SEMI C1-0705 © SEMI 1978, 2005 17
7.6.5.3 Place the generator flask in a water bath maintained at 25° ± 3°C and swirl the flask gently at 10 minute
intervals. (The addition of a small volume of 2-propanol to the generator flask may make the gas evolution more
uniform.) After 45 minutes, disconnect the tubing from the generator flask, transfer the silver diethyldithiocarbamate
solution to a 1 cm photometer cell, and measure the absorbance at 525 nm versus the water blank.
7.6.5.4 Carry a water blank and standard through the procedure, using the identical quantities of the same reagents.
The standard consists of a volume of the arsenic working solution corresponding to the amount of arsenic stated in
the procedure for the individual liquid chemicals. Any red color in the silver diethyldithiocarbamate solution from
the sample preparation should be no greater than that obtained with the standard.
7.7 Use of a pH Meter — The principles of pH measurement are given in documents and instrument manuals. The
purpose here is to describe in general terms how the technique is used with the procedures for individual liquid
chemicals, either to assure that a “pH of Solution” specification is met or to adjust the acidity or alkalinity of a
solution as required in the conduct of a procedure.
7.7.1 Equipment — The pH meter should be capable of reading to 0.02 pH units and should best be equipped for
temperature compensation. The electrodes should be calomel and glass and should be kept in appropriate solutions
when not in use, following the supplier’s recommendations.
7.7.2 Standardization — At time of use, the pH meter and electrodes should be standardized against two pH
standard buffers that preferably should bracket the expected pH of the solution to be measured. The meter with the
electrodes immersed in the first buffer should be adjusted to the stated value for that buffer. The electrodes should
then be rinsed, and the meter, with the electrodes immersed in the second buffer, should be adjusted to the stated
value for that buffer. This adjustment procedure should be repeated until two consecutive measurements for one of
the buffers gives the same value within ± 0.02 pH units. Where a temperature is specified for an individual liquid
chemical, either as part of the specification parameter or the relevant procedure, that temperature should be
maintained in the standardization with the corresponding assigned pH values for the pH standard buffers being
employed. During measurements, each buffer should be magnetically stirred at the same constant rate.
7.7.3 Quantitation — For the measurement of the pH of a solution, the meter and electrodes should be standardized
as given above. The electrodes should then be rinsed with water and immersed in the solution (held at the specified
temperature), and the pH read once the reading has stabilized. The solution should be magnetically stirred at the
same rate as used in the standardization. If the reading does not stabilize within 1 minute, the cause should be found
and eliminated. Any water used to dilute a sample for the “pH of Solution” test should be carbon dioxide-free.
7.8 Determination of Various Ions by Ion Chromatography — The theory of ion chromatography is given in many
documents, reviews, etc. The purpose here is to delineate some practical aspects for the determination of various
ions by Ion Chromatography. A complete detailed procedure for the ion chromatographic determination is not given
since available instruments vary from laboratory to laboratory.
7.8.1 Equipment — An ion chromatograph equipped with ion suppression is recommended. Single column ion
chromatography can also be used. The choice of reagents and columns is determined by the ions of interest.
7.8.2 Sample — Sample preparation varies depending upon the matrix effects of the liquid chemicals being
analyzed. Standard sample preparation techniques that may be used are (this list is not all inclusive):
7.8.2.1 Evaporation of the reagent to remove the matrix and then diluting the residual with the appropriate grade of
18.2 Mohm-cm water that meets ASTM Standard Guide D5127.
7.8.2.2 Diluting the reagent with the appropriate grade of 18.2 Mohm-cm water that meets ASTM Standard Guide
D5127 to lessen the matrix effects.
7.8.2.3 Direct injection of the reagent.
7.8.3
Columns — Many types of columns exist and can be used. Typical selection criteria for columns consist of
ion or ions of interest, the matrix of the reagent and eluent to be used.
