CP030电流探头规格书.pdf - 第4页

SPECIFICA TIONS & ORDERING INFORMA TION Specications CP030-3M CP030B CP031 CP031A CP150B (CP150-6M) CP500 Electrical Characteristics Max. Continuous Input Current 30 A rms 150 A rms 500 A rms Bandwidth 10 MHz 50 MHz…

100%1 / 4
2
Maximum Input Current vs. Frequency*
Insertion Impedance vs. Frequency*
(typical)
CP031/ CP031A CP030B
Maximum Input Current vs. Frequency
Insertion Impedance vs. Frequency
(typical)
*
This performance does not apply to CP030-3M model.
Current
(A
rms
)
Frequency (Hz)
3
CP150B CP500
Maximum Input Current vs. Frequency*
Insertion Impedance vs. Frequency*
(typical)
Maximum Input Current vs. Frequency
Insertion Impedance vs. Frequency
(typical)
*This performance does not apply to CP150-6M model.
SPECIFICATIONS & ORDERING INFORMATION
Specications
CP030-3M CP030B CP031 CP031A
CP150B
(CP150-6M)
CP500
Electrical Characteristics
Max. Continuous
Input Current
30 A
rms
150 A
rms
500 A
rms
Bandwidth
10 MHz 50 MHz 100 MHz
10 MHz
(5 MHz)
2 MHz
Rise Time (typical)
≤ 35 ns
≤ 7 ns
≤ 3.5 ns
≤ 35 ns
(≤ 70 ns)
≤ 175 ns
Max. Peak Current
50 A
peak
(non-continuous)
300 A
peak
(non-continuous);
500 Apeak ≤ 30 μs
700 A
peak
(non-continuous)
Output Voltage 0.1 V/A 0.1 V/A & 1 V/A 0.1 V/A 0.1 V/A & 1 V/A 0.01 V/A & 0.1 V/A
Max Continuous Input Current
at 1 V/A (100mA/div or less)
– 5 A – 5 A –
Offset Range at 1V/A
(100mA/div or less)
– ±5 A – ±5 A –
Minimum Sensitivity 10 mA/div 1 mA/div 10 mA/div 1 mA/div 100 mA/div
Low-Frequency Accuracy 1%
AC Noise at 20 MHz BWL
≤ 2.5 mA ≤ 150 µA ≤ 2.5 mA ≤ 150 µA ≤ 1.8 mA ≤ 8.0 mA
Coupling AC, DC, GND
General Characteristics
Cable Length 1.5 m
(3 m)
1.5 m
2 m
(6 m)
6 m
Weight 240 g
(290 g)
236 g 240 g 260 g
490 g
(600 g)
630 g
Max. Conductor Size
(Diameter)
5 mm 20 mm
Interface ProBus, 1 M
Ω only
Usage Environment Indoor
Operating Temperature
0
˚ C to 40˚ C
Max. Relative Humidity 80%
Max. Altitude 2000 m
Measurement Category No rated measurement category
“Not intended for measurements on circuits directly connected to Mains supply or within Measurement Categories II, III, or IV. These probes are compatible
with any Teledyne LeCroy oscilloscope with a ProBus interface running rmware version 9.3.x.x or greater”
Product Description Product Code
30A; 10 MHz Current Probe - AC/DC, 30 Arms; 50 A Peak Pulse, 3 meter cable (not EMC compliant) CP030-3M
30 A; 50 MHz High Sensitivity Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP030B
30 A; 100 MHz Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP031
30 A; 100 MHz High Sensitivity Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP031A
150 A; 10 MHz Current Probe – AC/DC; 150 A
rms
; 300 A Peak Pulse, 2 meter cable CP150B
150 A; 5 MHz Current Probe – AC/DC; 150 Arms; 500 A Peak Pulse, 6 meter cable (not EMC compliant) CP150-6M
500 A; 2 MHz Current Probe – AC/DC; 500 A
rms
; 700 A Peak Pulse, 6 meter cable CP500
Deskew Calibration Source DCS025
current-probes_ds_07sep23
© 2023 Teledyne LeCroy, Inc. All rights reserved. Specifications, prices, availability, and delivery subject to change without notice.
Product or brand names are trademarks or requested trademarks of their respective holders.
1-800-5-LeCroy
teledynelecroy.com
Local sales oces are located throughout the world.
Visit our website to nd the most convenient location.
Customer Service
Teledyne LeCroy oscilloscopes and probes are designed, built, and tested to ensure high reliability. In the unlikely event you experience
diculties, our digital oscilloscopes are fully warranted for three years and our probes are warranted for one year. This warranty includes:
•
No charge for return shipping
•
Long-term 7-year support
•
Upgrade to latest software at no charge
Ordering Information