semi合集-English.pdf - 第4624页

SEMI C3-0699 © SEMI 19 86, 1999 5 4.2.5.3 In sp ect the sampler vessel for a n y obvious physical defects, suc h as dents, gouges, b ent fittings, etc. Since the sampler is a pressurized v e ssel, it should be remov ed f…

100%1 / 7923
SEMI C3-0699 © SEMI 1986, 1999 4
3.2.2 By connecting the container being sampled
directly to the analytical equipment.
3.2.3 By selecting a representative c ylinder from the
cylinders in the lot.
3.3 Liquid Sampling (Vaporized)Vaporized liquid
samples shall be representative of the liquid supply.
Sampling shall be in accordance with one of the
following:
3.3.1 By vaporizing liquid from the supply container
in the sample tubing.
3.3.2 By flowing liquid from the sup ply container into,
or through, a suitable container in which a
representative sample is collected and then vaporized.
3.4 Liquid Samples (Liquified Compressed Gases)
A direct connection between the liquid phase of
liquified compressed gas containers and the analytical
equipment can be achieved, provided suitable flash
vaporization is obtained.
3.5 Lot Acceptance Tests — These are analyses
performed on the gas/liquid in the shipping container,
or a sample thereof, which is representative of the lot.
(The terms “lot” and “batch” may be used
interchangeably.)
3.6 Lots One of the following is to be used:
3.6.1 No specific quantity or any qu antity of product
agreed upon between the supplier and the customer.
3.6.2 All of the product supplied during the contract
period.
3.6.3 All of the product supplied or containers filled
during a calendar month.
3.6.4 All of the product supplied or containers filled
during seven consecutive days.
3.6.5 All of the product supplied or containers filled
during a consecutive 24-hour period.
3.6.6 All of the product supplied or containers filled
during one eight-hour shift.
3.6.7 All of the product supplied in one shipment.
3.6.8 All of the product supplied in one shipping
container.
3.6.9 All of the product supplied in the container(s)
filled on one manifold at the same time.
3.7 Number of Samples Per Lot The number of
samples per lot shall be in accordance with one of the
following:
3.7.1 One sample per lot.
3.7.2 Any number of samples agreed upon by the
supplier and the customer.
4 Sampling
4.1 For gases provided in cylinder s, a sample can be
taken directly for analysis. For gases provided in bulk
quantities or cylinders where direct sampling is not
appropriate, a sample may be taken per SEMI sampling
procedures.
4.2 Sampling Procedures
4.2.1 Cryogenic Liquid Sample — Liquid Samples for
Oxygen, Nitrogen, and Argon, using the TTU-131/E
sampler.
WARNING: DO NOT USE THIS PROCEDURE
FOR THE SAMPLING OF LIQUID HYDROGEN.
4.2.2 Applicable Document — Military Specification
MIL-S-27626D 16 August 1979, Amendment 1, 24
April 1981.
4.2.3 General Description — The TTU-131/E
cryogenic sampler is a small, portable pressure vessel
used to receive, vaporize, and contain a representative
sample of cryogenic liquid from a supply source. The
vaporized sample is withdrawn as a gas for analytical
purposes.
4.2.4 Theory — The sampler is used to isolate a small
but representative quantity of cryogenic liquid and
vaporize the major component and all volatile
impurities to form a homogeneous gas sample suitable
for analysis. The cryogenic liquid is used to cool a
shielding space and sampling cup prior to admitting the
liquid to the sampling cup. The purpose of pre-cooling
the sampling cup is to prevent concentrating impurities
which could result when the warm cup causes the liquid
to vaporize, leaving behind impurities with higher
boiling points.
4.2.4.1 When the cup is adequately cooled, the
sampling valve is opened, allowing liquid to fill the
cup. When the sampling valve is closed, the liquid is
trapped in the cup and will vaporize as a result of
atmospheric heating. Once the sample is trapped, the
flow of liquid is stopped.
4.2.5 Preparation for Sampling
4.2.5.1 The sample should be kept in the same product
service to avoid sample contamination. If a product
change is required, always thoroughly purge the
sampler or evacuate to 100 microns prior to taking the
sample.
4.2.5.2 When taking the sample, the s ampler is to
remain secured to the bottom half of the case and
MUST be in an upright position.
