semi合集-English.pdf - 第4238页

SEMI F76-0303 © SEMI 2003 2 5.1.4 regulator — a valve desi gned to redu ce a high incoming pressure to a lower ou tlet pressure by automatically opening to allo w flow until a desired, pre-set pressure on the outlet side…

100%1 / 7923
SEMI F76-0303 © SEMI 2003 1
SEMI F76-0303
TEST METHOD FOR EVALUATION OF PARTICLE CONTRIBUTION
FROM GAS SYSTEM COMPONENTS EXPOSED TO CORROSIVE GAS
This test method was technically approved by the Global Gases Committee and is the direct responsibility of
the North American Gases Committee. Current edition approved by the North American Regional Standards
Committee on November 22, 2002. Initially available at www.semi.org December 2002; to be published
March 2003.
1 Purpose
1.1 This is a test method to compare gas handling
components for potential particle generation in
corrosive gas service. It is intended as a practical
means of generating performance data for a group of
components to be compared in a selection process.
2 Scope
2.1 This method applies to valves, particle filters, and
low pressure regulators.
NOTICE: This standard does not purport to address
safety issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish
appropriate safety and health practices and determine
the applicability of regulatory or other limitations prior
to use.
3 Limitations
3.1 This method is not intended to provide insight into
fundamental corrosion mechanisms.
3.2 Since the test is to be conducted on fully functional
gas system components, it is not designed to determine
the suitability of specific materials of construction for
corrosive gas service. Such factors as component
design, flow path, weld quality, cleanliness of
construction and manufacturing methods can impact the
results of this test making the material of construction
only one of many factors.
3.3 This method is not designed to provide reliability
or failure data. It is expected that the data produced
will be relative; that is, the data will be most useful
when groups of components are compared to each
other.
3.4 This method is not designed to evaluate parameters
besides particle contribution, for instance leak rate or
regulator drift.
3.5 Components to be tested must be capable of
withstanding the bake temperature of 100°C for filters
and 70°C for valves and regulators.
3.6 This test applies to HCl which is a reducing
environment. Results may not be indicative of
performance in corrosive oxidizing environments such
as Cl
2
.
4 Referenced Standards
4.1 SEMI Standards
SEMI C3.49 — Standard for Bulk Nitrogen (N
2
),
99.99999% Quality
SEMI E66 — Test Method for Determining Particle
Contribution by Mass Flow Controllers
4.2 ASTM Standard
1
ASTM F1394-92 — Standard Test Method for
Determination of Particle Contribution from Gas
Distribution System Valves.
4.3 SEMATECH Documents
2
SEMASPEC #93021510A-STD — Test Method for
Determination of Particle Contribution by Low Pressure
Regulators in Gas Distribution Systems.
SEMASPEC #93021511A — Test Method for
Determination of Particle Contribution by Filters in Gas
Distribution Systems.
NOTICE: Unless otherwise indicated, all documents
cited shall be the latest published versions.
5 Terminology
5.1 Definitions
5.1.1 cycle purge — a procedure which employs
alternating cycles of vacuum and pressure for the
purpose of evacuating a gas system or a portion of a gas
system of impurities or active gases.
5.1.2 particle filter — a device that removes small
particles from a gas stream with high efficiency.
5.1.3 point-of-use purifier — an inline device that
removes homogeneous impurities such as moisture and
oxygen gases. An integral particle filter is typically
included in the same housing to remove particles.
1 American National Standards Institute, 1819 L Street, NW, Suite
600, Washington, DC 20036, website: www.ansi.org
2 SEMATECH, 2706 Montopolis Drive, Austin, TX 78741, website:
www.sematech.org
SEMI F76-0303 © SEMI 2003 2
5.1.4 regulator — a valve designed to reduce a high
incoming pressure to a lower outlet pressure by
automatically opening to allow flow until a desired,
pre-set pressure on the outlet side is reached, and then
automatically throttling to stop further pressure
increase.
5.1.5 valve — a device that controls the flow or
pressure of a gas. Valve functions can include shut-off,
metering, backflow prevention, and pressure relief.
6 Summary of Method
6.1 This test method is designed to provide comparative
results for similar components. It is not intended to
replicate process condition or actual use conditions.
Components will be compared for their particle
shedding performance after being exposed to moisture
and corrosive gases, in this case hydrogen chloride
(HCl) gas. The component will be exposed alternately
to corrosive gas and then to moist 100 ppm nitrogen gas
several times. The particle shed performance before
and after exposure will be compared. The overall test
sequence is outlined in Figure 2.
