semi合集-English.pdf - 第771页
SEMI E104-0303 © SEMI 2000, 2003 7 wavelength of the incident li ght beam , diameter, s hape, and refractive index of the particle and the suspension fluid, as well as on the geom etrical layout of th e collection optics…

SEMI E104-0303 © SEMI 2000, 2003 6
8.3 Humidity
8.3.1 The LPPD should work correctly at an ambient
humidity range RH as specified in ISO 14644-5. If the
use of an LPPD at an extended ambient humidity range
is required, the specific LPPD should comply with
these conditions.
8.4 Electromagnetic Compatibility
8.4.1 The equipment should comply with SEMI E33.
8.4.2 Sensors will be incorporated into equipment
either as original equipment or retro-fitted. In either
case, the sensors or the equipment of which they are a
part should comply with the current regulations
covering EMC in the country or region where the
equipment or sensor is used.
8.5 Vibrations
8.5.1 Process equipment designers are advised to
consider the impact of vibration on the performance of
the sensor while it is collecting data. Therefore, they
should minimize vibrations. The designers of the LPPD
sensors are also advised to consider the impact of
vibrations on the equipment at a time the sensor is not
collecting any data. It might be possible that these
vibrations are stronger than those occurring while the
sensor is collecting data.
9 Reference Calibration Procedure
9.1 The response of real contamination particles,
typically with refractive indices and shapes different
from calibration particles, will differ slightly from the
results obtained by the procedures in this document. It
is known that LPPDs with different optical design may
not produce the same data from identical aerosol
samples. This may happen even with similar LPPDs if
calibration differences have occurred. Therefore, before
the first use the sensor should be calibrated by the
manufacturer. This calibration should be compliant
with or should be reviewed with the following reference
calibration equipment and procedure. This reference
calibration allows the characterization of the
performance of the LPPD under test. The LPPD should
be recalibrated at regular intervals and also in case of
unusual measurement readings to ensure correct results.
9.1.1 The parameters calibrated for LPPDs with sizing
capability are:
• Sizing calibration,
• Resolution,
• Zero counting,
• Counting efficiency, and
• Sensitivity.
9.1.2 The parameters calibrated for LPPDs with non-
sizing capability are:
• Zero counting, and
• Counting efficiency.
9.1.3 Field calibration may not necessarily require the
calibration of all parameters performed by calibration at
the sensor manufacturers site.
9.1.4 Due to the fact that different LPPDs might be
working with different detection areas A
L
, the number
of counts should be printed out in counts per mm
2
detection area. Therefore, a comparison of the
measurement results of different LPPDs is possible.
The size of the detection area A
L
shall be reported in
the calibration report form. Any changes of calibration
parameters or of the calibration setup shall be reported
in the calibration report form. A copy of the calibration
report form shall be delivered with the sensor.
9.2 Apparatus
9.2.1 Materials
9.2.1.1 Particles
9.2.1.1.1 Calibration particles are polymer spheres
composed of polystyrene or a similar polymer, having a
refractive index of 1.58-1.61+ 0i (absorption coefficient
α = 0), a sizing accuracy of at least 95%, and a size
distribution in which the coefficient of variation is 5%
or less. They should be traceable to a nationally or
internationally recognized standard (e.g. NIST
9
).
9.2.1.1.2 The calibration particles are normally
supplied in concentrations too high to be used directly
in aerosol generators. The particles should be dispersed
and diluted in either deionized, distilled water in
accordance with ASTM D1193, Type 1, or Isopropanol.
The diluent should be cleaned using a filter with a pore
size no more than 10% of the size of the particles being
used. The solution should be stored in a clean container.
For generation and dilution of the suspension see
Appendix 3. After generation the particles should be
neutralized to avoid surface charge. For all tests
described in this document, the concentration should be
no more than 25% of the maximum recommended
concentration limit specified by the manufacturer. For
calibration a suitable set of particle sizes shall be used.
This set should contain at least 5 sizes that cover the
LDL size to at least 80% of the specified maximum size
measurement capability of the LPPD.
