semi合集-English.pdf - 第4667页
SEMI C3.32-03 01 © SEMI 1987 , 2001 5 Figure 2 Oxygen and Nitrogen Determination Figure 3 Metals Set up (Liquid Phase Analy sis)

SEMI C3.32-0301 © SEMI 1987, 2001 4
4.4.3 Calibration Standards — Construct a calibration
curve which contains at least two points covering the
range of interest.
4.4.4
Operating Procedure
4.4.4.1 Leak check the entire system using helium and
a helium leak detector prior to use.
4.4.4.2 Pre-purge through the regulator/cross-purge
assembly with a dry gas of known moisture content
until the moisture content is measured below 1 ppm
(vol/vol).
4.4.4.3
Direct samples of two standards, independently
verified with a dewpoint/frostpoint hygrometer,
spanning the range of moisture content of interest to the
analyzer. After a stable reading is obtained, record the
instrument's response to each standard.
4.4.4.4
Obtain a representative sample of the gas to be
analyzed and direct it to the unit as with the standards.
After a stable reading is obtained, record the
instrument’s response.
4.4.4.5
Construct a calibration curve from the standard
data and determine the moisture content in the sample
gas. The result may not exceed the specification in
Section 2 of this standard.
4.5
Notes
NOTE 1: Sample system should be constructed entirely of
stainless steel or other suitable material due to the corrosive
nature of the product.
NOTE 2: Sample must be appropriately vented for health and
safety of operating personnel.
NOTE 3: Individual standards are required for accuracy and
identification by retention time.
NOTE 4: Other accepted instrumental techniques are also
appropriate for this analysis. In addition, operating procedures
for ICP-MS instruments vary, depending on the manufacturer.
In all cases, refer to individual instrument vendor instructions.
NOTE 5: All sampling flasks are leached with a 25%
HNO
3
/H
2
O solution for at least 24 hours, rinsed thoroughly
with DI H
2
O, and filled with DI H
2
O for at least 24 hours
prior to use. On the day of sampling, flasks are rinsed
thoroughly with DI H
2
O and dried with N
2
prior to system
setup.
NOTE 6: The tubing and Fluorocarbon valves are soaked in a
25% HNO
3
/DI H
2
O solution prior to use (either apart or pre-
assembled). The tubing and valves are rinsed thoroughly with
DI H
2
O and dried with high-purity N
2
prior to system setup.
NOTE 7: A different digestion solution may be used,
depending on possible interferences with a given analysis
technique.
NOTE 8: All wetted surfaces, including P
2
O
5
hygrometer and
sample system, are to be of 316 stainless steel or other
suitable material.
Figure 1
Carbon Monoxide, Carbon Dioxide, and Hydrocarbons Determination

SEMI C3.32-0301 © SEMI 1987, 20015
Figure 2
Oxygen and Nitrogen Determination
Figure 3
Metals Setup (Liquid Phase Analysis)

SEMI C3.32-0301 © SEMI 1987, 2001 6
Figure 4
NOTICE: This specification does not purport to address safety issues, if any, associated with its use. It is the
responsibility of the user of this specification to establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use. SEMI makes no warranties or representations as to the suitability
of the specification set forth herein for any particular application. The determination of the suitability of the
specification is solely the responsibility of the user. Users are cautioned to refer to manufacturer’s instructions,
product labels, product data sheets, and other relevant literature respecting any materials or equipment mentioned
herein. This specification are subject to change without notice.
The user’s attention is called to the possibility that compliance with this specification may require use of
copyrighted material or of an invention covered by patent rights. By publication of this specification, SEMI takes no
position respecting the validity of any patent rights or copyrights asserted in connection with any item mentioned in
this specification. Users of this specification are expressly advised that determination of any such patent rights or
copyrights, and the risk of infringement of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.