m330_01_gas_cabinet气柜 - 第24页

START UP G AS CABINET R EFERENCE M ANUAL 4-2 • Mark pressure reading • Leave gassystem pressurized for at least 12 hours • Mark pressure reading. Any drop in pressure indicates a gas leak. • Use suitable gasleak detector…

100%1 / 31
START UP
4.Start Up
4.1 Introduction
After the gas cabinet is installed and connected correctly, start-up can begin.
Valve operation and gasflows (including pressure control for LPCVD) will be controlled by
the DPC.
In case of emergency or system failure, all system power can be switched off by the EMO-
switch, forcing all NC valves to be closed and gaslines to be shut off. See the outline drawing
for the exact locations of the EMO-switch on the gascabinet.
4.2 Prerequisites
First the furnace, loadstation, gascabinet and powercabinet installation must have been
completed.
Second, the start-up procedure of the furnace as described in the “Diffusion Furnace”
manual, chapter 5.0 (Start-Up) must be completed.
4.3 Gas Cabinet start up
For each gasline:
Check connections
Make sure gaslines are connected to the processtube
Make sure the tube is sealed properly (boat in)
Close door
Close MFC’s
Close valves
Close handvalves
Close regulators
Close facility gas supply
Close gas bottle (depending on customers facilities)
Atmospheric
Open facility gassupply
Set regulator to 30 psi N
2
and
20 – 25 psi H
2
, O
2
Open handvalves on all tube
levels
Close facility gas supply
LPCVD
Start the pump
Evacuate the tube
Open process valve (manifold)
Pump until base pressure is
reached
Maximize MFC setpoint
GAS CABINET REFERENCE MANUAL
4-1
START UP
GAS CABINET REFERENCE MANUAL
4-2
Mark pressure reading
Leave gassystem pressurized for
at least 12 hours
Mark pressure reading. Any drop
in pressure indicates a gas leak.
Use suitable gasleak detector
(such as Snoopy or electronic) to
trace any leak
When no pressure drop is
detected open the facility gas
supply.
The gassystem is ready for use.
Pump until base pressure
Open handvalve
Pump until base pressure
Open regulator fully
Pump until base pressure
Open facility gas supply slowly
Pump until base pressure
Leakcheak max 20 mtor/min right
up to the bottle
Open N
2
purge at gas bottle
cabinet briefly. Pressure should
not exceed the safety setpoint of
1.4 torr measured at the tube.
Pump until base pressure
Repeat N
2
purge/ Pump cycle 20
times.
Leakcheck max. 20 mtor/min
Close facility gas supply
Close regulator
Close handvalve
Close MFC
Close process valve
Slowly open the gas bottle
Slowly open facility gas supply
Slowly open the regulator
Set pressure at 20 psi. If gaslines
are shared with other tubes use 30
psi for N
2
and 20-25 psi for O
2
Slowly open the handvalve at all
applicable tube levels
Close facility gas supply
Mark pressure reading
Leave gassystem pressurized for at
least 12 hours
Mark pressure reading. Any drop
in pressure indicates a gasleak.
Use suitable gas leak detector
(such as Snoopy or electronic) to
trace any leak
When no pressure drop is
detected open the facility gas
supply.
The gassystem is ready for use.
MAINTENANCE
GAS CABINET REFERENCE MANUAL
5-1
5.Maintenance
5.1 Introduction
Maintenance for the gascabinet and its components is limited due to the use of high quality
components and high purity elements.
Routine checks include pressure gauge readings every day and inline filter replacement every
year.
5.2 Leak-checks
Whenever a component in the gasline (either the preplumbing or gaspanel) is replaced a full
leakcheck is required.
Two different kinds of leak-checks are possible. There are a helium- and a pressure leak-
check.
5.2.1
Helium Leak check
A Helium leak check is used to test the integrity of the components or complete gasline.
Connect the leaktester on a gas-inlet;
Open the pressure regulator fully to maximum;
Open the handvalve at tube level;
Open the MFC (this must be done by the DPC, for more information, look at the
manual of the DPC);
Open all valves in the gassystem (this must also be done by the DPC);
Leak check may now begin;
If no leak is found, repeat steps, mentioned above, in reversed and opposed
order. However, if a leak has been found, tighten/repair the relating nut/coupling
(etc.).
NOTE
Leak must be below 1 * 10
-8
atm. cm
3
/sec.
NOTE
If a checkvalve is present then a leak-check must be done
at both sides of this valve.