Maintenance
Exchanging the AXI transport belts
218
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
8. Mount the traverse by tightening the screws on the movable
transport plate with the socket wrench (5 mm Allen socket) and
the screws on the movable transport plate with the shortened
socket wrench (5 mm Allen socket).
The traverse must lie against the end stop.
9. Make sure the end stop rail guide catches on the traverse
correctly.
10.Use the socket wrench (2 mm Allen socket) to screw down the
cable carrier mount.
Maintenance
Exchanging the AXI transport belts
219
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
11.Connect the motor cable to the motor coupling.
12.Set the connector and the cable into the clips.
13.Bind the cable with cable ties.
14.Use the socket wrench (4 mm Allen socket) to install the hold-
down clamp.
15.Repeat steps 1 to 14 for the transport belt on the fixed trans-
port side plate.
16.Place a PCB in the transport track and check whether the end
stop guide rails of the movable and the fixed transport side
plates are parallel to each other.
The transport belts are installed.
The AXI transport belts are exchanged.
Maintenance
Maintenance of the sealed X-ray tube
220
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Maintenance of the sealed X-ray tube
This chapter describes the maintenance procedure for X-ray
system components.
The following procedures are covered:
Warm-up procedure (Page 220)
Exchanging the battery (Page 220)
Reconditioning of sealed X-ray tube (Page 222)
Warm-up procedure
Working with the sealed X-ray tube requires a warmup of a dura-
tion depending on how long the tube is switched off (tube high
voltage or supply voltage off). The warmup is started automatically
through the X-ray driver.
The duration of the warm-up depends on the length of time the
tube was switched off:
8h - 1 month: 15 min.
1 - 3 months: 40 min.
over 3 months: 120 min.
Exchanging the battery
This chapter describes the procedure for exchanging the battery of
the sealed X-ray tube.
Precondition: The X-ray tube must be dismantled.
Required: Screw driver (corresponding in size)
Battery, type CR2032 from the service box (VTN 16.901.B.02)
NOTICE
After re-installing the X-ray tube carry out a geometric calibration.