DFD6361-Maintenance - 第313页
D-28 E0019 : Insu fficient clean air pr essure. Error c ode E0019 Message Insuffici ent clean a ir pressu re. W hat wa s detected The clean a ir supply pressure has dropped to below the lower lim it defined on the SENSOR…

D-27
E0015 : Water case overflow error.
Error code
E0015
Message
Water case overflow error.
What was
detected
There was an input to the overflow sensor because the water level in the
water case exceeds the prescribed water level.
During what
operation it
occurred
While the wheel coolant water is ON and during spinner cleaning.
State after
detection
The Z-axis retreats to the upper end. The spindle stops rotating. The X-, Y-
and θ-axes immediately stop. Cleaning operation stops, if the error occurs
during spinner cleaning.
Phenomenon Probable causes Remedy Reference
What to do
Thedrainageinthe
water case is poor.
The wastes strainer
of the drain is
clogged.
Clean the drain
wastes strainer.
The error occurs
regardless of the
water level in the
water case.
The overflow sensor
is defective.
Replace the sensor.
The overflow sensor
line is defective.
Contact your
nearest DISCO
Service Office.
Recovery
(1) Cancel the alarm.
(2) Clean the wastes strainer in the water case.
(3) Turn ON the wheel coolant water only and check whether the water
drains properly.
(4) Press the <START> button to resume the operation or press the <EXIT>
button to discontinue the operation.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear

D-28
E0019 : Insufficient clean air pressure.
Error code
E0019
Message
Insufficient clean air pressure.
What was
detected
The clean air supply pressure has dropped to below the lower limit defined
on the SENSOR THRESHOLD DATA screen [screen 6.6].
During what
operation it
occurred
At any time.
State after
detection
The Z-axis retreats to the upper end. The spindle stops rotating. The X-axis
returns to a location under the microscope. The Y- and θ-axes decelerate
and stop.
Phenomenon Probable causes Remedy Reference
What to do
The error occurs
when the clean air
supply pressure is
lower than the
minimum limit.
The supply pressure
of clean air is too
low.
Check the supply
pressure of clean
air.
Check that the
clean air piping is
not clogged or bent.
The error occurs
when the clean air
supply pressure is
as same as or
higher than the
minimum limit.
The limit setting of
the pressure sensor is
improper.
Correct the setting
of the pressure
sensor to 0.2MPa.
The pressure sensor
is detective
Replace the sensor.
Recovery
(1) Cancel the alarm.
(2) Press the <Sensor> button to call up the SENSOR STATUS screen.
(3) Make sure that the air pressure gauge reads 0.5MPa.
(4) Effect system initialization.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear

D-29
E0020 : Handling cover sensor error.
Error code
E0020
Message
Handling cover sensor error.
What was
detected
There was an input on the safety switch of the arm section cover.
During what
operation it
occurred
While the X-, Y-, Z-, θ -axis, lower arm and upper is operating.
The cover does not open during the spindle rotation.
State after
detection
The Z-axis retreats to the upper end and the spindle stops.
The X-, Y-, θ-, spinner-axis and the lower arm immediately stop.
The upper arm stops after completing one cycle of operation.
Phenomenon Probable causes Remedy Reference
What to do
The error occurs
even when the
cover is closed.
Thesafetyswitchis
defective.
Replace the safety
switch.
Contact your
nearest DISCO
Service Office.
The sensor line is
broken.
Contact your
nearest DISCO
Service Office.
Recovery
(1) Close the arm section cover.
(2) Cancel the alarm.
(3) Effect the system initialization.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear