4OM-1840 1-4-7 1604-001 4.2 V acuum System V acuum Pressure Every Day Inspection Required T ime: 1 minute Fig. 4A4-8 4.2 V acuum System

100%1 / 315
4OM-1840
1-4-61604-001
 
Filter Holder
Fig. 4A4-7
 
 
 
Note

4.1 Air Source
4OM-1840
1-4-71604-001
4.2 Vacuum System
Vacuum Pressure
Every Day Inspection
Required Time: 1 minute
Fig. 4A4-8
4.2 Vacuum System
4OM-1840
1-4-81604-001
4.2.1 Vacuum Pressure Check
Cycle : Every Day (Required Time: 1 minute)
Vacuum Pressure
Fig. 4A4-9
Procedure
Check the vacuum pressure visually.
Vacuum Valve Fully Closed : Min. -93 kPa (Min. 70 cmHg) or more
Running : Min. -60 kPa (Min. 45 cmHg) or more
Note

line or the vacuum pump may have broken down.