Process Lens PL Service Manual_EN - 第52页

3 Replacing spare parts 52 Se rv ic e Ma nu al P ro ce ss L en s PL - 0 3/ 20 25 3.3.3 Checking the alignment of the encoder reader of the Y gantry Whenever the encoder readers and/or scale bars have been removed, they n…

100%1 / 168
3 Replacing spare parts
Service Manual Process Lens PL - 03/2025 51
Fig.60: Calibration of the new reader
Check the gap between the encoder
reader(1) and the scale bar(2) by using
the spacer(3) provided.
The gap between the encoder reader and
the scale bar must be 2.1 mm.
Calibrate the new reader. See the external
Power Point file “Y1Y2 ENC install & ad-
just procedure”.
Move the gantry slowly the whole travel
range backwards. Whilst doing that check
the light at the indicator.
If the lights are not within the color range
the holder needs adjustments. See 3.3.2
"Color code for the encoder readers"
[}51].
Proceed the same way with all other en-
coder readers.
3.3.2 Color code for the encoder readers
Color Indication Adjustment requirement
Purple Optimum set-up No
Blue Good set-up No
Green Good set-up No
Orange Acceptable set-up but below recommendation level. Yes, angle and gap.
Red Poor set-up, signal maybe too low for reliable oper-
ation
Yes, angle and gap.
3 Replacing spare parts
52 Service Manual Process Lens PL - 03/2025
3.3.3 Checking the alignment of the encoder reader of the Y gantry
Whenever the encoder readers and/or scale bars have been removed, they need to be realigned.
Requirements
Machine is switched on.
Optical head is centered on the X position.
Checking the alignment
Fig.61: “Unit Function” tab
Select the Unit Function tab(2) on the Diagnostic Page(1).
Check the value differences between YL & YR under Encoder Info(3). Adjust them if necessary
(see 3.3 "Encoder" [}50])
Click Motion Initialisation(4) to reset all NuMotion values.
Fig.62: “System Test” tab
3 Replacing spare parts
Service Manual Process Lens PL - 03/2025 53
Select the System Test tab(1).
Ignore the message “Please lock the flexure first” in the message window (2) under Gantry Par-
allel Calibration. This step is not necessary.
Click Start(3) to continue, the calibration status will update on the message window.
A pop-up message will show the encoder count. Adjust the encoder if the value does not fall
within 10,000-digit. See also 3.3 "Encoder" [}50].
Fig.63: Message showing encoder counts
Click Ok, a success message will appear at the end.
Fig.64: Success message
Click Ok and the calibration is completed.
3.3.4 Checking the alignment of the x, y and z encoder reader
Furthermore the alignment of the x, y and z readers need to be checked whenever the readers or the
scale bar has been changed.
Requirements
Machine is switched on.
Checking the alignment
Fig.65: Checking the alignment
Go to the Diagnostic page.
Click on the Unit function(1) tab.
Go to the Gantry section(2).
Select the X button(3) and click on Home(4).
Select the Y button(3) and click on Home(4).
Select the Z button(3) and click on Home(4).