00195963-03_MM SWS-EN - 第59页
59 Ch ange to Me nu Manu al operat ions [1] - > Manu al oper ations subs ystems [2] - > IOs [3] ( see scre enshot below ) Change t o Menu Manu al op erations USBIO [1] and choose int o folder USB outputs [2] (s…

Checking the Rotary Part for Ease-of-Movement
Interval: 3 months
Equipment:
None
Move the rotary part within the raceway
back and forth in the direction of the ar-
rows.
It must be easily turnable and must not get
caught anywhere.
In case of a defect of the linear bearings
the whole flip unit has to be replaced (see
the Service Manual for further details).
If there is a rough movement of the ball
baering at the edges of the raceway from Z-
Axis to Die-Attach, the position from the
“Retract”-Cylinder must be adjusted according
Servicemanual
Checking the Nozzle Attachment for Leaks
Interval: 3 months
Equipment:
Vacuum
CAUTION
For this check the Service user level is required in SWS GUI. Only appropriately trained personal
is allowed to carry out this check.
- SWS must be referenced.
- Open any cover from the CA-Placement machine, to get access to the SWS
The power supply to the gantry axes and the SWS axes is interrupted immediately. The gantry
axes and all SWS axes will come to a standstill. A corresponding message is issued on the
screen. Skip the message.

59
Change to Menu Manual operations [1] -> Manual operations subsystems [2] -> IOs [3]
(see screenshot below)
Change to Menu Manual operations USBIO [1] and choose into folder USB outputs [2] (see
screenshot below) and switch off all Vacuum consumers (Dxx Vacuum …) by pressing OFF
buttons (see. 3), in case one of these outputs is active. This has to be done also for all air
supply consumer outputs (Dxx Pressure ...)

Change from previous screenshot into Manual operations -> Manual operations
subsystems -> IOs -> Manual operations USBIO -> USB programs [1] and read out the
main vacuum value Main Vacuum [mbar] (see screenshot below).
Take care, that the Vacuumpump is Switched Off, like in the Screenshot below..
The Main Vacuum Value should have at least 700 mbar. See Screenshot above.
Write down this value.
Change to Menu Manual operations [1] -> Manual operations subsystems [2] -> IOs [3]
(see screenshot below)