semi合集-English.pdf - 第7388页
SEMI MF1526-95 © SEMI 2004 4 10. Test Specimen Meas urement 10.1 Load the test specimen into t he TXRF instrument in a contam ination free e nvironm ent of Class 100 or better. 10.2 Measure the TXRF spectrum for the test…

SEMI MF1526-95 © SEMI 2004 3
6. Interferences
6.1 The interferences in conventional X-ray fluorescence spectroscopy are common to TXRF also. These include, but are
not limited to: overlap of fluorescence lines, escape peak and sum peak overlap, energy gain calibration drift, X-ray source
stability, and instrumental background peaks. However, no X-ray fluorescence corrections for secondary fluorescence or for
matrix absorption are required for TXRF. Interferences common to software procedures and calculations can be evaluated by
comparing data sets; see Annex A1.
6.2 In addition to conventional interferences, there are interferences that are unique to TXRF as follows:
6.2.1 If the glancing angle calibration is not reproducible, variability is introduced to the measurement,
6.2.2 If the glancing angle calibration is inaccurate, bias is introduced to the measurement,
6.2.3 If the anglescan of the known elemental impurity on the calibration specimen is different from the elemental impurity
anglescan on the test specimen, this may introduce a bias to the quantification. An example can be the measurement of
particulate metal contamination on a test specimen while using a calibration specimen which has the calibration metal
localized within the native oxide,
6.2.4 Mechanical vibration may degrade the detector energy resolution, and this may degrade detection limits,
6.2.5 If the specimen is not chem-mechanically polished, there will be a loss in detection capability, a bias in quantification,
and an increase in variability of the measurement. There is some possibility that the surface roughness and waviness
differences from different cleaning processes may cause these interferences also. The semiquantitative effect of surface
roughness upon the TXRF measurement is under study.
6.2.6 Bias in the assigned elemental areal density of the calibration specimen can introduce bias into the TXRF measured
areal densities,
6.2.7 Surface contamination introduced during handling of the test specimens or during the measurement itself will
introduce a bias to the measurement, if this surface contamination contains the element(s) to be measured,
6.2.8 Bias in the RSF of a fluorescence line can introduce bias to the measurement,
6.2.9 Nonlinearity of detected fluorescence signal versus impurity areal density may occur due to high deadtime of some
detectors under the condition of high-total signal count rates, and
6.2.10 Fluorescence curve smoothing may affect the quantification accuracy.
6.2.11 Instrumental peaks may be generated by the incident X-ray beam diffracting from the silicon crystal and the
diffracted beam entering the detector to excite metals in the detector window or detector housing. This effect can be tested for
by appropriate experiments.
7
7. Apparatus
7.1 TXRF Instrument, equipped with a monochromatic X-ray source, test specimen handling equipment, a method for
glancing angle calibration, an energy-dispersive spectrometer X-ray detector, software for background subtraction, peak
integration, analysis and RSFs, and an analysis ambient without argon (for example, vacuum of 10
2
torr, or helium gas). The
methods for glancing angle calibration are presently proprietary for each TXRF instrument manufacturer. A TXRF
manufacturer may include a subtraction routine for escape peaks and these signals may already be removed.
7.2 Reference Wafer— The suitability of the apparatus shall be determined with the use of a reference wafer and its
associated data set in accordance with the procedures of Annex A2, or by performance of a statistically-based instrument
repeatability study to ascertain whether the equipment is operating within the manufacturer's stated specification for
repeatability.
7.3 Class 100 Air Environment—The area for sample transfer to the instrument measurement stage must be enclosed in
this.
8. Sampling
8.1 A sampling procedure must be used to evaluate the characteristics of a group of silicon wafers. No general sampling
procedure is included as part of this test method, because the most suitable sampling plan will vary considerably depending
upon individual conditions. For referee purposes, a sampling plan shall be agreed upon before conducting the test. See
Practice E 122 for suggested choices of sampling plans.
9. Specimen Requirements
9.1 Test specimens must be chem-mechanical polished on the side used for analysis.
7 Yakushiji, K., Ohkawa, S., Yoshinga, A., and Harada, J.,“ Origins of Spurious Peaks of Total Reflection X-Ray Fluorescence Analysis of Si Wafers
Excited by Monochromatic X-Ray Beam W1-beta,” Japanese Journal of Applied Physics, Vol 33, 1994, pp. 1130—1135.

