SEMI F29-0997 © SEMI 1997, 1103 1 SEMI F29-0997 (Reapproved 1103) TEST METHOD FOR PURGE EFFI CACY OF GAS SOURCE SYSTEM PANELS This test method was technically approved by the Global Gases Committee and is the dire ct res…

SEMI F28-1103 © SEMI 1997, 2003 7
6.4.2.6 In order to determine if there is statistical
evidence of the component having an effect on the
observed particle count, a test must be conducted to see
if the actual difference between the average background
count, X
B
, and the average test count, X
t
, exceeds two
standard errors in a distribution of differences between
means. The average particle count generated by the test
component, X
c
, is given by:
X
c =
X
t −
X
B
6.4.2.7 The standard error in a distribution of
differences between means is expressed as:
S
ec
= S
et
2
+ S
eB
2
6.4.2.8 The 95% confidence interval of the test
component itself is then determined by:
X
c ± 2 S
ec
6.4.2.9 If this confidence interval includes 0, then this
implies that there is not strong statistical evidence of
the component having an effect on the observed particle
count. If this interval does not include 0, then this
implies there is strong statistical evidence that the
component does have an effect on the observed particle
count.
6.4.2.10 Present in tabular form the average particle
count and the associated upper and lower confidence
limits (as calculated in Section 6.4.2) for each test state.
In addition, present the background test data separately
as a distinct element.
7 Related Documents
Agarwal, J. K. and Sem, G. J., “Continuous Flow,
Single Particle Counting Condensation Nucleus
Counter”, Journal of Aerosol Science, v.11.4. July
1950: 343–357
Fissan, H. and Schwientek, “Sampling and Transport of
Aerosols”, TSI Journal of Particle Instrumentation,
v.2.2. July–December 1987: 3–10
Hinds, W. C. Aerosol Technology: “Properties,
Behavior, and Measurement of Airborne Particles”,
John Wiley & Sons. 1982: 187–194
VanSlooten, R. A., “Statistical Treatment of Particle
Counts in Clean Gases”, Microcontamination, v.4.2.
February 1986: 32–38
ANSI/ASME B46.1
2
— “Specification for Surface
Texture - Surface Roughness, Waviness, and Lay”
NOTICE: SEMI makes no warranties or
representations as to the suitability of the standards set
forth herein for any particular application. The
determination of the suitability of the standard is solely
the responsibility of the user. Users are cautioned to
refer to manufacturer’s instructions, product labels,
product data sheets, and other relevant literature
respecting any materials mentioned herein. These
standards are subject to change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.
2 American Society of Mechanical Engineers, Three Park Avenue,
New York, NY 10016-5990, USA. Telephone: 800.843.2763
(U.S./Canada), 95.800.843.2763 (Mexico), 973.882.1167 (outside
North America)Website: www.asme.org
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI F29-0997 © SEMI 1997, 1103 1
SEMI F29-0997 (Reapproved 1103)
TEST METHOD FOR PURGE EFFICACY OF GAS SOURCE SYSTEM
PANELS
This test method was technically approved by the Global Gases Committee and is the direct responsibility of
the North American Gases Committee. Current edition approved by the North American Regional Standards
Committee on September 3, 2003. Initially available at www.semi.org September 2003; to be published
November 2003. Originally published September 1997.
1 Purpose
1.1 This document defines the purge efficacy test
method recommended for determining the minimum
acceptable level of purge efficacy for gas source
systems used in semiconductor manufacturing. It is also
intended as an aid to the procurement of gas source
equipment.
NOTE 1: Users of this specification are encouraged to submit
suggested revisions or other comments to SEMI.
2 Scope
2.1 This specification applies to gas source equipment
used in semiconductor manufacturing facilities and
comparable research and development areas. It includes
contamination testing requirements for gas source
systems.
2.2 The tests covered by this document are as follows:
2.2.1 Purge efficacy with a non-interactive gas using
manufacturers’ standard purge sequence.
2.2.2 Purge efficacy with a non-interactive gas using
the test method specified purge sequence.
2.2.3 Purge efficacy with an interactive gas using
manufacturers’ standard purge sequence.
2.2.4 Purge efficacy with an interactive gas using the
test method specified purge sequence.
NOTICE: This standard does not purport to address
safety issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish
appropriate safety and health practices and determine
the applicability of regulatory or other limitations prior
to use.
3 Limitations
3.1 The document is not intended to determine the safe
operation of a gas source system. The test method
described here is for determining the removal efficacy
of hazardous gas from the pigtail portion of the gas
panel during routine cylinder change operations (pre-
purge). It does not address the purge condition after the
change of the cylinder (post-purge) nor the purge of the
entire gas source system which is usually performed
following its initial installation or during maintenance
(maintenance purge).
3.2 Because the response time of moisture and oxygen
analyzers utilizing different technologies can differ
significantly, this can lead to the possibility that
different instruments will report different peak heights
when measuring identical contaminant spikes.
