SEMI S2-0703a E © SEMI 1991, 2004 36 there is a failure, and then sligh tly increasing the flow rate until the test is successful, can be used to help minimize air flow specifications. A2-2.4.4 Chem ical air or wipe m on…

SEMI S2-0703a
E
© SEMI 1991, 2004 35
Hood Type Recommended Test Methods Typical Design and Test Exhaust
Parameters (See NOTE 1.)
References
Parts-Cleaning
Hoods
Primary: face velocity, vapor
visualization
Supplemental: tracer gas, air sampling
0.40–0.64 m/s (80–125 fpm) face
velocity
ASHRAE Standard 110
SEMI F15
ACGIH Industrial
Ventilation Manual
Pump and
Equipment
Exhaust Lines
Primary: static pressure
Supplemental: tracer gas
–6 to –25 mm (–0.25 to –1.0 in.)
H
2
O static pressure
125% maximum volume flow rate
from pump
ACGIH Industrial
Ventilation Manual
SEMI F15
Glove Boxes Primary: static pressure, tracer gas
Supplemental: vapor visualization, air
monitoring
No consensus for a reference at the
time of publication of this guideline.
ACGIH Industrial
Ventilation Manual
SEMI F15
Drying/ Bake/
Test Chamber
Ovens
Primary: static pressure, tracer gas
Supplemental: vapor visualization, air
monitoring
–1.3 to –2.5 mm (–0.05 to –0.1 in.)
H
2
O static pressure
SEMI F15
ACGIH Industrial
Ventilation Manual
Spin-Coater
(cup only)
Primary: vapor visualization,
velometry
Supplemental: air sampling
(see SEMI S2 Sections 23.5.1–3) ACGIH Industrial
Ventilation Manual
Supplemental
Exhaust
Primary: capture velocity, vapor
visualization, air sampling
0.50–0.76 m/s (100–150 fpm)
capture velocity
ACGIH Industrial
Ventilation Manual
NOTE 1: All measurements should be within ± 20% of average for face velocity, ± 10% of average along the length of each slot for slot velocity,
and ± 10% of average between slots for slot velocity.
A2-2 Exhaust Optimization
A2-2.1 Exhaust optimization is the use of good
ventilation design to create efficient equipment exhaust.
The design and measurement methods discussed below
confirm that equipment exhaust is acting as the
manufacturer intended. This information is not meant to
prohibit alternate methods of achieving or verifying
good ventilation design. References for ventilation
design are included at the end of this Appendix.
A2-2.2 Design Recommendations
A2-2.2.1 Equipment exhaust design can attempt to
reduce inefficient static pressure losses caused by:
friction losses from materials; openings, and duct
geometries (elbows, duct expansions or contractions);
turbulent air flow; fans; internal fittings such as blast
gates and dampers; directional changes in airflow.
A2-2.2.2 Other good design principles can include
minimizing distance between the source and hood, and
reducing enclosure volumes.
A2-2.2.3 For non-chemical issues such as heat from
electrical equipment, heat recapture rather than exhaust
may be appropriate.
A2-2.2.4 The possible impact of highly directional
laminar airflow found in most fabs should be
considered when designing equipment exhaust.
A2-2.3 Recommended Equipment Controls — The
location of internal blast gates or dampers inside
equipment, and their appropriate settings, should be
clearly identified. The number of equipment dampers
and blast gates should be minimized. Gates/dampers
should be lockable or otherwise securable. Static
pressure or flow sensors installed on equipment by the
manufacturer should have sufficient sensitivity and
accuracy to measure exhaust flowrate fluctuations that
place the equipment out of prescribed ranges.
A2-2.4 Recommended Measurement/Validation
Method — Measurements should be made to identify
optimal exhaust levels and confirm that safety and
process requirements are being addressed. The
manufacturer should be able to identify any critical
equipment locations for chemical capture, and quantify
appropriate exhaust values. Multiple validation/
measurement methods may be needed.
A2-2.4.1 Measurements should be done after
equipment components are assembled.
A2-2.4.2 Computer modeling can be done to predict
exhaust flow and hazardous material transport in
equipment by solving fluid mechanics conservation of
energy and mass equations. Modeling can be used in
the equipment design stage or to improve existing
equipment. Computer models should be verified
experimentally, using one or more of the methods
discussed below.