SEMI C3-0699 © SEMI 1986, 19995
4.2.5.3 Inspect the sampler vessel for any obvious
physical defects, such as dents, gouges, bent fittings,
etc. Since the sampler is a pressurized vessel, it should
be removed from service if any damage is apparent.
4.2.6 Sampling Procedure
4.2.6.1 Remove cover.
4.2.6.2 Loosen inlet and outlet fittings dust caps
located on side of the vessel.
CAUTION: SAFETY GLASSES AND
PROTECTIVE GLOVES ARE REQUIRED WHEN
OPERATING THIS EQUIPMENT. SAMPLING
SHOULD BE DONE ONLY IN WELL
VENTILATED AREAS.
4.2.6.3 Relieve pressure in vessel thro ugh vents in
dust caps by cautiously opening inlet sampling valve.
4.2.6.4 When gauge indicates atmosph eric pressure
and flow ceases, close inlet sampling valve and remove
dust caps.
4.2.6.5 Connect fill hose to supply tan k and inlet
fittings of sampler.
4.2.6.6 Open supply tank outlet valve. Gas and liquid
will begin to flow from the sampler outlet.
CAUTION: WHEN OXYGEN IS SAMPLED,
SOME ADDITIONAL PRECAUTIONS MUST BE
OBSERVED. IF THERE IS NO FACILITY SUCH
AS A CLEAN CONCRETE PAD ON WHICH THE
LIQUID OXYGEN CAN EVAPORATE SAFELY,
IT WILL BE NECESSARY TO CATCH THE
LIQUID OXYGEN IN A CLEAN, PREFERABLY
SEAMLESS ALUMINUM BUCKET. USE A LINE
WITH MINIMUM INSIDE DIAMETER OF ONE-
HALF INCH TO DIRECT THE LIQUID OXYGEN
INTO THE BUCKET. FURTHERMORE, IT IS
IMPORTANT THAT THERE IS NO
RESTRICTION IN THIS LINE BECAUSE THE
BACK PRESSURE DEVELOPED MAY BE
SUFFICIENT TO CAUSE THE LIQUID IN THE
CUP TO OVERFLOW INTO THE LARGE
CHAMBER. ON WARMUP, THE PRESSURE IN
THE SAMPLER COULD BECOME EXCESSIVE,
PARTICULARLY IF THE SAMPLE WERE
SUBCOOLED, NECESSITATING THE
REPLACEMENT OF THE SAFETY RELIEF
DEVICE.
4.2.6.7 Allow sampler to cool until a s teady flow of
liquid appears at outlet.
CAUTION: AVOID CONTACT WITH THE
FLOW OF CRYOGENIC LIQUID. THE
EXTREMELY LOW TEMPERATURE CAN
CAUSE PAINFUL INJURIES.
4.2.6.8 Open inlet sampling valve completely to allow
liquid to enter sampling cup.
4.2.6.9 After 30 seconds, close sampling valve.
4.2.6.10 Close supply tank outlet valve.
4.2.6.11 Disconnect fill hose.
4.2.6.12 Invert sampler for five minutes to allow
sampling cup to empty and provide vaporization of
liquid.
4.2.6.13 At ambient temperature the sa mpler should be
at 400-500 psig, indicating that a good sample was
obtained. A lower pressure would indicate a leak in the
sampler or that the cup was not filled with liquid. If
there is any doubt, release gas from sampler and take
another sample.
4.2.6.14 Re-install inlet and outlet fitting caps. Do not
tighten.
4.2.6.15 Affix a tag identifying the product to the inlet
sampling valve handwheel.
4.2.6.16 Install cover.
4.3 Gas Phase Sample — (To Be Determined)
4.4 Liquid Sample — (To Be Dete rmined)
4.5 Delivery to Analytical Instruments — (To Be
Determined)
5 Quantification (To Be Dete rmined)
5.1 Linear Response
5.1.1 Direct Comparison
5.1.2 Calibration Curve
5.2 Non-Linear Response
5.2.1 Direct Comparison
5.2.2 Calibration Curve
6 Gas Chromatography
6.1 The analysis of many gaseous impurities in bulk
and specialty gases is done by means of a gas
chromatography separation of impurities from the gas
matrix and quantification using a broad range of
sensitive and sometimes selective detectors. This
section will provide a guide for the format and
representation of such a procedure. Figure 1 is designed
to represent the format for submission of such
methodology.