7.8.4 Conditions — Appropriate conditions will vary from one ion chromatographic system to another. Analytical
conditions using ion suppression, which have been found acceptable using ion suppression, are listed under each
individual standard. The two broad classifications of ions are anions and cations. A typical eluent used in the
analysis of anions by ion suppression is a sodium carbonate and sodium bicarbonate in a solution with the
appropriate grade of 18.2 Mohm-cm water that meets ASTM Standard Guide D5127. A typical eluent used in the
SEMI C1-0705 © SEMI 1978, 2005 18
analysis of cations is a dilute solution of hydrochloric acid in a solution with the appropriate grade of 18.2 Mohm-
cm water that meets ASTM Standard Guide D5127.
7.8.5 Eluent concentration, eluent flow rate, and system pressure are variables, which affect the chromatography.
Each must be determined for the chromatographic system being used.
7.8.6 Detector — A conductivity detector is used to monitor the changes in the eluent conductivity due to ions in
the sample. An ultraviolet absorbance detector can be used for detection of some ions. The detector output, after
amplification, is used to produce the chromatogram, a plot of component quantity versus elution time. Modern
systems digitize the analog output allowing direct printout of peak heights, peak areas, and retention times.
7.8.7 Calculation of Results
Ion Concentration =
Pk. Ht. S
I
= Peak Height of Sample Ion
Pk. Ht. B
I
= Peak Height of Blank Ion
Pk. Ht. Std
I
= Peak Height of Standard Ion
Conc. of Std
I
= Concentration of Standard Ion
Peak area may be used instead of peak height. Calibration linearity must be established before use.
7.9 Determination of Trace Elements by Atomic Absorption Spectrometry (AAS) — The purpose here is to describe
in general terms how the technique should be used when specified in the procedures for an individual liquid
chemical to assure that the specifications for one or more stated elements are met. No general procedure is given
since operating details will vary with instrument design; consequently, the manufacturer’s manual should be
followed.
7.9.1 Equipment — The source is a hollow cathode lamp, or equivalent, for the relevant element. The burner is
slotted and will vary in design and path length according to the gas(es) to be used. The monochromator should have
a resolution of at least 0.1 nm. The signal from the detector is read from a meter, digital display, recorder tracing or
custom software. A background correction system should be available.
7.9.2 Flame and Burner Conditions — The flame gases to be used are specified in the procedure for the individual
liquid chemical. The burner position and gas flow rates should be optimized for a given determination.
7.9.3 Solution Conditions — The solutions are prepared as given in the procedures for the individual liquid
chemical.
7.9.4 Working Standard Solutions — Working standard solutions should be prepared on the day of use by dilution
of a stock standard solution.
7.9.5 Quantitation — Commonly, the method of standard addition is employed. It is applicable at low
concentrations where a linear relationship exists between signal and concentration. By this technique the standards
have a matrix similar to the sample. It is essential that the sample solution, along with the additions, has absorbance
in the range for which linearity is expected: routinely, 0.1 to 0.5 absorbance unit. The use of standard additions does
not eliminate interferences, but assures that the element of interest behaves similarly in the sample preparation and
those with additions. Background correction is usually required, and additives may be called for (see below).
7.9.5.1 For specific tests, the signals are measured for five solution preparations: the reagent blank (which may be
only water), the sample solution, and three made by standard additions to the sample solution. The signal for the
reagent blank preparation is subtracted from the signals for each of the four other preparations. These four signals
are used graphically or mathematically to determine the content of the element in the sample. Where several
elements are to be determined, they are added simultaneously to the third, fourth, and fifth preparations.
Commonly, the additions correspond to one half the specification limit, equal to the limit, and twice the limit.
7.9.5.2 A calibration line obtained for aqueous standards can be used if its slope and the slope of the standard
addition line are identical. In that event, liquid chemical (matrix) effects are negligible.