NOTE 4: Most of the ISPM sensors are based on light
scattering. The intensity of the scattered light detected by a
photodetector depends on intensity, polarization state, and
9 National Institute of Standards and Technology, 100 Bureau Drive,
Gaithersburg, MD 20899-0001, USA

SEMI E104-0303 © SEMI 2000, 2003 7
wavelength of the incident light beam, diameter, shape, and
refractive index of the particle and the suspension fluid, as
well as on the geometrical layout of the collection optics and
detector. In the particle size range near the wavelength (0.1λ
< d
P
< 10λ), large oscillations can be seen in the intensity
curve of the light scattered by spherical particles as a function
of all these parameters. This phenomenon should be taken
into account when selecting a suitable set of particle sizes for
calibration. The used calibration particles should be within a
monotonic response range of the LPPD response curve.
9.2.1.2 Zero Gas
9.2.1.2.1 Clean air or nitrogen filtered with a ULPA
filter.
NOTE 5: The calibration will be executed under atmospheric
pressure and zero gas. In semiconductor manufacturing,
pressure and process gases will differ from the calibration
conditions. This will affect the refractive index ratio of the
particles and the process gas and consequently the scattering
from the particles.
9.2.1.3 Surfaces
9.2.1.3.1 The materials of pump lines and other
components should be conductive to minimize
electrostatic interaction with the particles.
9.2.2 Instrumentation (see Figure 1) — Some LPPDs
may require specialized equipment not generally
available. Please contact the LPPD manufacturer.
9.2.2.1 Fan System
9.2.2.1.1 The fan or pump system should be adjustable
to transport the aerosol and the zero gas within the
stipulated velocity range v
L, min
... v
L,max
, specified by
the manufacturer of the LPPD.
9.2.2.2 Filter System
9.2.2.2.1 The ULPA filter system is used for gener-
ation of zero gas. The filter system should be capable of
removing particles at the minimum size detectable by
the LPPD or the reference particle counter.
9.2.2.3 Aerosol Generator
9.2.2.3.1 An atomizer converts the monodisperse
particle suspension to an aerosol by using compressed
zero gas for generation and transportation of the
particles. The aerosol generator should generate
monodisperse particles as defined in Section 9.2.1.1 in
constant and reproducible concentration C
G
under
constant and reproducible volume flow rate V´
G
. The
variation in particle concentration shall be no more than
10% as measured by the reference particle counter over
a time period of 10 times or more of the sample
measurement time. The generation should comply with
the German VDI-Richtlinie 3491, or an equivalent
standard in other countries.
9.2.2.4 Aerosol dryer
9.2.2.4.1 The monodisperse polymer spheres in the test
aerosol shall be thoroughly dry to avoid that the
particles have a water layer which would increase their
size. A diffusion dryer, another appropriate instrument,
or adequately dry dilution air should be used to dry the
particles. The diffusion dryer uses silica gel desiccant to
remove the moisture. The desiccant shall either be new
or freshly regenerated. The design flow rate of the
aerosol dryer shall at least match the output flow rate of
the aerosol generator.
NOTE 6: Some diffusion dryers may precipitate polymer
spheres and add other particles when the aerosol gets in direct
contact with the silica gel.
9.2.2.5 Neutralizer
9.2.2.5.1 An aerosol neutralizer should be connected in
line with the dryer to reduce electrostatic charges on the
dry polymer spheres and to avoid electrostatic
interaction with each other or the line wall. The design
flow rate of the aerosol neutralizer shall at least match
the output flow rate of the aerosol generator.
NOTE 7: Some electrostatic neutralizers may produce a large
number of ultra-fine particles which will combine with the
calibration aerosol.
9.2.2.6 Aerosol Size Separator
9.2.2.6.1 In the case of calibration with ultra-fine
particles, a system should be used to separate single
polymer spheres from the residual particles resulting
from vaporization of solutions and aggregate particles
consisting of several spheres. The size separation could
be achieved with an electrostatic classifier or a
diffusion battery.
9.2.2.7 Aerosol Dilution
9.2.2.7.1 If the particle concentration behind the
aerosol generator is too high, the particle flow shall be
diluted to achieve the required concentration and to
avoid coincidence errors. The dilution should comply
with the German VDI-Richtlinie 3491, or an equivalent
standard in other countries.
9.2.2.8 Aerosol Line System
9.2.2.8.1 The system consisting of
• the aerosol generator,
• the aerosol dryer,
• the neutralizer,
• the particle size separator,
• the dilution stage, and
• tubing connecting the devices with each other and
the filtered, dried and compressed zero gas.

