4M-1751-005w_G5S - 第72页
4OM-1751 1-22 Filter Holder F4A13 …

4OM-1751
1-211304-001
4.1.4 Filter Cleaning
Cycle : Every 3 Months (Required Time : 5 minutes)
Filter
F4A11
Procedure
case removing jig to remove the cover.
F4A12
4.1 Air Source

4OM-1751
1-22
Filter Holder
F4A13
Note
1304-001
4.1 Air Source

4OM-1751
1-231304-001
4.2 Vacuum System
4.2.1 Vacuum Pressure Check
Cycle : Every Day (Required Time: 1 minute)
Vacuum Pressure
F4A14
Procedure
Check the vacuum pressure visually.
Vacuum Valve Fully Closed : Min. -93 kPa (Min. 70 cmHg) or more
Running : Min. -60 kPa (Min. 45 cmHg) or more
Note
line or the vacuum pump may have broken down.
4.2 Vacuum System