Therefore, results of tests should only be compared
when similar analyzers are used.
4 Referenced Standards
None.
5 Terminology
5.1 Abbreviations and Acronyms
5.1.1 ppb — Molar parts per billion, same as ppbv.
5.1.2 ppm — Molar parts per million, same as ppmv.
5.2 Definitions
5.2.1 clean dry air (CDA) — filtered air filtered to 0.02
mm and dried to a dew point of at least -80° C.
5.2.2 dry down — removal of residual moisture in a
gas delivery or distribution system, often accomplished
by flowing a stream of high purity dry inert gas
continuously through the system for an extended period
of time.
5.2.3 dwell time — the time for which vacuum or
pressure is applied during the evacuation or
pressurization steps of a purge operation.
5.2.4 interactive gas — a gas that will readily adsorb to
the surface of a vessel used to contain or transport it.
Examples of interactive gases are hydrogen chloride
and moisture.
5.2.5 non-interactive gas — a gas that will not adsorb
to the surface of a vessel used to contain or transport it.
Examples of non-interactive gases are oxygen and
nitrogen.
5.2.6 pigtail — the pigtail is the part of the gas source
equipment that is the flexible connection between the
cylinder and the gas panel.

SEMI F29-0997 © SEMI 1997, 1103 2
5.2.7 pigtail bleed — a pigtail bleed is a reverse flow
of purge gas from the pigtail to minimize atmospheric
intrusion into the gas panel.
5.2.8 purge cycle — following the initial process vent
step, a cycle is defined as a pressurization step followed
by an evacuation step.
5.2.9 purified nitrogen — nitrogen purified to meet the
following characteristics:
Moisture < 20 ppb
Oxygen < 10 ppb
Total hydrocarbons < 1 ppm
CO
2
< 1 ppm
CO < 1 ppm
6 Ordering Information
6.1 Orders for equipment or services requiring purge
efficacy testing in accordance with this test method
shall include the following:
6.1.1 This test method number and date of issue.
6.1.2 Test requirements as defined in the detailed
specifications for the specific product being purchased.
6.1.3 Whether certifications of tests and a report of the
test results is required.
7 Requirements
7.1 Personnel Qualification — Personnel performing
tests in accordance with this specification shall have
suitable training and experience. Such personnel
should, as a minimum,
7.1.1 be trained and experienced in operation of gas
source control equipment,
7.1.2 be trained and experienced in the use of oxygen
and moisture analyzers, and
7.1.3 be familiar with the operation and calibration of
the specific equipment used in performing the tests.
8 Apparatus
8.1 Oxygen Analyzer — The oxygen analyzer should
meet or exceed the following characteristics:
Limit of Detection 10 ppb
Accuracy ± 10 ppb
8.2 Moisture Analyzer — The moisture analyzer
should meet or exceed the following characteristics:
Limit of Detection 10 ppb
Accuracy ± 10 ppb
8.3 Pressure Transducer — The transducer should be
of a high-purity type with minimal dead volume and
meet or exceed the following characteristics:
Accuracy ± 3.4 kPa (0.5 psi)
Hysteresis ± 3.4 kPa (0.5 psi)
Non-Repeatability ± 0.7 kPa (0.1 psi)
8.4 Mass Flow Controller or Flow Meter — The mass
flow controller or flow meter should be accurate to ±
0.2 slpm.
8.5 Moisture Source — The moisture source must be
capable of providing a 2 slpm continuous flow of 2.0 ±
0.2 ppm moisture in nitrogen mixture at 138 kPa (g) [20
psi (g)] for a minimum of 30 minutes.
9 Procedure
9.1 Purge efficacy determination with non-interactive
gas at manufacturer’s recommended pressure, vacuum,
and dwell times.
9.1.1 Connect gas source equipment as in Figure 1.
NOTE 2: To avoid biasing the test results by substantially
increasing piping volume to be purged, the dead space
between the cylinder connection and the valves isolating the
analyzer and the contaminant source must be small compared
to the total pigtail volume. The use of mono-block valve and
flow-through type pressure transducer is highly
recommended.
9.1.1.1 However, the pigtail is connected to the
cylinder valve during the purge in an actual cylinder
change operation. The small additional volume to be
purged in the test setup may be used to account for the
volume of the cylinder valve outlet cavity, which must
be purged during a cylinder change.
9.1.2 To challenge the gas panel with a known
contaminant gas, completely fill and pressurize the gas
source system to 207 kPa (g) [30 psi (g)] with CDA as
measured at the pigtail with the pressure transducer.
9.1.3 Using purified nitrogen as the purge gas,
complete one purge cycle at the manufacturer’s
recommended pressure, vacuum, and dwell times.
9.1.4 Flow purified nitrogen using a pigtail bleed at 1
slpm or at the oxygen analyzer manufacturer’s
recommended flow rate, whichever is higher, into the
oxygen analyzer for 10 minutes or until below the
detection limit of the oxygen analyzer.