SEMI E33-94 © SEMI 1994, 19965
ELF emissions shall be measured using a calibrated
ELF meter. Equipment emissions shall be measured at a
distance of 1.5 meters from the entire perimeter of the
EUT at a height of 1 meter. Facility ELF shall be
measured at a height of 1 meter in the region where
ELF-sensitive equipment installation is intended.
Measurement points shall be no greater than 0.5 meter
apart in the regions specified above. Facilities electrical
systems and adjacent equipment should be in normal
operation to assure accurate measurement of the actual
operating environment.
At each measurement point, the ELF sensor should be
oriented to produce the maximum reading. The
maximum from all the measurement points is then used
to determine the level from the table below:
Level A — ELF of less than 0.25 milliGauss rms
Level B — ELF of less than 0.50 milliGauss rms
Level C — ELF of less than 1.00 milliGauss rms
Level D — ELF of less than 2.00 milliGauss rms
Level E — ELF of 2.00 milliGauss rms and greater
6 Equipment Installation and Grounding
Special installation, power, and grounding requirements
necessary to meet this standard shall be thoroughly
documented by the equipment supplier. FIPS
Publication 94 and the IEEE Emerald Book are
suggested sources of information for recommended
practices in this area.
Removable shielded panels or doors that must be
properly secured to meet this standard should be labeled
“SECURE THIS PANEL TO REDUCE
ELECTROMAGNETIC EMISSIONS.” This labeling
requirement is not applicable to interlocked panels that
must be in place for the equipment to operate.
7 Documentation
Upon request of the purchaser a report of the test results
against this standard shall be furnished. Similarly, the
supplier may use this standard as a reference in
specifying the environmental requirements of
equipment.
8 Alternate Testing
Testing completed under a different, but substantially
similar standard may be substituted for the basic
standards specified herein if the performance criteria
and test levels are similar to those above. Such
substitution shall be noted in the test report, and must
be agreed to by the concerned parties. Examples
include substitution of ANSI C63.4-1991 testing for
CISPR 22.
Similarly, equipment which satisfies the European
EMC Directive and bears the CE mark meets this
specification.
NOTICE: These standards do not purport to address
safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
SEMI makes no warranties or representations as to the
suitability of the standards set forth herein for any
particular application. The determination of the
suitability of the standard is solely the responsibility of
the user. Users are cautioned to refer to manufacturer’s
instructions, product labels, product data sheets, and
other relevant literature respecting any materials
mentioned herein. These standards are subject to
change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.

SEMI E33-94 © SEMI 1994 6
APPENDICES
The appendices are not an official part of this standard.
However, they contain relevant information for
customizing the basic standards for testing of
semiconductor equipment and facilities, and point out
areas of concern for which no standard currently exists.
A.1 Appendix A: Charged Wafer Testing
Due to overriding microcontamination concerns, non-
static-dissipative wafer carriers must be used during
semiconductor fabrication. It is suggested that any
equipment which handles wafers be tested with wafers
charged to 18 kV/in. The charge generator used for
801-2 testing can be used to charge the wafer. Some
experimentation will be required to determine the best
means to carry out this test. Alternately, the handling
mechanism may be challenged by air or contact
discharge directed to the mechanism.
It is suggested that wafer handlers be designed to
dissipate charge on incoming wafers in a controlled,
non-sparking manner. A simple AM radio tuned
between stations and placed near the handler has been
found to be a good qualitative means to detect sparking.
Sparking will result in pops and static being heard on
the radio.
A.2 Appendix B: ESD Test Methods
Although the contact method of ESD testing has been
found to be much more reproducible than the air
discharge, the air discharge is a more realistic
simulation of possible events that can occur in the
semiconductor factory. Also, the air discharge method
is the only realistic way to test objects that are primarily
fabricated from insulators, such as wet stations and
keyboards.
In addition to testing surfaces that come in contact with
humans and wafers, it is important to test any area of
the equipment that can be exposed to ESD. For
instance, wet stations are composed of assemblies
involving large masses of insulators in proximity to
metallic sensors and controllers. It is possible for the
insulating surfaces to build up charge, which then
sparks to a sensor or controller — all internal to the
equipment itself.
A.3 Appendix C: Future Directions
Items in this section are not requirements of this
standard, but are included to indicate currently ill-
defined goals for further improvements to
semiconductor factory EMC.
A.3.1 Static Charge — Minimization of static charge
promotes cleanliness, and avoids damage to material in
process and reticles.