SEMI MF1526-95 © SEMI 2004 4
10. Test Specimen Measurement
10.1 Load the test specimen into the TXRF instrument in a contamination free environment of Class 100 or better.
10.2 Measure the TXRF spectrum for the test specimen.
10.3 Calculate the net integrated counts/second for detected elemental peaks.
10.4 Using the calibration wafer data for the calibration element and the RSFs for the other elements, calculate the
elemental areal densities for each detected element on the test specimen according to Eq. 1.
D
um
=
()
1/F
sm
×
()
CPS
um
×
()
CPS
s
1
×
()
A
s
(1)
where:
D
su,m
= the areal density in atoms/cm
2
of element “m” on the
test specimen “u”,
F
s,m
= the RSF of element “m” with respect to the
calibration element “s”,
CPS
u,m
= the integrated counts/second of element “m”
detected on test specimen “u”,
CPS
s
= the integrated counts/second of the calibration
element on the calibration standard, and
A
s
= the assigned areal density in atoms/cm
2
of the
calibration element.
11. Calibration Standard
11.1 The calibration standard and the method of its preparation must be agreed upon between the parties.
NOTE 1—Several methods for preparing calibration standards are in practice.
8
The present state-of-the-art calibration standard is a chem-mechanical polished silicon substrate with a surface metal areal density of between
1 by 10
12
and 1 by 10
14
atoms/cm
2
located in the analysis area. The k-alpha fluorescence signal from the calibration standard metal is free
of interferences, that is, escape peaks, sum peaks, and fluorescence peaks from other contamination, and free from external contamination
sources. The preferred element should not be easily added as a contaminated element (for example, iron) or diffuse away from the analyte
depth over time (for example, gold or copper). The preferred element is nickel or vanadium. Calibration can also be done using a set of
wafers with different known areal densities of a particular element.
The calibration standard should have been measured by an appropriate analytical method to determine the metal areal density of the
calibration metal. Some analytical methods which have been used for assigning quantitative values to calibration standards include:
(a) Nitrogen-Beam Rutherford Backscattering Spectrometry (N-RBS)—If this method is used, there should not be any other metals present at
levels greater than 1 % of the calibration metal areal density within a mass range of ±5 atomic mass unit. The N-RBS measurement is
absolute and must be made within the analysis area of the TXRF. Other backscattering methods are also used to calibrate elemental areal
densities. These include forward scattering Rutherford backscattering (F-RBS) and heavy ion backscattering spectrometry (HIBS).
(b) Vapor Phase Decomposition—This followed by atomic absorption spectrometry (VPD/AAS) is used to calibrate a spin coating
contamination process for making calibration standards. The VPD/AAS measurement is destructive. The contamination process should also
be shown to be uniform across the wafer by using TXRF or SIMS mapping. The VPD/AAS absolute assignment is based upon Atomic
Absorption Spectroscopy standards. An error in the elemental recovery rate of VPD is a source of error in accuracy.
(c) Implant—Ion implant the reference element into a pre-amorphized silicon surface and using solid phase epitaxy to regrow the amorphous
silicon into crystal silicon.
9
This procedure sweeps the ion implanted reference element to the sample surface if the reference element is
much more soluble in the amorphous silicon than in the crystal silicon. The assignment of the quantitative level in atoms/cm
2
is done using
the ion implant current to determine the implant dose.
(d) Diluted Atomic Absorption Standard Solution—Deposit onto a polished silicon substrate a diluted atomic absorption standard solution of
a metal onto a localized spot much less than the size of the TXRF analysis area. This approach assumes none of the calibration metal is lost
8 Hockett, R. S., “TXRF Reference Standards: A Discussion,” Contamination Control and Defect Reduction in Semiconductor Manufacturing III, Vol 94-9,
edited by Dennis N. Schmidt, The Electrochemical Society, Pennington, NJ, 1994, pp. 323–334.
