Maintena nce Maintenance of th e sealed X-ray tube 220 Inspection system X7056 | Operating manual | Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a Maintenance of the sealed X-ray tube This chapter describes the maint ena…

100%1 / 280
Maintenance
Exchanging the AXI transport belts
219
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
11.Connect the motor cable to the motor coupling.
12.Set the connector and the cable into the clips.
13.Bind the cable with cable ties.
14.Use the socket wrench (4 mm Allen socket) to install the hold-
down clamp.
15.Repeat steps 1 to 14 for the transport belt on the fixed trans-
port side plate.
16.Place a PCB in the transport track and check whether the end
stop guide rails of the movable and the fixed transport side
plates are parallel to each other.
The transport belts are installed.
The AXI transport belts are exchanged.
Maintenance
Maintenance of the sealed X-ray tube
220
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Maintenance of the sealed X-ray tube
This chapter describes the maintenance procedure for X-ray
system components.
The following procedures are covered:
Warm-up procedure (Page 220)
Exchanging the battery (Page 220)
Reconditioning of sealed X-ray tube (Page 222)
Warm-up procedure
Working with the sealed X-ray tube requires a warmup of a dura-
tion depending on how long the tube is switched off (tube high
voltage or supply voltage off). The warmup is started automatically
through the X-ray driver.
The duration of the warm-up depends on the length of time the
tube was switched off:
8h - 1 month: 15 min.
1 - 3 months: 40 min.
over 3 months: 120 min.
Exchanging the battery
This chapter describes the procedure for exchanging the battery of
the sealed X-ray tube.
Precondition: The X-ray tube must be dismantled.
Required: Screw driver (corresponding in size)
Battery, type CR2032 from the service box (VTN 16.901.B.02)
NOTICE
After re-installing the X-ray tube carry out a geometric calibration.
Maintenance
Maintenance of the sealed X-ray tube
221
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Remove the
battery
DANGER
Beryllium is poisonous!
The X-ray output window of the sealed X-ray tube contains Beryl-
lium.
After removing the sealed X-ray tube, a protective cover has to
be attached to the X-ray window.
Perform the following work steps:
1. Loosen the screws of the battery cover.
2. Press the battery in direction of the arrow and carefully lift it on
the left side and remove it from the bracket.
The battery is removed.
Inserting the
battery
Perform the following work steps:
1. Insert the new battery.
The battery positive terminal points upwards.
2. Fasten the cover of the battery compartment.
The battery is inserted
The battery is replaced.