E690008 “Diode measurement is not available unless probe shuts down. |
Please shut probe then try again”
[Cause] The probe has to be closed to measure diode polarity.
The probe is not closed.
[Treatment] Please close the probe then try the measurement.
E690009 “The measurement is impossible since measurement mode is not
set to Resistance measurement. | Please switch to the Resistance
measurement mode then try again”
[Cause] In order to measure resistance, the measurement mode has to be the Resistance
measurement mode.
Measurement mode is not set to the Resistance measurement mode.
[Treatment] Please set the measurement mode to the Resistance measurement mode and try
again.
E690010 “The measurement is impossible since measurement mode is not
set to Condenser measurement. | Please switch to the Condenser
measurement mode then try again”
[Cause] In order to measure condenser capacity, the measurement mode has to be the
Condenser measurement mode.
Measurement mode is not set to the Condenser measurement mode.
[Treatment] Please set the measurement mode to the Condenser measurement mode and try
again.
E690011 “The measurement is impossible since measurement mode is not
set to Diode measurement. | Please switch to the Diode
measurement mode then try again”
[Cause] In order to measure diode polarity, the measurement mode has to be the Diode
measurement mode.
Measurement mode is not set to the Diode measurement mode.
[Treatment] Please set the measurement mode to the Diode measurement mode and try
again
E690012 “The measurement is impossible since the gotten measurement
mode is set to Resistance measurement”
[Cause] The measurement mode gotten from the tester is not the Resistance
measurement.
The specified measurement mode and the measurement mode gotten from the
tester are clashing with each other.
The Component Verification is broken.
[Treatment] If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690013 “The measurement is impossible since the gotten measurement
mode is set to Condenser measurement”
[Cause] The measurement mode gotten from the tester is not the Condenser
measurement.
The specified measurement mode and the measurement mode gotten from the
247
tester are clashing with each other.
The Component Verification is broken.
[Treatment] If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690014 “The measurement is impossible since the gotten measurement
mode is set to Diode measurement
[Cause] The measurement mode gotten from the tester is not the Diode measurement.
The specified measurement mode and the measurement mode gotten from the
tester are clashing with each other.
The Component Verification is broken.
[Treatment] If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690015 “The resistance when zero-adjusting is abnormal. | Please check if
probe short-circuits or not”
[Cause] The resistance that was gotten from the tester when adjusting the resistance is
irregular.
The probe does not short-circuit.
The Component Verification is broken.
An obstruction that has resistance is caught in the probe.
The Component Verification is broken.
[Treatment] Please check and see if the probe short-circuits.
If the error occurs when the probe does not have any problems, please turn
off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690016 “A time-out occurred while reading resistance in”
[Cause] The tester does not start outputting data when starting to measure the resistance.
Component Verification is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690017 “A time-out occurred while reading volume in”
[Cause] The tester does not start outputting data when starting to measure the condenser
capacity.
Component Verification is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
248
E690018 “A time-out occurred while reading voltage in”
[Cause] The tester does not start outputting data when starting to measure the voltage.
Component Verification is broken.
[Treatment] Please turn off the power supply and re-boot your machine.
If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690019 “The measured resistance does not get stable. | Please check the
probe status”
[Cause] The resistance gotten from the tester does not get stable.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690020 “The measured volume does not get stable. | Please check the
probe status”
[Cause] The condenser capacity gotten from the tester does not get stable.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690021 “The measured voltage does not get stable. | Please check the
probe status”
[Cause] The voltage gotten from the tester does not get stable.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690022 “Something is wrong with the measured resistance”
[Cause] The resistance gotten from the tester is irregular.
The object(component) of the measurement is not contacting the probe well.
Component Verification is broken.
[Treatment] Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.