00195372-0102_AI_Vakuumpumpe_D1D2_DE EN - 第45页

D1 / D2 2 Assembly instructions for vacuum pump 11/2006 Edition 45 : Note the measured value. : Remove the measuring tube from the vacuum distribution block and put the plugs back in. : Close all doors. : Repeat the abov…

100%1 / 54
2 Assembly instructions for vacuum pump D1 / D2
11/2006 Edition
44
2.9.2.2 Measure the vacuum with an external pressure measurement instrument:
at the measuring connection of the vacuum distribution blocks entrance (see Fig. 2.9.4) and
afterwards
at an open nozzle in the holding circuit of the placement head (see Fig. 2.9.5). All other
nozzles must be closed.
Pressure may amount to max. 20-30 mbar! Otherwise everything is not sealed. 2
Fig. 2.9.4 Measurement of the vacuum distribution block (machine entrance)
: Remove for the plugs from the measuring connection at the vacuum distribution block.
: Attach the external measuring instrument.
: Switch the vacuum pump on.
: Open if necessary the inserted vacuum shut-off valves
: Open the compressed air shut-off valves.
: Switch the main switch on.
D1 / D2 2 Assembly instructions for vacuum pump
11/2006 Edition
45
: Note the measured value.
: Remove the measuring tube from the vacuum distribution block and put the plugs back in.
: Close all doors.
: Repeat the above with all attached mounting machines.
Fig. 2.9.5 Measuring a nozzle
: Press the gauge connecting piece of the external measuring instrument to the open nozzle.
If the values specified above are not reached, the cause
usually lies in a not correctly suspended
tube in the head area. Proceed as follows: 2
: Press all hoses down with
pliers.
: Repeat the vacuum measurement.
2
With closed nozzles with attached vacuum pump the vacuum is usually approx. 30-50 mbar
lower in the holding circuit than at a Venturi producer in the pickup circuit (in each case mea-
sured with external measuring instruments at an (open
) pipette in the pickup circuit and one (open)
in the holding circuit). 2
This difference can be accepted, if with 521 m NN (a
ltitude Munich, Start-up SIPLACE ) a value
800 mbar of at least is reached. 2
2 Assembly instructions for vacuum pump D1 / D2
11/2006 Edition
46
2.9.2.3 Checking the holding circuit of the vacuum system
The flow can be controlled “indirectly “over the vacuum characteristic. 2
Proceed as follows: 2
: Re
move the nozzles from the placement head (in order to allow maximum flow.
: Ope
n sequentially in the SITEST the segment in fetching position and turn it by means of as-
terisking into the holding circuit.
The vacuum measurement in the holding circuit sh
ows
with every additionally opened nozzle a
vacuum waste of approx. 60-90 mbar. 2
2
With six opened nozzles the vacuum could for example drop from -880 mbar to -520 mbar.
With six C&P heads on two mounting machines with a common vacuum pump, the decrease of
pressu
re is larger; with only one placement machine smaller (four or two C&P heads). 2
2.9.2.4 Vacuum Characteristics
The following diagrams also apply to the D1 and D2 (provided with other mounting equipment) 2
Fig. 2.9.6 C&P6 (red) and C&P12 (blue): Pressure drop at 1, 2, 3... open nozzles
Vacuum [mbar]
Number of open nozzles
Gantry 1
Gantry 2
Vacuum curve for S27HM with ELMO 2BL 1 1000 vacuum pump and 720 nozzles
Gantry 1: 6-segment head / Gantry 2: 12-segment head