9 Hockett, R. S., and Jacobson, D. C., “A New Approach to TXRF Standards,” Extended Abstracts, Vol 93-1, The Electrochemical Society, Inc., Pennington,
NJ, 1993, p. 1289; Jacobson, D. C., Poate, J. M., Higashi, G. S., Boone, T., Eaglesham, D. J., and Hockett, R. S., “Ion Implanted Calibration Standards for Si
Surface Contamination Detection by TXRF,” Materials Research Society, Proceedings of the Spring '93 MRS Meeting, San Francisco, April 26–29, 1993.

SEMI MF1526-95 © SEMI 2004 5
during the drying of the solution. The deposition spot must be easily found by the TXRF instrument in order to analyze the entire deposited
dried solution. The absolute assignment is based upon the AAS standards. The deposited solution must produce one of two TXRF anglescan
forms:
(1) A metal fluorescence count rate versus angle that is independent of glancing angle for angles 80 % below the critical angle (see Curve (a)
in Fig. 2
10
).
(2) A metal fluorescence count rate versus angle that is characteristic of metal contamination localized within a 3-nm thickness of the surface
(see Curve (b) in Fig. 2).
11.2 An anglescan of the calibration standard must be made to verify if the anglescan of the calibration element is similar to
Curve (a) (residue) or Curve (b) (in the film) in Fig. 2.
11.3 An appropriate non-TXRF measurement of the calibration standard must be agreed upon between parties in order to
assign the elemental areal density to the calibration standard element.
11.4 The calibration of the other elements detectable by the TXRF instrument must be completed using relative sensitivity
factor (RSFs) developed by the instrument manufacturer and stored in the instrument computer program. These RSFs are a
function of the X-ray source energy, the atomic number of the fluorescing element, and the fluorescence energy level, so if the
X-ray source energy is changed, a different set of RSFs must be used.
12. Procedure
12.1 Turn on the instrument in accordance with the manufacturer's instructions.
12.2 Analytical Conditions:
12.2.1 Choose and record the analytical conditions for the test specimen measurement. This includes:
12.2.1.1 Voltage to the X-ray source,
12.2.1.2 Current to the X-ray source,
12.2.1.3 Ambient in the analysis chamber,
12.2.1.4 Glancing angle,
12.2.1.5 Integration time,
12.2.1.6 X-ray source energy, and
12.2.1.7 Analysis location on the test specimen.
13. Calibration
13.1 Load the calibration specimen into the TXRF instrument in a contamination free environment defined by Class 100 or
better.
13.2 Measure the TXRF spectrum of the calibration specimen under the same operating conditions of current and voltage
applied to the X-ray source and of glancing angle to be used for the measurement of the test specimen. The measurement time
and the ambient (for example, vacuum, air, nitrogen, helium, etc) in the analysis chamber may be different between the
calibration and test specimen measurements. If the calibration anglescan is similar to Curve (b) of Fig. 2, use a glancing angle
which is 70 to 80 % of the critical angle. If the calibration anglescan is similar to Curve (a) of Fig. 2, use any glancing angle
that is below 85 % of the critical angle.
13.3 A polished silicon substrate with no detectable elemental contamination (that is, a BLANK), or a series of polished
silicon substrates each of which shows no detectable metal contamination for a selected element of interest (that is, a set of
element-specific BLANKs) and which collectively represent a blank for all elements of interest, must be measured by the
TXRF measurement under the condition of the test specimen to verify three are no instrumental background signals. An
example BLANK for all elements except sulfur is shown in Fig. 3.
13.4 Have the TXRF instrument integrate the calibration elemental fluorescence signal and subtract the background to
obtain a net integrated counts/second that corresponds to the assigned elemental areal density. The background may be
subtracted by a deconvolution routine or by a linear fit routine.
14. Report
14.1 Record the following information (see also 12.2.1):
14.1.1 Test specimen identification,
14.1.2 Calibration specimen identification,
10 Eichinger, P., Rath, H. J., and Schwenke, H., “Application of Total Reflection X-Ray Fluorescence Analysis for Metallic Trace Impurities on Silicon
Wafer Surfaces,” Semiconductor Fabrication: Technology and Metrology, ASTM STP 990, ASTM, 1989, pp. 305–313.