A-34 Microscope sec tion mechanism The microscope section is structured as follows: Function The workpiece front side image is magnified b y the microscope and displayed on the monitor screen. Feature - Automatic focusin…

A-33
Microscope section component names (Continued)
[2-axis]
FRONT
High magnification microscope
CCD camera
[High magnification microscope]
High magnification (
×
7.5)
Micro direct light
source (LED)
Fiber of micro
oblique light source

A-34
Microscope section mechanism
The microscope section is structured as follows:
Function The workpiece front side image is magnified by the
microscope and displayed on the monitor screen.
Feature - Automatic focusing is done with an auto focus
function.
- Switch between high and low magnifications.
Illumination
adjustment
The microscope illumination level can be adjusted with
the LIGHTING ADJUSTMENT screen.

A-35
2-11. Spindle-axis Section
Spindle-axis section component names
The spindle-axis section consists of the following components.
FRONT
Z1-axis
spindle holder
Z2-axis
spindle holder
Z1-axis spindle Z2-axis spindle
Spindle-axis section mechanism
The spindle-axis section is structured as follows:
Driving source Driven by a DC brushless motor.
Axis supporting section An air bearing is used.