(Pk. Ht. S
I
Pk. Ht. B
I
) (Conc. of Std
I
)
Pk. Ht. Std
I
Pk. Ht. B
I
Dilution Factor
SEMI C1-0705 © SEMI 1978, 2005 19
7.9.6 Background Correction — Where spurious absorption or other background correction is required, details are
given in the procedure for an individual liquid chemical. For an element with the stated resonance line in the
ultraviolet region (<350 nm), the correction should be performed with a deuterium discharge lamp. For an element
with the stated line in the visible region (>350 nm), the adjacent line technique should be employed. In this method,
the background absorption is determined at a wavelength adjacent, ±5 nm, to the line being used for the element of
interest; that is, it is measured at a wavelength where significant atomic absorption by that element is absent.
Zeeman Background correction may also be used.
7.9.7 Additives — Where additives are included in the solution preparations, including an ionization suppressant,
details are provided in the procedures for the individual liquid chemical.
7.9.8 Remarks — The manufacturer’s manual should be consulted for details on the venting of burner gases and for
safety precautions in the operation of the instrument and burner. The user should check the performance of an
instrument from time to time to assure that adequate sensitivity is being attained for the needs of the particular
determination.
7.10 Determination of Trace Elements by Flame Emission Spectrometry — The purpose here is to describe in
general terms how the technique should be used when specified in the procedures for an individual liquid chemical
to assure that the specifications are met for one or more stated elements (commonly sodium or potassium, or both).
No general procedure is given since operating details will vary with instrumental design; consequently, the
manufacturer’s manual should be followed.
7.10.1 Equipment — The flame photometer should have a monochromator with a resolution of at least 1 nm. The
detector should have a red-sensitive phototube, a filter to remove the sodium D-line for potassium determinations,
and a photomultiplier. (Instruments having a photomultiplier specially sensitive to the potassium line do not need a
red-sensitive phototube.) The slit width must be adjustable. A sensitivity control should be available. Also, the
burner should have controls for adjusting its position and gas flow rates. The read-out device may be a meter,
digital display, or recorder.
7.10.2 Flame and Burner Conditions — The flame gases to be used are specified in the procedure for the individual
liquid chemical. The burner position and gas flow rates should be optimized for a given determination.
7.10.3 Solution Conditions — The solutions are prepared as given in the procedures for the individual liquid
chemical.
7.10.4 Standards — A stock standard solution for each relevant element should be accurately prepared, usually at a
concentration of 0.1% wt/vol, as given either in the procedure for the individual liquid chemical or in §4. Working
standard solutions should be prepared on the day of use by dilution of a stock standard solution.
7.10.5 Quantitation — Commonly, a single addition of standard is employed. It is applicable at low concentration
where a linear relationship exists between signal and concentration. By this technique, the standard has a matrix
similar to the sample.
7.10.5.1 The amount of standard to be added to the sample to make the standard solution is given in the procedures
for the individual liquid chemical, and corresponds to the specification value of the element being determined.
7.10.5.2 For specific tests, the signals are measured for the sample solution and the standard solution. The
instrument is optimized with the standard solution at the wavelength specified. It is then zeroed with water, and the
sample is aspirated into the flame and its signal read. Without changing the slit, the wavelength is changed to a
nearby setting and the background value determined. The difference between the background value and the value
for the sample should not exceed the difference between the values obtained for the sample and standard solutions.
7.10.6 Remarks — The manufacturer’s manual should be consulted for details on the venting of burner gases and
for safety precautions in the operation of the instrument and burner. The user should check the performance of the
instrument from time to time to assure that adequate sensitivity is being attained for the needs of the particular
determination.
7.11 Determination of Trace Elements by Graphite Furnace Atomic Absorption Spectrometry — The purpose here
is to describe in general terms how the technique should be used when specified in the procedures for an individual
liquid chemical to assure that the specifications for one or more stated elements are met. No general procedure is
given since operating details will vary with instrument design; consequently, the manufacturer’s manual should be
followed.