Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a

Operation 147
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Operation
This chapter describes the operation of the inspection system.
It contains the following sections:
Safety (Page 148)
Operating the safety features (Page 149)
Switching on the inspection system (Page 154)
Operating the inspection system (Page 155)
Switching off the inspection system (Page 159)

Operation
Safety
148
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Safety
Follow these instructions when operating the inspection system:
Check the system regularly for externally perceivable defects
and damages as well as changes in operating behavior.
In case of disruptions of electrical supply and/or damage to the
electrical equipment, immediately shut off the main power and
prevent it from being accidentally restarted.
Promptly report and correct all malfunctions and damage
displayed by the system or noted in any other way.
Each time before the system is switched on, check all safety
features.
The safety circuits protect personnel and may not be be used to
shut off the radiation. Frequently opening the doors under load
burns out the contacts and can also damage the high voltage
generator control electronics.
Only authorized persons may use the key switch.
Whenever work is done that requires access to the inspection
system interior, make absolutely certain that the radiation is
turned off.
High voltage in exposed places. To avoid high voltage, turn off
the X-ray tube through the key switch and set the main switch
to the position OFF.