It is suggested that electrostatic fields be held to under
200 V/cm (500 V/in). This would be applicable to
personnel, wafers, cassettes, cassette boxes, reticles,
reticle boxes, and all equipment and facility surfaces in
proximity to these items as well as all surfaces exposed
to filtered airflow. Measurements shall be made using a
properly-grounded electrostatic field meter in
accordance with its manufacturer's instructions.
A.4 Appendix D: National Requirements
A.4.1 U.S. — Section 5.1.1 Table 1, row 1.2: Radiated
immunity testing outside a shielded enclosure is legally
accomplished by obtaining an experimental license
from the FCC using FCC form 442. The FCC
recommends using techniques such as scanning or
hopping to avoid interference.
A.4.2 Europe — The earth port: Refer to Table 1.6 of
the informative annex to EN 50 082-2 for test
requirements applicable to the earth.
NOTICE: These standards do not purport to address
safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
SEMI makes no warranties or representations as to the
suitability of the standards set forth herein for any
particular application. The determination of the
suitability of the standard is solely the responsibility of
the user. Users are cautioned to refer to manufacturer’s
instructions, product labels, product data sheets, and
other relevant literature respecting any materials
mentioned herein. These standards are subject to
change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI E34-95 © SEMI 19951
SEMI E34-95
GUIDELINE FOR MASS FLOW DEVICE RETURN
NOTICE: This safety guideline does not purport to
address all of the safety issues associated with its use. It
is the responsibility of the users of this guideline to
establish appropriate safety and health practices and
determine the applicability of regulatory limitations
prior to use.
1 Purpose
Mass flow devices (mass flow controllers, mass flow
meters and flow control valves) are used as an integral
part of the chemical delivery system in semiconductor
processing equipment. Many of the chemicals used in
semiconductor processes can be hazardous to
equipment and personnel. Once exposed to these
chemicals, the contaminated mass flow devices can
present a hazard to repair personnel and equipment if
not handled properly.
Implementing the procedures outlined in this guideline
provide a mechanism to reduce the risk associated with
shipping and/or receiving any potentially hazardous
mass flow device used to monitor or control gases or
liquids.
2 Scope
This guideline is intended to:
Recommend and encourage the purging of devices used
with hazardous materials prior to removal to reduce the
risk of handling them.
Alert the recipient of possibly hazardous mass flow
devices to the nature of the hazard.
Establish mechanism for properly identifying and
documenting possibly hazardous mass flow devices by
using an orange tag and a health and safety disclosure
form.
3 Referenced Documents
49 CFR
1
— Title 49 of the Code of Federal
Regulations (CFR)
NFPA 704
2
— Standard System for the Identification
of Fire Hazards of Materials
1 United States Government Printing Office, Washington, D.C. 20402
2 National Fire Protection Association, Batterymarch Park, Quincy,
MA 02269
4 Terminology
4.1 hazardous materials — Those chemicals or
substances that are physical hazards or health hazards
as defined and classified in NFPA 704 whether the
materials are in use or in waste conditions.
4.2 mass flow controller (mfc) — A self-contained
device, consisting of a mass flow transducer, control
valve, and control- and signal-processing electronics,
commonly used in the semiconductor industry to
measure and regulate the mass flow of gas.
4.3 mass flow meter (mfm) — A self-contained device,
consisting of a mass flow transducer and signal-
processing electronics, commonly used in the
semiconductor industry to measure the mass flow of
gas.
4.4 material safety data sheet (msd s) — Written or
printed material concerning a hazardous material which
is prepared in accordance with the provisions of 29
CFR 1910.1200.
4.5 protective container — A sealable plastic bag or
other container which will keep hazardous material
from the mass flow device from contaminating the
outer package. The container should be transparent, if
possible.
4.6 purge — To dilute potentially harmful material in
the mass flow device by flowing an inert gas through
the device. Purging a chemical delivery line containing
a mass flow device with an inert substance is intended
to dilute hazardous materials and reduce the level of the
hazard. To be effective, the inert substance must be able
to reach all points within the chemical delivery system
in sufficient quantity to dilute the hazardous material to
safe levels. The nature of the hazard and the physical
configuration of the chemical delivery system must be
considered when developing a purge procedure.
4.7 safe — Free of conditions that can cause
occupational illness, injury, or death to personnel or
damage to or loss of equipment or property or the
environment.
5 Procedure
5.1 Disclosure Form
• Complete the disclosure form for mass flow device
return.
• Send the completed form via Fax or other suitable
method to the recipient of the mass flow device.