A2-2.4.3 Tracer gas testing provides a method to test
the integrity of hoods by simulating gas emission and
measuring the effectiveness of controls. Testing until

SEMI S2-0703a
E
© SEMI 1991, 2004 36
there is a failure, and then slightly increasing the flow
rate until the test is successful, can be used to help
minimize air flow specifications.
A2-2.4.4 Chemical air or wipe monitoring can be used
to confirm that chemical transport is not occurring into
unintended areas of the equipment.
A2-2.4.5 Velocity profiling will confirm expected
airflows, the direction of flow, and the effect of
distance.
A2-2.4.6 Vapor visualization will confirm expected
airflows, the direction of flow, and the effect of
distance. Vapor visualization is the observation of
aerosols (e.g., aerosols generated by using water, liquid
nitrogen, or dry ice) so that exhaust flow patterns can
be observed. Smoke tubes or aerosols may also be used,
however they can produce contamination.
A2-3 Chemical Laboratory Fume Hoods, Parts
Cleaning Hoods
Lab fume hoods and part cleaning hoods are designed
to control emission by enclosing a process on five sides
and containing the emission within the hood.
A2-3.1 Design Recommendations
A2-3.1.1 Fully enclosed on five sides, open on one side
for employee access and process/parts placement and
removals.
A2-3.1.2 Front (employee access side) should be
provided with sliding door and/or sash.
A2-3.1.3 Minimize size of the hood based on process
size.
A2-3.1.4 Minimize front opening size based on size of
process and employee access needs.
A2-3.1.5 Ensure hood construction materials are
compatible with chemicals used.
A2-3.2 Control Specifications — Face velocity is the
specification generally used with hoods open on only
one side.
A2-3.2.1 Generally acceptable laboratory fume hood
face velocities range from 0.40 to 0.60 m/s (80–120
fpm) with no single measurement 20 % of average.
0.64 to 0.76 m/s (125–150 fpm) is recommended for
hoods in which carcinogens or reproductive toxicants
may be used.
A2-3.2.2 Velocities as low as 0.30 to 0.40 m/s (60–80
fpm) can be effective but require no cross drafts or
competing air movement in the work area.
A2-3.2.3 An average face velocity of 0.50 m/s (100
fpm) is generally found to be acceptable in most
applications.
A2-3.2.4 Face velocities of 0.64 to 0.76 m/s (125 to
150 fpm) may be required when a lab hood is installed
in an area with laminar air flow.
A2-3.2.5 Face velocity above 0.76 m/s (150 fpm)
should be avoided to prevent eddying caused by a lower
pressure area in front of an employee standing at the
hood.
A2-3.3 Recommended Measurement/Validation
Method
A2-3.3.1 The preferred method is measurement of
average face velocity and hood static pressure.
Measurements are taken with a velometer or
anemometer. Multiple measurements are taken in a
grid, at least 10 to 40 per square meter (1–4 per square
foot) of open area, in the plane opening of the hood.
This allows representative, evenly spaced
measurements to be taken (see also open-surface tanks).
A2-3.3.2 Additional confirmation by visualization
check of containment using smoke or vapor testing.
A2-3.3.3 ASHRAE Method 110, or equivalent (use
appropriate sections), for tracer gas testing of lab hoods
may be used as a supplemental verification provided
that an accurate emission rate can be defined.
(ASHRAE 110 lists 3 tests: “as manufactured,” “as
used,” and “as installed.” The “as manufactured” test is
the test that is used most frequently.)
A2-4 Wet Stations
Wet stations are slotted hoods designed to capture
laminar air flow while also capturing wet process
emissions from the work area. Wet stations can be open
on the front, top and both sides (it is usually preferable
to enclose as much as possible).
A2-4.1
Design Recommendations
A2-4.1.1 Slots should be provided uniformly along the
length of the hood for even distribution of airflow.
A2-4.1.2 Additional lip exhaust slots should be
provided around tanks or sinks to control emissions.
A2-4.1.3 The plenum behind the slots should be sized
to ensure even distribution of static pressure. These
slots should be designed to ensure adequate airflow is
provided by the side slots, and to minimize turbulence
that could reduce exhaust performance.
A2-4.1.4 Velocity along length of slot should not vary
by more than 10% of the average slot velocity.