6.2 Sample IntroductionBy its nature, gas
chromatography implies the use of carrier gases into
which the gaseous sample is injected and carried into
the separation medium, either solid or liquid in nature.
SEMI C3-0699 © SEMI 1986, 1999 6
The introduction of a gaseous sample into a carrier gas
stream is accomplished by means of a sample injection
valve. Valves for this purpose can be any of several
types, although the most commonly used are either
rotary or diaphragm valves using loops of stainless steel
tubing of a known volume. It is not necessary to
diagram such sample introduction systems for a
procedure unless the introduction system is unique to
the application by way of special conditions for safety
or other reasons critical to the accuracy of the sampling
technique.
6.3 SeparationsIt is an acceptable and common
practice to quantitate several gaseous impurities with a
single gas chromatographic separation. Occasionally a
single injected sample volume will undergo
"multidimensional" separation to achieve the necessary
analysis. Rotary valves, diaphragm valves or Dean's
pressure switching again are used to move impurities
from one separation medium to another, or from one
column to another. Where multidimensional techniques
are used, they should be accompanied by flowpath
diagrams which clearly indicate valving and their
appropriate positions. The valving sequences used
should have adequate explanation to clarify the
separations.
6.4 Columns — Columns used are generally of 3.2
mm (1/8 in.) o.d. and 2.2 mm (0.085 in.) i.d. stainless
steel of varying lengths and packed with one of many
solid supports. Columns may be specified by length and
packing material if these standard dimensions apply. If
another dimension or material is used for the column
itself, it should be specified by material type, and by
o.d. and i.d. in mm. Packing materials should be
specified by material, % coating and coating type if
applicable, and mesh size. Capillary columns are
assumed to be fused silica unless otherwise specified
and should be specified by i.d. in mm, film thickness in
µm, and length in meters. Column temperatures and
applicable program rates should be given for each
independently heated zone in degrees Celsius.
6.5 Carrier and Support GasesCarrier and support
gas flow rates for all separations should be specified.
Carrier and support gases used should fall into the
general purity requirements in Table 2 and purity need
not be specified unless the analysis has specific purity
requirements for safety, accuracy, or component
lifetime. Flows should be specified in mL/min or
L/min, or as linear velocity for open tubular capillary
columns.
6.6 Detectors — Detector technolo gy should be
specified, and alternative detection may be assumed
equivalent for detectors of like selectivity and
sensitivity. Table 2 gives a reference of relative
detector sensitivity and is provided as a guideline only.
Detector parameters are less likely to be easily
formatted since detector technologies vary widely in
their specific parameters. All elements critical to
achieving like sensitivity must be included, such as
temperature, make-up gases and their flows, flame,
combustion or reaction gas ratios or flows, voltages,
currents, or any additional settings as outlined by the
detector manufacturer.
6.6.1 Detection limits must be speci fied for each
impurity for the actual methodology used. Because the
lower detection limit can be greatly influenced by
retention time for any impurity, it is important that the
detection limit be derived for the impurity within the
context of the method, and not solely on the
detectability of the detector. Detection limits are
assumed to be calculated on a mole/mole basis.
6.7 Operating ProceduresThe standard procedure
assumed for the gas chromatographic analysis of most
gases is as follows:
1. Inject the calibration standard onto the column
using a gas sampling valve. Record the retention
time(s) and peak area(s) for all impurities detected.
2. Analyze the sample to be tested in the same
manner as the calibration standard.
3. Repeat 1.
4. Compare the average peak area of the calibration
standard with that of the sample being tested. (See
Calculation of Concentration, in Section 6.9.)
6.7.1 The method should specify an y additional valve
switching or special parameters which are necessary for
the successful analysis of the impurities specified for
the procedure. It must also specify the order of elution
if multiple impurities are detected.
6.8 Calibration Calibration for these analytical
procedures is generally by external, single point
calibration. Calibration gas mixtures should be
specified and are assumed to be within the range
specified in Section 2.26. Calibration standards are
assumed to be made in a balance of gas representative
of the sample unless otherwise specified. If calibration
techniques other than external, point calibrations are
used, they must be described.
6.9 Calculation of Concentration — The calculation of
the concentration of impurity within the sample is based
on a comparison of the average peak area of the
impurity in the calibration standard, to the average peak
area of the impurity within the sample, based on the
formula:
Sample Peak Are
a
Standard Peak Are
a
×
Concentratio
n
of Standard
=
Concentratio
n
of Sample