Installation First use 98 Inspection system X7056 | Operating manual | Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a 4. Under S TART MENU > M AINTENA NCE PROGRAM , sele ct the prog ram X RAY C LOSED T UBE C ONTROL . …

100%1 / 280
Installation
First use
97
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
11.Enter the value from si_trans.conf in mm in JAT MAINTENANCE
Tool in the tab Movement - Absolute.
12.Confirm with Enter; click the button START.
The lift gate is closed.
The Interlock connections in the program XRAYCLOSEDTUBE-
C
ONTROL are closed.
The lift gate functioning is checked
Checking the
shutter func-
tioning
Precondition: The inspection system is switched on.
The AXI transport track must be in the AXI section.
Perform the following work steps:
1. End the Viscom software, see „Ending the Viscom software“
(Page 158).
2. Actuate the STOP mode, see „Actuating the STOP mode“ (Page
149).
The shutter gate is opened.
3. Cancel the STOP mode, see „Ending the STOP mode“ (Page
150).
Installation
First use
98
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
4. Under START MENU > MAINTENANCE PROGRAM, select the program
X
RAYCLOSEDTUBECONTROL.
The start menu of the VXC software appears.
The Interlock connections are open.
5. Under S
TART MENU > MAINTENANCE PROGRAM, select the program
JAT MAINTENANCE.
The dialog SELECT AXES opens.
6. In the menu, select the shutter from (X
R_Z); confirm with OK.
The program JAT MAINTENANCE opens.
Installation
First use
99
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
7. Select SWITCH ON.
The shutter motor switches on.
A commutation takes place the first time the motor is switched
on.
After the switching process is complete, the SWITCH ON button
is bordered in green.
8. Select S
TART HOME.
A reference run is made.
After the referencing is complete, the START HOME button is
bordered in green.
9. In directory  select the file
si_trans.conf.