m330_01_gas_cabinet气柜 - 第26页

MAINTENANCE G AS CABINET R EFERENCE M ANUAL 5-2 5.2.2 Pressure Leak-check A pressure test is used to test all connections at supply pressures see Gas Cabinet start up 4.3. 5.3 Disassembly and cleaning of MFC’s MFC’s are …

100%1 / 31
MAINTENANCE
GAS CABINET REFERENCE MANUAL
5-1
5.Maintenance
5.1 Introduction
Maintenance for the gascabinet and its components is limited due to the use of high quality
components and high purity elements.
Routine checks include pressure gauge readings every day and inline filter replacement every
year.
5.2 Leak-checks
Whenever a component in the gasline (either the preplumbing or gaspanel) is replaced a full
leakcheck is required.
Two different kinds of leak-checks are possible. There are a helium- and a pressure leak-
check.
5.2.1
Helium Leak check
A Helium leak check is used to test the integrity of the components or complete gasline.
Connect the leaktester on a gas-inlet;
Open the pressure regulator fully to maximum;
Open the handvalve at tube level;
Open the MFC (this must be done by the DPC, for more information, look at the
manual of the DPC);
Open all valves in the gassystem (this must also be done by the DPC);
Leak check may now begin;
If no leak is found, repeat steps, mentioned above, in reversed and opposed
order. However, if a leak has been found, tighten/repair the relating nut/coupling
(etc.).
NOTE
Leak must be below 1 * 10
-8
atm. cm
3
/sec.
NOTE
If a checkvalve is present then a leak-check must be done
at both sides of this valve.
MAINTENANCE
GAS CABINET REFERENCE MANUAL
5-2
5.2.2
Pressure Leak-check
A pressure test is used to test all connections at supply pressures see Gas Cabinet start up 4.3.
5.3 Disassembly and cleaning of MFC’s
MFC’s are complex components, which require specialist tools and skills. It is
recommendable to return the MFC to the supplier for cleaning and/or repair (make sure to
include the proper contamination reports).
Evacuate upstream and downstream of the MFC in case it is inside a gasline for toxic or
flammable gases.
5.4 Disassembly and Cleaning of all flowmeters
The metering tube, float and float stops may be removed as an assembly for cleaning or
replacement, without removing the meter from the line. To disassemble, use the following
procedure:
1. Remove the plastic safety shield.
2. Loosen the seal spindle by turning it counter clockwise with a 4 mm hex wrench. The
tube must be pushed into the direction of the seal spindle and may now be canted out of
the meter housing. If it is desired to clean the metering tube, float and float stops as an
assembly (using a solvent or ultrasonic cleaner) only the above two-steps are required.
However, if complete disassembly is required, use the remaining steps.
3. Using a small hook, remove both Teflon float stops from the metering tube and remove
the float. Be careful not to chip tube ends.
4. Packing seats and packing inserts now may be removed.
5. With the metering tube out, the seal spindle may be rotated clockwise for removal. It
should not be necessary to remove the seal spindle unless the 0-ring that seals the spindle
requires replacement. The O-ring may be used as long as it is not torn or distorted.
6. The needle control valve assembly may be removed by turning the valve body counter
clockwise. The valve seat, stem and packing then may be carefully removed from the
valve body for cleaning or replacement.
7. The metering tube, made of Borosilicate glass, may be cleaned with any solvent that does
not attack glass.
Reassembly Procedure
1. Use the reverse of steps 1 through 7 of the disassembly procedure to reassemble the
meter.
2. When replacing the packing seats in the flowmeter body be sure the packing inserts are
approximately 1,5 mm above the top of the packing seat. Also be certain the tube seats
firmly on the packing seats and does not overlap onto the end block.
3. The seal spindle serves to radial compress the tube seat gasket and exert a uniform
pressure on the metering tube to prevent any possibility of leakage. Do not over tighten
the seal spindle.
MAINTENANCE
GAS CABINET REFERENCE MANUAL
5-3
5.5 Preventive maintenance Gassystem
Daily maintenance:
Check gas pressures approx. 25 psi.
Check for visible signs of leakage (liquid supply lines)
Check Touchscreen and/or TSC system for any alarms.
Check gasflows of MFC’s take in account that certain hardware interlocks are
implemented in the gassystem.
-------------------------------------------------------------------------------------------------------
Weekly maintenance
For Metal Alloy tubes the Hydrogen filaments should be visually checked.
If applicable check the liquid source level
If applicable check the External Torch on flame shape (there should be no
red color in the flame “tip burning”, and check for condensation.
-------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------
Monthly maintenance:
Check amount of exhaust flow (see facility drawing for values).
-------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------