X7056_Operating manual_en.pdf - 第79页
Installation Electrical in stallation 79 Inspection system X7056 | Operating manual | Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a The inspection system is properly set in place. NOTICE Each system must be assigned a…

Installation
Electrical installation
78
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Electrical installation
This section describes the proper electrical installation of the
inspection system.
The following steps are described:
Electrical specifications (Page 78)
Providing a standard connection (Page 78)
Checking the safety conductor PE (Page 80)
Connecting FE (TE) (Page 80)
Check motor overload switch settings (Q1EA) (Page 81)
Electrical specifications
NOTICE
The system can optionally be equipped with a network trans-
former, see „Product description“ > „Network transformer“
(Page 37).
Providing a standard connection
Precondition: The inspection system is connected to the power supply with a
16A/6 h CEE plug.
The inspection system requires the following electrical connec-
tion: 230/400 V AC, 50/60 Hz, 3 phase, N, PE
The main switch is set to OFF.
The connection requirements are met, see section „Electrical
specifications“ (Page 78).
Electrical specifications
Power supply 230/400 VAC, 50/60 Hz, 3 phase, N, PE
Power supply with internal
transformer for main power
adaptation
190 V - 600 VAC, 50/60 Hz, 3 P/N/PE
Average power consumption 0.78 kW/h
Max. Power 1.5 kW
Min. Power 0.62 kW
Motor overload switch See current circuit diagram
Automatic cutout 8 A characteristic "C"

Installation
Electrical installation
79
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
The inspection system is properly set in place.
NOTICE
Each system must be assigned a separate leakage circuit breaker.
DANGER
Electrical shock
Danger to life from improper conduct.
Before all electrical work, make sure that:
the main power supply is disconnected and
the system is protected from accidental switching-on.
CAUTION
Danger of tripping
Injury due to improperly laid cable.
When laying the cable, make sure there are no kinks, places to
trip or tangles.
Perform the following work steps:
1. Open the rear swing doors of the inspection system.
2. Take the connection cable with the CEE plug.
3. Connect the inspection system to the main electrical power
supply.
The standard connection is done.

Installation
Electrical installation
80
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Checking the safety conductor PE
DANGER
Electrical shock
Danger to life from disconnected safety conductor.
The safety conductor must be connected to prevent electrical
shock.
The system may not be operated with a disconnected safety
conductor under any circumstances.
Perform the following work steps:
1. Open the front door.
2. Open the pivoting frame, see „Open the pivoting frame“ (Page
87).
3. Make sure that the safety conductor is properly connected.
The safety conductor is checked.
Connecting FE (TE)
NOTICE
The FE (TE) connection improves electromagnetic compatibility
(EMC).
Perform the following work steps:
1. Open the front door.
2. Open the pivoting frame, see „Open the pivoting frame“ (Page
87).
3. Connect a flexible conductor with a minimum cross section of
10 mm
2
, which is connected to the potential equalization point.