7 Apparatus
7.1 Corrosive Gas Exposure Apparatus
7.1.1 The corrosive gas exposure will be carried out
using the apparatus shown in Figure 1.
7.1.2 Moisture Generator — Moisture generation is
achieved using a permeation-source-type moisture
generator. The permeation source should be selected to
provide 100 ppm H
2
O in a flow of 500 sccm nitrogen.
Records should be kept on the age of the permeation
source so that a reliable source of moisture can be
maintained.
7.1.3 Vacuum Generator — A Venturi or similar
device capable of producing a rough vacuum of 85000
pascals (25 inches of mercury).
7.1.4 Point of HCl Introduction — The point at which
gaseous hydrogen chloride HCl and N
2
join the system
(point A in Figure 1) should be designed so that the
dead volume between the check valves is minimized.
The check valves following V1 and V2 should be
nearly joined to each other so as to minimize this
volume.
N2
CORROSIVE GAS EXPOSURE
MOISTURE
GENERATOR
TEST PIECE
SCRUBBER
ANALYZER
PORT
V1
V2
V3
V4
V5
V6
V7
N2 purifier
HCl
PURIFIER
APPARATUS
HCl
vacuum
generator
A
Figure 1
Corrosive Gas Exposure Apparatus
Legend to figure 1
SYMBOL LEDGEND
Mas Flow Contoller
Impurity Removal Purifier
Particle Removal Filter
Pneumatically Actuated Valve
Check Valve
8 Reagents and Materials
8.1 Nitrogen — Nitrogen with less than 50 ppbv
moisture is to be used. This can be accomplished with
a point-of-use purifier as shown in Figure 1 or by using
99.99999% quality bulk nitrogen according to SEMI
C3.49.
8.2 Hydrogen Chloride — Gaseous HCl should contain
less than 1 ppmv H
2
O. The HCl purifier shown in
Figure 1 is optional unless the source gas contains
greater than 1 ppmv H
2
O.
SEMI F76-0303 © SEMI 2003 3
PARTICLE
CLEANLINESS TEST
REPEAT FOR A TOTAL
OF 8 TIMES
HOLD STATIC IN HCl
FOR 16 HOURS
RUN EXPOSURE
SEQUENCE ONCE
PER HOUR 6 TIMES
COOL OVERNIGHT
UNDER N2 PURGE
BAKE 2 HOURS
UNDER N2
INSTALL IN SYSTEM
START
FINAL DRY N2 PURGE
CYCLE PURGE 30
CYCLES
DRY N2 PURGE FOR
2 HOURS
EVALUATE PARTCLE
CLEANLINESS IN N2
REMOVE AND CAP
SAMPLE IN GLOVE
BAG
DONE
Figure 2
Test Flow
9 Safety Precautions
9.1 This method involves the use of hazardous
materials. The user should be experienced in the
handling and use of corrosive specialty gases.
9.2 Safety precautions should include, but are not
limited to:
continuous air monitoring for HCl in the working
environment as well as the ventilated enclosure.
SCBA or other supplied air respiratory protection
capable of use in gas exposure levels that may
exceed immediately dangerous to life and health
(IDLH) levels should be used during cylinder
installation and start-up.
Complete enclosure of the HCl exposure apparatus
in a forced-ventilation cabinet.
The user should consult the safety, health and
environmental personnel at their site for advice on
appropriate personal protective equipment,
monitoring and ventilation requirements.
10 Test Specimen
10.1 Starting Condition — The test specimen should
be a new component in its original packaging as
supplied by the manufacturer.
10.2 Fittings — The component must have either 1/4
face seal connections or be of a surface mount design
with all metal seals.
10.3 Sample Size — The number of samples to be
tested is to be determined by the user depending upon
their test objectives. It is beyond the scope of this
method to determine a statistically valid sample size,
however it is recommend that multiple samples of each
type are tested.
11 Preparation of Apparatus
11.1 Mass Spectrometer-based leak checking of the
HCl exposure system after assembly, after the
installation of each test component and after opening
any connection.
11.2 Before each experiment and during idle times, the
apparatus is to be kept in a constant purge condition.
The following valve positions would apply during idle
times:
VALVE POSITION
V1 CLOSED
V2 OPEN
V3 OPEN
V4 CLOSED
V5 OPEN
V6 CLOSED
V7 OPEN
11.3 Moisture Stabilization — At least four hours prior
to the start of a test, a constant flow of wet nitrogen
(100 ppm H
2
O) should be established through the
sampling port so as to maintain a steady moisture
output from the moisture generator. This is
accomplished by opening valve V7. During this time,
V4 is closed.
11.4 Moisture Level Verification — Once the moisture
level is stabilized in Section 11.2, the moisture value