KE2010.Instruction Manual.Ver.2.01,Rev.08 - 第1331页
248 E690018 “A time-out occurred w hile reading voltage i n” [Cause] T he tester does not star t outputting data when starting to m easure the voltage. • Component Verif ication is br oken. [Treatm ent] • Please turn of …
247
tester are clashing with each other.
• The Component Verification is broken.
[Treatment] • If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690014 “The measurement is impossible since the gotten measurement
mode is set to Diode measurement”
[Cause] The measurement mode gotten from the tester is not the Diode measurement.
The specified measurement mode and the measurement mode gotten from the
tester are clashing with each other.
• The Component Verification is broken.
[Treatment] • If the error occurs even after a re-try, please turn off the power supply and
re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690015 “The resistance when zero-adjusting is abnormal. | Please check if
probe short-circuits or not”
[Cause] The resistance that was gotten from the tester when adjusting the resistance is
irregular.
• The probe does not short-circuit.
• The Component Verification is broken.
• An obstruction that has resistance is caught in the probe.
• The Component Verification is broken.
[Treatment] • Please check and see if the probe short-circuits.
• If the error occurs when the probe does not have any problems, please turn
off the power supply and re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690016 “A time-out occurred while reading resistance in”
[Cause] The tester does not start outputting data when starting to measure the resistance.
• Component Verification is broken.
[Treatment] • Please turn off the power supply and re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690017 “A time-out occurred while reading volume in”
[Cause] The tester does not start outputting data when starting to measure the condenser
capacity.
• Component Verification is broken.
[Treatment] • Please turn off the power supply and re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
248
E690018 “A time-out occurred while reading voltage in”
[Cause] The tester does not start outputting data when starting to measure the voltage.
• Component Verification is broken.
[Treatment] • Please turn off the power supply and re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690019 “The measured resistance does not get stable. | Please check the
probe status”
[Cause] The resistance gotten from the tester does not get stable.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690020 “The measured volume does not get stable. | Please check the
probe status”
[Cause] The condenser capacity gotten from the tester does not get stable.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690021 “The measured voltage does not get stable. | Please check the
probe status”
[Cause] The voltage gotten from the tester does not get stable.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690022 “Something is wrong with the measured resistance”
[Cause] The resistance gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
249
E690023 “Something is wrong with the measured volume”
[Cause] The condenser capacity gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690024 “Something is wrong with the measured voltage”
[Cause] The voltage gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690025 “Something is wrong with the measured resistance(unit)”
[Cause] The unit for the resistance gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690026 “Something is wrong with the measured volume(unit)”
[Cause] The unit for the condenser capacity gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690027 “Something is wrong with the measured voltage(unit)”
[Cause] The unit for the condenser capacity gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.