A2-4.1.5 Additional use of end or side panels/baffles
can reduce negative impact of side drafts.

SEMI S2-0703a
E
© SEMI 1991, 2004 37
A2-4.1.6 Exhaust volume settings should consider
laminar air flow volumes and be balanced to minimize
turbulence and to ensure capture.
A2-4.1.7 The station design should consider airflow
patterns in the operating zone to minimize turbulent
horizontal airflow patterns into and across the work
deck.
A2-4.1.8 Additional considerations to reduce exhaust
demand include providing covered tanks, and recessing
tanks below deck level.
A2-4.2 Control Specifications
A2-4.2.1 Wet station specifications are complicated by
the fact that wet stations generally do not have an easily
definable face velocity to measure. A number of
methods have been used and are all acceptable if used
consistently and provided documentation indicates
chemical containment meets the 1% of the OEL at
distances beyond the plane of penetration at the exterior
of the wet station.
A2-4.2.2 Maintain an average capture velocity of 0.33
to 0.50 m/s (65–100 fpm) immediately above a bath.
A2-4.2.3 Calculate the total exhaust volume
requirement by determining the total volumetric flow of
laminar air hitting the deck and increasing this value by
20 to 25%.
A2-4.2.4 For some wet stations that are partially
enclosed from the top, an artificial plane opening
(“face”) can be defined where the downward laminar
air flow penetrates the capture zone (at “face velocity”)
of the wet station. Depending on the hood design and
laminar air flow provided, average face velocities can
range from 0.20 to 0.50 m/s (40 to 100 fpm). The
measurement location can greatly influence the
measured face velocity; therefore, this method should
be supplemented with at least one of the preceding
methods for greater accuracy and reproducibility at the
user’s facility.
A2-4.3 Recommended Measurement/Validation
Method
A2-4.3.1 Confirmation of capture using vapor
visualization.
A2-4.3.2 Confirmation of laminar flow of make up air
into the station using vapor visualization.
A2-4.3.3 Tracer gas testing may be used as
supplemental verification, provided an emission rate
can be accurately defined.
A2-5 Supplemental Exhaust
Supplemental exhaust, if not designed into the
equipment, can be provided by a flexible duct with a
tapered hood. This can be placed in the work area to
remove potential contaminants before they enter the
breathing zone. Supplemental exhaust is frequently
used during maintenance or service.
A2-5.1 Design Recommendations
A2-5.1.1 Retractable or movable non-combustible flex
ducting for easy reach and placement within 150 to 300
mm (6 to 12 inches) of potential emissions to be
controlled.
A2-5.1.2 Manual damper at hood to allow for local
control, i.e., shut off when not required.
A2-5.1.3 Tapered hood with a plane opening as a
minimum. The additional use of flanges or canopies to
enclose the process will result in improved efficiency.
A2-5.2 Control Specifications
NOTE A2-1: This is one equation that is most commonly
used. Other equations may be appropriate; see also ACGIH
Industrial Ventilation Manual, and Semiconductor Exhaust
Ventilation Guidebook.
A2-5.2.1 A minimum capture velocity of 0.50 m/s (100
fpm) is required at the contaminant generation point for
releases of vapor via evaporation or passive diffusion.
Ventilation should not be relied upon to prevent
exposures to hazardous substances with release
velocities (e.g., pressurized gases). For a plane open
ended duct without a flange, the air flow required at a
given capture velocity can be calculated by:
Q = V(10X
2
+ A)
Where: Q = required exhaust air flow in m
3
/s (cfm)
V = capture velocity in m/s (fpm) at distance X from
hood
A = hood face area in square meters (square feet)
X = distance from hood face to farthest point of
contaminant release in meters (feet). NOTE: This is
only accurate when X is within 1.5 diameters of a round
opening, or within 0.25 circumference of a square
opening.
A2-5.3 Recommended Measurement/Validation
Method
A2-5.3.1 Measurement of capture velocity at farthest
point of contaminant release. Measurements taken with
a velometer or anemometer.
A2-5.3.2 Confirmation by visualization check of
capture using vapor capture testing.
A2-6 Equipment Gas Panel Enclosures
Equipment gas panel enclosures, also known as gas
boxes, jungle enclosures, gas jungle enclosures, valve