D-86 V0624 : Pressure of S/ T high pressure w a ter risen. Error c ode V0624 Option al accesso ry Message Pressure of S/T hig h pressu re water ri sen. W hat wa s detected During oper ation of the sp inner high- pressure…

100%1 / 720
D-85
V0615 : Error occurs in UV lamp.
Error code
V0615
Message
Error occurs in UV lamp.
What was
detected
Interception of light was detected by the ultraviolet light receiving sensor,
located at the lower part of the UV irradiation lamp.
During what
operation it
occurred
While the UV irradiation lamp is ON.
State after
detection
UV irradiation will stop. The glass plate will open. N
2
supply will stop.
When the value of <UV Lamp ON Timing> on UV SENSOR
ADJUSTMENT screen [screen 6.5] is set to "UV", the UV irradiation
lamp will light off.
Phenomenon Probable causes Remedy Reference
What to do
The error occurs
even though the
UV irradiation
lamp functions
normally.
The sensor wiring is
broken.
Contact your
nearest DISCO
Service Office.
The sensor is
defective.
Replace the
sensor.
The UV irradiation
lamp is burned
out.
Replace the UV
irradiation lamp.
Recovery
(1) Cancel the alarm, and then the ERROR RECOVERY (UV) screen is
displayed.
(2) Perform the error recovery steps indicated on the screen.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear
D-86
V0624 : Pressure of S/T high pressure water risen.
Error code
V0624
Optional accessory
Message
Pressure of S/T high pressure water risen.
What was
detected
During operation of the spinner high-pressure pump, the jet pressure of the
cleaning water is higher than the sensor setting.
During what
operation it
occurred
During spinner cleaning.
State after
detection
The cleaning operation stops. The nozzle axis moves to its origin.
The spinner table moves upward.
Phenomenon Probable causes Remedy Reference
What to do
Increase of the jet
pressure.
The spinner nozzle is
clogged.
Replace the spray
tip (high-pressure
type).
F-1-5
The error occurs
even though the jet
pressure is normal.
The pressure sensor
wiring is broken.
Contact your
nearest DISCO
Service Office.
The sensor is
defective.
Replace the sensor.
The pressure sensor
setting is too low.
Correct the sensor
setting.
Recovery
(1) Cancel the alarm, and then the ERROR RECOVERY (CLEANING)
screen is displayed.
(2) Perform the error recovery steps indicated on the screen.
Maintenance level Maintenance personnel level Hazard level
Measuring equipment used Tools Replacement parts Protective gear
D-87
V1014 : Up error (Lower arm Upper cylinder).
Error code
V1014
Message
Up error (Lower arm Upper cylinder).
What was
detected
There is no input from Up-end sensor, when the lower arm upper cylinder
moves upward.
During what
operation it
occurred
Lower arm up operation.
State after
detection
Transport stops.
Phenomenon Probable causes Remedy Reference
What to do
The error occurs
even though the
lower arm upper
cylinder ascends
normally.
The sensor position
is deviated.
Adjust the position
so that there is
sensor input when
the lower arm
upper cylinder is in
the up state.
The sensor is
defective.
Replace the sensor.
The lower arm
upper cylinder does
not ascend.
The upper cylinder
piping of the lower
arm section is bent of
clogged.
Contact your
nearest DISCO
Service Office.
Recovery
(1) Cancel the alarm, and then the ERROR RECOVERY (HANDLING)
screen is displayed.
(2) Perform the error recovery steps indicated on the screen.
Maintenance level Maintenance personnel level Hazard level