X7056_Operating manual_en.pdf - 第191页
Maintenanc e Lubricating the flat panel detector positioning unit 191 Inspection system X7056 | Operating manual | Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a Y -axis runner block Legend No. Name Size Numbe r Grease i…

Maintenance
Lubricating the flat panel detector positioning unit
190
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Lubricating the flat panel detector
positioning unit
This section describes how to lubricate the runner blocks and spin-
dles of the flat panel detector positioning unit
The following sections describe how to lubricate the flat panel
detector positioning unit:
Positioning unit flat panel detector
Lubricating the runner blocks and spindles
Positioning unit flat panel detector
The illustrations indicate the position of the runner blocks and
spindles to be lubricated. The lubricating nipples f the runner
blocks and spindle nuts may be mounted on the sides or the front.
X-axis runner block
Legend
No. Name Size Numbe
r
Grease in cm3
1 X-axis runner block 15 2 0.4
1
X
Y

Maintenance
Lubricating the flat panel detector positioning unit
191
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Y-axis runner block
Legend
No. Name Size Numbe
r
Grease in cm3
1 Y-axis runner block 15 4 0.4
1
X
Y

Maintenance
Lubricating the flat panel detector positioning unit
192
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
X- and Y-axis spindles
Lubricating the runner blocks and spindles
Required: Lubricating grease KP2K, NLGI class 2 according to DIN 51818
Grease gun
Extension attachment, approx. 120 mm
Lubricating the
X-axis runner
block
Perform the following work steps:
1. Stand behind the inspection system and open the rear swing
doors.
2. Pull the X-axis toward yourself.
3. Check guide carriage and guide rails for contamination; clean if
required.
Legend
No. Name Numbe
r
Grease in cm3
1 X- and Y-axis spindles 2 1.5
1
X
Y