Unpacking Unit
Unpacking
1. Make sure the packaging is laid out in the correct orientation before opening the box.
2. Remove all packaging material around the PM5000 Centrifuge unit before attempting to remove unit
from box.
3. Carefully remove The PM5000 Centrifuge unit from the box.
4. Check unit for damage before discarding shipping packaging and box.
5. Should any damage be detected, please call your local representative of EFD Technical Services for
assistance.
Caution
1. Be careful not to drop the PM5000 Centrifuge, or place / rest the unit upside down. Unit is
heavy and you may require assistance in lifting the PM5000 Centrifuge unit.
2. Do not place the PM5000 Centrifuge in the high draft are, direct sunlight or near equipment
emitting heat and / or electromagnetic conduction.
3. Conditions for transportation & storage:
Temperature 10ºC – 40ºC
Humidity 40% - 90%
4. Check the specification label located on back of this equipment, the power source must
conform to your local codes.

10
www.nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 Asia +86 (21) 3866 9006
Sales and service of EFD dispensing systems are available worldwide.
Set-up
Placement Surface
The PM5000 weighs approximately 35.2 Lbs. / 16 Kg, and will be spinning at high speeds. Select a
level, sturdy and stable surface to place the PM5000 Centrifuge unit before use.
Power Source
The ProcessMate PM5000 is a multi voltage unit with a built in power converter. The unit is capable and
will operate in the ranges of 100 – 240 VAC 50/60 HZ. Locate the power cord shipped with the unit,
plug one end into the unit and the other to a suitable power source. Make sure the ground pin on the
power cord is connected to ground on the power source.

11
www.nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 Asia +86 (21) 3866 9006
Sales and service of EFD dispensing systems are available worldwide.
Preparation
Before using this centrifuge, the following steps should be checked to make sure the PM5000 is ready
and safe for use in a production setting:
1. Plug power cord in a grounded outlet.
2. Turn Power Switch to ON position.
The power lamp ON/OFF, Speed Indicator and Timer must be illuminated.
3. Press LID OPEN switch then push Lid Release knob on the right hand of the housing to place
contents into the centrifuge.
4. Inspect and check if rotor is screwed on securely.
5. Shut lid and set timer by pressing the corresponding keypads “▲” “▲” to set centrifuge spinning
time. Timer is adjustable from 0 – 99 minutes, and timer will count down displaying remaining time
left of spin cycle.
6. Adjust the RPM by turning the control knob. Once the start button is presses, the RPM may not be
altered or changed unless the spin cycle has timed out or has been reset.
Warning:
The maximum RPM depends on the rotor, size of Syringe barrels and weight of contents.
7. Press STOP switch to verify that the spin cycle may be canceled.
8. While in spin cycle, press LID OPEN key and Lid release button to see if the lid can be opened
while in operation. The safety feature on the PM5000 should not allow you to open the lid until the
rotor comes to a complete stop.
9. Switch off the power to verify that power can be shut off in case of emergency.
If the above procedures work as detailed, the centrifuge is operating as designed and ready for
production use.
Note:
IF the ERROR sign of Err-5 is displayed on screen at any time, the cause is usually due to an
“imbalance rotor”. Check contents within the centrifuge and make sure the syringe barrels are equally
loaded and filled.