semi合集-English.pdf - 第4699页
SEMI C3.39-0304 © SEMI 1989, 2004 5 Figure 1 Configuration for the Analysis of CF 4 , CO 2 , N 2 O, SF 6 , and CO in NF 3 Figure 2 Configuration for the An alysis of O 2 /Ar and N 2 in NF 3

SEMI C3.39-0304 © SEMI 1989, 2004 4
Table 1 Calibration Concentrations
Added Volume of
Working Standard
(ml)
Total Volume
(ml)
Resulting F
-
Concentration (M)
0.2 100.2 1.0 × 10
-6
0.2 100.4 2.0 × 10
-6
0.4 100.8 4.0 × 10
-6
0.4 101.2 5.9 × 10
-6
0.8 102.0 9.8 × 10
-6
1.0 103.0 1.5 × 10
-5
2.0 105.0 2.4 × 10
-5
5.6.5.5 Put 50 ml 0.2 N NaOH into each of two
bubblers connected in series.
5.6.5.6 Establish a flow of < 1000 sccm of NF
3
through
the bubblers using a suitable flow controller or
flowmeter.
5.6.5.7 Sample approximately 15 liters of NF
3
. Record
flowrate and time of sampling to determine total
volume sampled (flowrate × time). A wet test meter
can also be used to measure total volume. The amount
of gas sample must be the volume at STP. If the
flowmeter or wet test meter is not reference to 0º and
760 torr, use the formula below to correct sample
volume.
K 273
273T
P
Torr 760
Liters Measured STPat Liters
+
××=
P: Pressure of sampled gas (mm Hg)
T: Temperature of sampled gas or reference
temperature of the flow controller or meter in ºC.
5.6.5.8 Transfer contents of each bubbler to individual
100 ml volumetric flasks and add 50 ml Buffer solution
to each. Then, if necessary, add deionized or distilled
water to bring the volume up to 100 ml.
5.6.5.9 Transfer contents to a plastic beaker.
5.6.5.10 While stirring, measure and record mV
readings for each sample.
5.6.5.11 Determine F
-
concentration in solution using
calibration curve generated in Section 5.6.5.4.
5.6.5.12 Calculate gas phase hydrolyzable fluoride
concentration using the equation below. Note that the
equation assumes the hydrolyzable fluoride is hydrogen
fluoride.
6
v
10
Vs
l/mole 22.4
0.11 C )(ppm HF Phase Gas ×××=
C: Measured hydrolyzable fluoride concentration (M)
determined in Section 5.6.5.11.
Vs: Volume of NF
3
sampled (liters at STP)
5.6.5.13 The HF concentration of the second bubbler
should be insignificant compared to the first bubbler. If
significant HF levels are found in the second bubbler,
resample with a lower flowrate.
5.7 Notes
NOTE 1: Pass the carrier gas through a trap (3 m (10 ft.) by
6.4 mm (1/4 in.) OD) containing 50% by volume molecular
sieve 5A and 50% by volume molecular sieve 13X submerged
in liquid nitrogen.
NOTE 2: The methanizer catalyst can be destroyed by
nitrogen trifluoride passing through to at high temperature.
After the carbon monoxide peak is detected, the bulk nitrogen
trifluoride is bypassed to vent by means of a four-port
switching valve installed downstream of the separating
column and before the methanizer (see Figure 1).
NOTE 3: The sampling system and hygrometer must be
designed to operate under the sample pressure, or the sample
pressure must be reduced, by a regulator with a diaphragm of
stainless steel or other suitable material, to accommodate the
pressure restrictions of the analytical hygrometer.
NOTE 4: A passivation procedure is described in the Metals
Handbook, 8th ed., Vol. 2, American Society for Metals,
Metals Park, OH.
NOTE 5: Polytetrafluoroethlyene, polymethylpentene and
polypropylene are suitable plastic materials.

SEMI C3.39-0304 © SEMI 1989, 2004 5
Figure 1
Configuration for the Analysis of CF
4
, CO
2
, N
2
O, SF
6
, and CO in NF
3
Figure 2
Configuration for the Analysis of O
2
/Ar and N
2
in NF
3

SEMI C3.39-0304 © SEMI 1989, 2004 6
NOTICE: This standard does not purport to address safety issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish appropriate safety and health practices and determine the
applicability of regulatory or other limitations prior to use.
SEMI makes no warranties or representations as to the suitability of the standards set forth herein for any particular
application. The determination of the suitability of the standard is solely the responsibility of the user. Users are
cautioned to refer to manufacturer’s instructions, product labels, product data sheets, and other relevant literature
respecting any materials mentioned herein. These standards are subject to change without notice.
The user’s attention is called to the possibility that compliance with this standard may require use of copyrighted
material or of an invention covered by patent rights. By publication of this standard, SEMI takes no position
respecting the validity of any patent rights or copyrights asserted in connection with any item mentioned in this
standard. Users of this standard are expressly advised that determination of any such patent rights or copyrights, and
the risk of infringement of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.