00196101-04_Serviceanleitung Vakuumtest 08-2018_D-E - 第128页

Service manual vacuum check with closed nozzles Page 128 fr om 136 4.3.2 .3.2 Measurement with half a filter disk (sealing membr ane missing). a.) W ith type 2057 no zzles b.) W ith type 2069 no zzles AC T NOM AC T Open …

100%1 / 136
Service manual vacuum check with closed nozzles
Page
127 from 136
4.3.2.3 Measurements with modifications
4.3.2.3.1 Measurement with filter disk covered (soiled).
a.) With type 2057 nozzles
b.) With type 2069 nozzles
ACT
Holding circuit > nominal.
Open value > nominal
NOM
NOM
ACT
No features or only
slight features
No features or only
slight features
Service manual vacuum check with closed nozzles
Page
128 from 136
4.3.2.3.2 Measurement with half a filter disk (sealing membrane missing).
a.) With type 2057 nozzles
b.) With type 2069 nozzles
ACT
NOM
ACT
Open value < nominal
Closed value < nominal
Holding circuit < nominal
NOM
Closed value < nominal
No features or only
slight features
Service manual vacuum check with closed nozzles
Page
129 from 136
4.3.2.3.3 Measurement without filter disk.
a.) With type 2057 nozzles
b.) With type 2069 nozzles
NOM
ACT
ACT
Open value < nominal
Closed value < nominal
Holding circuit < nominal
NOM
Closed value < nominal
No features or only