DMM4020.pdf - 第8页
Datasheet DC Current Characteristic Description Input Protection T ool-accessible 1 1 A / 1000 V and 440 mA / 1000 V fuses Shunt Res ist ance 0.01 Ω for 2 A a nd 10 A range s 1 Ω for 2 0 mA an d2 0 0 m A Burden volta ge …

Digital Multimeters — Tektronix DMM4020
Resistance
Specifications
are for 4-wire resistance function, or 2-wire resistance with REL. If
REL is not used, add 0.2 Ω for 2 -wire resistance plus lead resistance.
Characteristic
Description
Measurement Method
Current source referenced to LO input
Max Lead Resistance
(4-wire ohms)
10% of range per lead for 200 Ω,2kΩ ranges. 1 kΩ
per lead on all other ranges
Input Protection 1000 V on all ranges
Input Characteristics
ResolutionRange
Full Scale
(5½ Digits)
Slow
Medium Fast
Current Source
200 Ω 199.999 Ω 0.001 Ω 0.01 Ω 0.01 Ω
0.8 mA
2kΩ 1.99999 kΩ 0.01 Ω 0.1 Ω 0.1 Ω
0.8 mA
20 kΩ 19.9999 kΩ 0.1 Ω 1 Ω 1 Ω
0.08 mA
200 kΩ 199.999 kΩ 1 Ω 10 Ω 10 Ω
0.008 mA
2MΩ 1.99999 MΩ 10 Ω 100 Ω 100 Ω
0.9 μA
20 MΩ 19.9999 MΩ 100 Ω 1kΩ 1kΩ
0.16 μA
100 MΩ 100.000 MΩ 1kΩ 10 kΩ 10 kΩ 0.16 μA||10MΩ
Accuracy
Uncertai
nty*
3
90 days 1 year
Range
23 °C ±5 °
C
23 °C ±5 °
C
Temp erat
ure Coefficient/°C
Outside 1
8-28°C
200 Ω
0.02 + 0.004 0.03 + 0.004 0.003 + 0.0006
2kΩ
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
20 kΩ
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
200 kΩ
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
2MΩ
0.03 + 0.003 0.04 + 0.004 0.004 + 0.0005
20 MΩ
0.2 + 0.003 0.25 + 0.003 0.01 + 0.0005
100 MΩ
1.5 + 0.004 1.75 + 0.004 0.2 + 0.0005
*
3
Uncertainty given as ±(% of reading + % of range).
www.tektronix.comdmm 7

Datasheet
DC Current
Characteristic
Description
Input Protection
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses
Shunt Resist
ance
0.01 Ω for 2 A a
nd 10 A ranges
1 Ω for 20 mA an
d200mA
Burden volta
ge <5 mV for 200 μA and 2 mA range
Input Characteristics
ResolutionRange
Full Scale
(5½ Digits)
Slow
Medium Fast
Burden Voltage
200 μA 199.999 μA0.001μA0.01μA0.01μA<5mV
2 mA 1999.99 μA0.01μA0.1μA0.1μA<5mV
20 mA 19.9999 mA 0.1 μA1μA1μA <0.05 V
200 mA 19.9999 mA 1 μA10μA10μA<0.5V
2 A 1.99999 A 10 μA100μA100μA<0.1V
10 A 10.0000 A 100 μA1mA 1mA<0.5V
Accuracy
Uncertainty*
3
90 days 1 year
Range
23 °C ±5 °C 23 °C ±5 °C
Te mp eratu re Coefficient/°C
Outside18-28°C
200 μA 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001
2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001
20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001
200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001
2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001
10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
*
3
Uncertainty given as ±(% of reading + % of range).
8 www.tektronix.comdmm

Digital Multimeters — Tektronix DMM4020
AC Current
The following A
C current specifications are for sinusoidal signals with amplitudes
greater than 5% of range. For inputs from 1% to 5% of range, add an additional
error of 0.1% of range.
Characteristic
Description
Input Protection
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses
Measurement Method
AC-coupled true RMS
Shunt Resistance 0.01 Ω for 2 A and 10 A ranges
1 Ω for 20 mA and 200 mA
AC Filter Bandwidth
20 Hz - 100 kHz
Maximum C rest Factor 3:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1-2, 0.05% of full scale
CrestFactor2-3,0.2%offullscale
Only applies to non-sinusoid signals
Input Characteristics
ResolutionRange
Full Scale
(5½ Digits)
Slow
Medium Fast
Burden Voltage
20 mA 19.9999 mA 0.1 μA1μA1μA <0.05 V
200 mA 199.999 mA 1 μA10μA10μA<0.5V
2 A 1.99999 A 10 μA100μA100μA<0.1V
10 A 10.0000 A 100 μA1mA 1mA<0.5V
Accuracy
Uncertainty*
3
90 d ays 1 year
Range Frequency
23 °C ± 5 °C 23 ° C ±5 °C
Temperature C oef ficien t/°C
Outside 18 - 28 °C
20 Hz - 45 Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.00520 mA
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
20 Hz - 45 Hz 0.8 + 0.05 1 + 0.06 0.015 + 0.005200 mA
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
20 Hz - 45 Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.0052A
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
20 Hz - 45 Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.00510 A
45 Hz - 2 kHz 1 + 0.1 0.5 + 0.12 0.015 + 0.005
*
3
Uncertainty given as ±(% of reading + % of range).
www.tektronix.comdmm 9