHz to 50 GHz
0.003
0.0006
0.005
0.012
50 GHz to 67 GHz
0.003
0.0007
0.006
0.013
67 GHz to 70 GHz
--
0.0010
0.007
0.016
1
Ratioed measurement, nominal power at test port.
2
With an LFE option installed and LFE disabled, applied to frequencies 100 MHz. Above 100 MHz, performance is the
same for both LFE enabled or disabled. For LFE enabled performance 100 MHz, see Table 56.