249 E690023 “Somethi ng is w rong w ith the measured volume” [Cause] T he condens er capac ity gotten fr om the tester is irregular. • The objec t(com ponent) of the m easurem ent is not contacting the pr obe well. • Com…
248
E690018 “A time-out occurred while reading voltage in”
[Cause] The tester does not start outputting data when starting to measure the voltage.
• Component Verification is broken.
[Treatment] • Please turn off the power supply and re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690019 “The measured resistance does not get stable. | Please check the
probe status”
[Cause] The resistance gotten from the tester does not get stable.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690020 “The measured volume does not get stable. | Please check the
probe status”
[Cause] The condenser capacity gotten from the tester does not get stable.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690021 “The measured voltage does not get stable. | Please check the
probe status”
[Cause] The voltage gotten from the tester does not get stable.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690022 “Something is wrong with the measured resistance”
[Cause] The resistance gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
249
E690023 “Something is wrong with the measured volume”
[Cause] The condenser capacity gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690024 “Something is wrong with the measured voltage”
[Cause] The voltage gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690025 “Something is wrong with the measured resistance(unit)”
[Cause] The unit for the resistance gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690026 “Something is wrong with the measured volume(unit)”
[Cause] The unit for the condenser capacity gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690027 “Something is wrong with the measured voltage(unit)”
[Cause] The unit for the condenser capacity gotten from the tester is irregular.
• The object(component) of the measurement is not contacting the probe well.
• Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
250
E690028 “Resistance can no be measured. Please check if probe shuts
down properly”
[Cause] The tester overflowed data when measuring the resistance.
• The object(component) of the measurement is not contacting the probe.
• The contact between the object(component) of the measurement and the
probe is not good.
• The resistance of the object(component) of the measurement is exceeding the
measurement range.
• The Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• Please check the object.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690029 “Volume can not be measured. Please check if probe shuts down
properly”
[Cause] The tester overflowed data when measuring the condenser capacity.
• The object(component) of the measurement is not contacting the probe.
• The contact between the object(component) of the measurement and the
probe is not good.
• The resistance of the object(component) of the measurement is exceeding the
measurement range.
• The Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• Please check the object.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E690030 “Voltage can not be measured. Please check if probe shuts down
properly”
[Cause] The tester overflowed data when measuring the diode polarity.
• The object(component) of the measurement is not contacting the probe.
• The contact between the object(component) of the measurement and the
probe is not good.
• The resistance of the object(component) of the measurement is exceeding the
measurement range.
• The Component Verification is broken.
[Treatment] • Please check and see if the object is contacting the probe well. If not, please
adjust the position of the object to contact the probe properly.
• Please check the object.
• If the error occurs when the contact is good, there is a possibility that the
Component Verification unit is broken. Please contact our technical service
dept. for information.
E691001 “Input/Output buffer status of CCL-48M does not get ready”
[Cause] The input/output buffer status of the backup table counter(CCL-48M) does not get
ready.
• Backup table counter is broken.
[Treatment] • Please turn off the power supply and re-boot your machine.
• If the error occurs even after the re-boot, there is a possibility that the Safety
board is broken. Please contact our technical service dept. for information.