EN 3.pdf - 第324页
Model DECAN Series Company SAMSUNG TECHWIN Module 3.18 V ACUUM DEVICE V ersion 01 Tit le V ACUUM PUMP REPLACEMENT Page 3-18-16 Note 3. Replace it with a new V acuum Pump and connect the air pressure tube.

Model
DECAN Series
Company
SAMSUNG TECHWIN
Module
3.18 VACUUM DEVICE
Version
01
Title
VACUUM PUMP REPLACEMENT
Page
3-18-15
Note
[VACUUM PUMP REPLACEMENT]
1. Separate the air pressure tube connected to the vacuum tube in the bottom front of equipment.
2. Remove the fixing bolts from the Vacuum Pump to separate it.

Model
DECAN Series
Company
SAMSUNG TECHWIN
Module
3.18 VACUUM DEVICE
Version
01
Title
VACUUM PUMP REPLACEMENT
Page
3-18-16
Note
3. Replace it with a new Vacuum Pump and connect the air pressure tube.

Model
DECAN Series
Company
SAMSUNG TECHWIN
Module
3.18 VACUUM DEVICE
Version
01
Title
VACUUM PUMP(OPTION) REPLACEMENT
Page
3-18-17
Note
[VACUUM PUMP(OPTION) REPLACEMENT]
1. Separate the air pressure tube connected to the vacuum tube in the bottom front of equipment.
2. Remove the fixing bolts from the Vacuum Pump to separate it.
3. Separate the fitting connected to the Vacuum Pump.
4. Replace it with a new Vacuum Pump and open the power distribution panel to wire the cable as follows.
5. Assemble the Vacuum Pump into the equipment and connect the air pressure tube.