246 E690008 “Diode measurement is not availabl e unless probe shuts dow n. | Please shut probe then t ry again” [Cause] T he probe has to be closed to m easur e diode polarity. • The probe is not c losed. [Treatm ent] Pl…

100%1 / 1361
245
E690001 “CVS is not installed”
[Cause] CVS is not installed on the machine.
[Treatment] If CVS is actually installed on your machine, please enable the unit in the MS
parameter option.
E690002 “Impossible to switch to Resistance Measurement mode since the
probe has not shut down. | Please shut the probe down then try
again”
[Cause] In order to set measurement mode to the Resistance mode, the probe has to be
closed to adjust resistance.
The probe is not closed.
[Treatment] Please close the probe then set the measurement mode.
E690003 “Impossible to switch to Condenser Measurement mode since the
probe has not shut down. | Please shut the probe down then try
again”
[Cause] In order to set measurement mode to the Condenser mode, the probe has to be
closed to adjust capacity.
The probe is not closed.
[Treatment] Please close the probe then set the measurement mode.
E690004 “The measurement mode is not correct
[Cause] The measurement mode parameter is not correct.
[Treatment] Please re-start the application.
Please re-install or version-up the application.
If the error occurs even after the treatments above were performed, please
contact our technical service dept. for information.
E690005 “Parameter to control probe is not correct”
[Cause] The parameter to control probe is not correct.
[Treatment] Please re-start the application.
Please re-install or version-up the application.
If the error occurs even after the treatments above were performed, please
contact our technical service dept. for information.
E690006 “Probe has to be shut down to measure resistance. | Please shut
probe then try again”
[Cause] In order to measurement resistance, the probe has to be closed.
The probe is not closed.
The probe is not closed.
[Treatment] Please close the probe then measure the resistance.
E690007 “Probe has to be shut down to measure condenser. | Please shut
probe then try again”
[Cause] In order to measurement condenser capacity, the probe has to be closed.
The probe is not closed.
The probe is not closed.
[Treatment] Please close the probe then measure the condenser capacity.
246
E690008 “Diode measurement is not available unless probe shuts down. |
Please shut probe then try again”
[Cause] The probe has to be closed to measure diode polarity.
The probe is not closed.
[Treatment] Please close the probe then try the measurement.
E690009 “The measurement is impossible since measurement mode is not
set to Resistance measurement. | Please switch to the Resistance
measurement mode then try again”
[Cause] In order to measure resistance, the measurement mode has to be the Resistance
measurement mode.
Measurement mode is not set to the Resistance measurement mode.
[Treatment] Please set the measurement mode to the Resistance measurement mode and try
again.
E690010 “The measurement is impossible since measurement mode is not
set to Condenser measurement. | Please switch to the Condenser
measurement mode then try again”
[Cause] In order to measure condenser capacity, the measurement mode has to be the
Condenser measurement mode.
Measurement mode is not set to the Condenser measurement mode.
[Treatment] Please set the measurement mode to the Condenser measurement mode and try
again.
E690011 “The measurement is impossible since measurement mode is not
set to Diode measurement. | Please switch to the Diode
measurement mode then try again”
[Cause] In order to measure diode polarity, the measurement mode has to be the Diode
measurement mode.
Measurement mode is not set to the Diode measurement mode.
[Treatment] Please set the measurement mode to the Diode measurement mode and try
again
E690012 “The measurement is impossible since the gotten measurement
mode is set to Resistance measurement”
[Cause] The measurement mode gotten from the tester is not the Resistance
measurement.
The specified measurement mode and the measurement mode gotten from the
tester are clashing with each other.
The Component Verification is broken.
[Treatment] If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690013 “The measurement is impossible since the gotten measurement
mode is set to Condenser measurement”
[Cause] The measurement mode gotten from the tester is not the Condenser
measurement.
The specified measurement mode and the measurement mode gotten from the
247
tester are clashing with each other.
The Component Verification is broken.
[Treatment] If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690014 “The measurement is impossible since the gotten measurement
mode is set to Diode measurement
[Cause] The measurement mode gotten from the tester is not the Diode measurement.
The specified measurement mode and the measurement mode gotten from the
tester are clashing with each other.
The Component Verification is broken.
[Treatment] If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690015 “The resistance when zero-adjusting is abnormal. | Please check if
probe short-circuits or not”
[Cause] The resistance that was gotten from the tester when adjusting the resistance is
irregular.
The probe does not short-circuit.
The Component Verification is broken.
An obstruction that has resistance is caught in the probe.
The Component Verification is broken.
[Treatment] Please check and see if the probe short-circuits.
If the error occurs when the probe does not have any problems, please turn
off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690016 “A time-out occurred while reading resistance in”
[Cause] The tester does not start outputting data when starting to measure the resistance.
Component Verification is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690017 “A time-out occurred while reading volume in”
[Cause] The tester does not start outputting data when starting to measure the condenser
capacity.
Component Verification is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.