14_PO-CON1714E.PDF

PO-CON1714E High-sensitivity and simultaneous analysis of Psychoactive drugs using LC-MS/MS with full-automated pretreatment system ASMS 2017 WP358 Daisuke Kawakami 1 , Toshikazu Minohata 1 1 Shimadzu Corporation. 1, Nis…

100%1 / 7
PO-CON1714E
High-sensitivity and simultaneous
analysis of Psychoactive drugs using
LC-MS/MS with full-automated
pretreatment system
ASMS 2017
WP358
Daisuke Kawakami
1
, Toshikazu Minohata
1
1 Shimadzu Corporation. 1, Nishinokyo-Kuwabaracho
Nakagyo-ku, Kyoto 604–8511, Japan
2
High-sensitivity and simultaneous analysis of Psychoactive drugs
using LC-MS/MS with full-automated pretreatment system
Introduction
Figure 1 Target drugs
LC-MS/MS has become a preferred method for the routine
analysis for forensic toxicology. LC-MS/MS allows for the
simultaneous analysis of multiple compounds in a single
run, thus enabling a fast and high throughput analysis. In
recent years that it seems the number of incident and
accident is increasing caused by dosed with psychotropic
drugs and the number of drug testing with LC-MS/MS is
also increasing to investigate the cause of death. However,
manual sample preparation often involves several
complicated manual steps which can introduce error into
the results. In this study, we investigated the processing
capability to analyze serum, whole blood and urine spiked
sixty psychotropic drugs by LC-MS/MS with automated
sample preparation unit.
Alprazolam Bromazepam Brotizolam Chlordiazepoxide Clorazepic
acid Clotiazepam Cloxazolam Diazepam Estazolam Ethyl
loazepate Etizolam Fludiazepam Flunitrazepam Flurazepam
Flutazolam Flutoprazepam Haloxazolam Lorazepam Lormetazepam
Medazepam Mexazolam Midazolam Nimetazepam Nitrazepam
Oxazolam Prazepam
Quazepam Rilmazafone Tosopam Triazolam Zolpidem
7-Aminounitrazepam 7-Aminonimetazepam 7-Aminonitrazepam
α-Hydroxyetizolam (M-VI) α-Hydroxyalprazolam α-Hydroxybrotizolam
α-Hydroxytriazolam Zolpidem M-1
Group 3. Thirty-nine Benzodiazepines and their metabolites
Group 1. Eight Barbiturate drug and Bromovalerylurea
Allobarbital Amobarbital Barbital Pentobarbital Phenobarbital
Secobarbital Thiamylal Thiopental Bromovalerylurea
Group 2. twelve Tri-/Tetra-cyclic antidepressant
Amitriptyline Amoxapine Clomipramine Desipramine Dosulepin Imipramine
Maprotiline Mianserin Nortriptyline Promethazine Setiptiline
3
High-sensitivity and simultaneous analysis of Psychoactive drugs
using LC-MS/MS with full-automated pretreatment system
Figure 2 CLAM-2000 and LCMS-8060 system
Methods and Materials
The analysis of 60 psychoactive drugs (eight Barbiturate
drug, thirty-nine Benzodiazepines and their metabolites,
twelve Tri-/Tetra- cyclic antidepressant and
bromovalerylurea) were performed using a fully automatic
LCMS preparation unit (CLAM-2000, Shimadzu) online
with HPLC-LCMS (NexeraX2-LCMS-8060, Shimadzu).
Samples were trapped on Imtakt Unison UK-C18
(10x2mm, 3.0μm), then separated by Imtakt Unison
UK-C18 (75x2mm, 3.0μm) with a binary gradient system.
Water with ammonium formate and methanol were used
for mobile phases.
-Serum
-Whole Blood
-Urine
Report creation
MS/MS
Library
search
Filtration
(PTFE,
0.45μm)
90sec
Shaking
30sec
Filtration
(PTFE,
0.45μm)
90sec
Shaking
60sec
To AutoSampler
Figure 3 Analytical ow of serum and whole blood
Reagent
Dispensing
50 µL of Water
Reagent
Dispensing
300 µL of MeOH
Sample
Dispensing
50 µL
of serum or whole blood
Shaking
60sec
To AutoSampler
Figure 4 Analytical ow of urine
Reagent
Dispensing
100 µL of MeOH
Sample
Dispensing
100 µL
of urine