DFD6361-Maintenance - 第345页

D-60 B0586 : Z 1-axi s non-c ontact set up reap pearance er ror. Error c ode B0586 Option al accesso ry Message Z1- axis non-contact setup reappearan ce error. W hat wa s detected The di fference b etween the fir st and …

100%1 / 720
D-59
B0585 : Z1-axis setup error (No detection.)
Error code
B0585
Message
Z1-axis setup error (No detection.)
What was
detected
Setup was not completed when the Z1-axis has reached the software defined
lower-end.
During what
operation it
occurred
During Z1-axis setup operation.
State after
detection
The Z1-axis returns to its origin and the setup operation terminates.
Setup data value registration does not take place.
Phenomenon Probable causes Remedy Reference
What to do
The error recurs
when setup is
attempted again.
The Z1-axis blade
outside diameter
entry is incorrect.
Correct the Z1-axis
blade outside
diameter value
entered from the
BLADE
REPLACEMENT
screen.
The servicing maker
data is improper.
Contact your
nearest DISCO
Service Office.
Recovery
(1) Cancel the alarm.
(2) Select the setup operation again.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear
D-60
B0586 : Z1-axis non-contact setup reappearance error.
Error code
B0586
Optional accessory
Message
Z1-axis non-contact setup reappearance error.
What was
detected
The difference between the first and second Z1-axis non-contact setups
exceeded the limit that had been defined in the SETUP DATA 1 screen
[screen 4.7].
During what
operation it
occurred
After Z1-axis setup operation.
State after
detection
The X- and Y-axes move to their origins and the alarm indication is given.
Setup data value registration does not take place.
Phenomenon Probable causes Remedy Reference
What to do
The error recurs
when setup is
attempted again.
The preselected limit
value is too small.
Check whether the
limit value setting
is 0.005 mm
(standard value).
Check whether the
non-contact setup
sensor is soiled.
Clean the non-
contact setup
sensor section and
perform sensor
calibration setup.
Erroneous detection
occurs during setup
operation due to
water drops on the
non-contact setup
sensor section.
Correct the setup
data value for the
setup blow time
setting.
Recovery
(1) Cancel the alarm.
(2) Select the setup operation again.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear
D-61
B0587 : Z1-axis blade wear check error.
Error code
B0587
Message
Z1-axis blade wear check error.
What was
detected
The Z1-axis setup data for the blade wear check limit value was exceeded
by the difference between the current and previous setup values.
During what
operation it
occurred
After Z1-axis setup operation.
State after
detection
Setup data value registration does not take place.
Phenomenon Probable causes Remedy Reference
What to do
The error recurs
when setup is
attempted again.
Erroneous detection
occurs during setup
operation due to
water drops on the
setup position
section.
Increase the setup
data value for the
setup blow time
setting.
The non-contact
setup sensor is
soiled.
Clean the non-
contact setup
sensor.
The setup data value
for the Z1-axis blade
wear check limit
setting is too small.
Enter a limit setting
appropriate for the
Z1-axis blade wear
amount.
Recovery
(1) Cancel the alarm.
(2) Perform the customer-defined procedure for excessive blade wear.
(3) Perform the setup operation again.
Maintenance level Maintenance personnel level Hazard level