SEMI E104-0303 © SEMI 2000, 2003 8
9.2.2.8.2 The line system should be smooth, conduc-
tive, and electrically grounded to minimize electrostatic
interaction of the particles with line walls and the parti-
cles themselves. The line should be as short and straight
as possible with no bends with a radius of curvature less
than 100 mm. Leak-free connections of the line and all
devices should be ensured using appropriate fittings.
9.2.2.8.3 Figure 1 illustrates a recommended
calibration aerosol generation system.
9.2.2.9 Flow Control
9.2.2.9.1 The velocity of aerosol and the zero gas
should be within the stipulated range v
L, min
... v
L,max
,
specified by the manufacturer of the LPPD. Dependent
on the line diameter D
L
, a stipulated aerosol flow is
necessary. The velocity meter (e.g. thermoanemometer)
or flow meter should be mounted in a line with known
diameter D
VM
.
NOTE 8: The aerosol drawn by the reference particle counter
influences the velocity or flow measured by the velocity
meter or flow meter if it is mounted behind the probe inlet.
This is taken into account when calculating and adjusting the
flow and velocity v
L
at the LPPD. If the velocity meter or
flow meter is mounted in front of the probe inlet of the
reference particle counter, the device should have no
influence on the particle size distribution measured by the
reference particle counter.
9.2.2.10 Calibration Line System
9.2.2.10.1 The system consisting of
• a device to inject the particles into the zero gas,
• a device to mix the aerosol sample with the zero
gas to obtain uniform particle concentration,
• a device to adapt the LPPD into the line (diameter
D
L
),
• a device to adapt the velocity meter or the flow
meter,
• a device (diameter D
RC
= D
L
, same particle
concentration as at the LPPD) to draw a defined,
isokinetic sample from the line for the reference
particle counter (the opening area of the probe inlet
should be reported in the calibration report form),
and
• tubing connecting the devices with each other and
with the fan and filter system.
9.2.2.10.2 The line system should be smooth, conduc-
tive, and electrically grounded to minimize electrostatic
interaction of the particles with line walls and the parti-
cles themselves. The line should be as short and straight
as possible with no bends with a radius of curvature less
than 100 mm. Leak-free connections of the line and all
devices should be ensured using appropriate fittings.
9.2.2.10.3 Figure 1 illustrates a recommended LPPD
calibration system.
NOTE 9: The distance between the device to adapt the LPPD
under test and the device for sample acquisition for the
reference particle counter should be as short as possible to
minimize particle loss and to ensure comparable particle
concentrations.
9.2.2.11 Reference Particle Counter (RPC)
9.2.2.11.1 The reference particle counter is required to
measure the actual concentration of the monodisperse
aerosol and the quality of the zero gas inside the line.
Therefore, the counting efficiency of the reference
particle counter is defined as 100% over the range of
particle sizes used in the test. The resolution should be
better than 10% at the lower detection limit of the
LPPD under test. The values of the measurement
should be printed out in counts per mm
2
opening area
A
RC
of the probe inlet. The sample transit line from the
probe inlet and the reference particle counter should be
as short and as straight as possible. Smooth, conductive,
and electrically grounded materials should be used.
9.2.2.12 Sensor Window Temperature
9.2.2.12.1 Some LPPDs have the capability to heat
their sensor windows to avoid coating of the window. A
device to measure and to adjust the sensor window
temperature within the operating range (measurable to
5%) should be installed in the calibration setup. The
thermometer should be calibrated with an accuracy of
0.2° C.
9.2.2.13 Pulse Height Analyzer (PHA)
9.2.2.13.1 The external analyzer is connected with the
analog sensor calibration output. The PHA should have
at least 64 channels and a resolution of at least 1% of
the average voltage that will be measured. The use of a
PHA, which is built into the sensor electronics, is
allowed if this PHA meets the criteria mentioned above.
The required range and speed will depend on the
performance of the LPPD under test. These parameters
should be obtained by the LPPD manufacturer.
9.2.2.14 Environmental Monitoring
9.2.2.14.1 The environmental temperature T
A
is
measured by a thermometer calibrated with an accuracy
of 0.2° C.
9.2.2.14.2 The environmental relative humidity RH is
measured by a hygrometer.
9.2.2.14.3 The atmospheric pressure p is measured by a
barometer calibrated with an accuracy of 133 Pa.