152 Exit the system. Contac t our servic e engineer or the agency. E605231 “Axis Driver: Support table (right) status error.” [Cause] The ax is m ove was block ed during or at the s tart of the operation. An initializati…
151
Exit the system. Contact our service engineer or the agency.
E605131 “Axis Driver: Support table status error.”
[Cause]
The axis move was blocked during or at the start of the operation.
An initialization error or alarm may have been generated for the axis.
This error can also be displayed against errors for other axes.
Or, the XMP board may have failed.
[Treatment]
Press the emergency stop switch, and then return the axes to the home position again.
If the error persists, exit the system and contact our service engineer or the agency.
E605132 “Axis Driver: Support table failed motion start.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605133 “Axis Driver: Support table failed motion stop.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605210 “Axis Driver: Support table (right) parameter error.”
[Cause]
An internal data error occurred in the system.
The combination of the system software and machine being used may be incorrect.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605212 “Axis Driver: Support table (right) failed originating.”
[Cause]
Failed to return to the home position. The possible causes are as follows:
1. Cannot turn the power ON.
2. The motor cannot start, or it rotate in the reverse direction.
3. The home sensor cannot be captured, or its state does not change.
4. The present position cannot be acquired.
[Treatment]
Exit the system. Check the conditions of the boards, connections, sensors and other possible
factors.
If the error persists, contact our service engineer or the agency.
E605213 “Axis Driver: Support table (right) error. Servo power is OFF.”
[Cause]
The power OFF was detected during or at the start of the operation.
An event that causes the servo-OFF may have occurred such as; the detection by the limit
sensor/area sensor/feeder float detector, or pressing of the emergency stop switch. Or, other
axis may have failed and resulted in cutting of the power.
[Treatment]
Remove the cause and return the axes to the home position.
If the error persists, exit the system and contact our service engineer or the agency.
E605230 “Axis Driver: XMP error occurred on right support table.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
152
Exit the system. Contact our service engineer or the agency.
E605231 “Axis Driver: Support table (right) status error.”
[Cause]
The axis move was blocked during or at the start of the operation.
An initialization error or alarm may have been generated for the axis.
This error can also be displayed against errors for other axes.
Or, the XMP board may have failed.
[Treatment]
Press the emergency stop switch, and then return the axes to the home position again.
If the error persists, exit the system and contact our service engineer or the agency.
E605232 “Axis Driver: Support table (right) failed motion start.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605233 “Axis Driver: Support table (right) failed motion stop.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605310 “Axis Driver: AWC parameter error.”
[Cause]
An internal data error occurred in the system.
The combination of the system software and machine being used may be incorrect.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605312 “Axis Driver: AWC failed originating.”
[Cause]
Failed to perform a return to home the position on the AWC. The possible causes are as
follows:
1. Cannot turn the power ON.
2. The motor cannot start, or it rotate in the reverse direction.
3. The home sensor cannot be captured, or its state does not change.
4. The present position cannot be acquired.
[Treatment]
Exit the system. Check the conditions of the boards, connections, sensors and other possible
factors.
If the error persists, contact our service engineer or the agency.
E605313 “Axis Driver: AWC error. Servo power is OFF.”
[Cause]
The power OFF was detected during or at the start of the operation.
An event that causes the servo-OFF may have occurred such as; the detection by the limit
sensor/area sensor/feeder float detector, or pressing of the emergency stop switch. Or, other
axis may have failed and resulted in cutting of the power.
[Treatment]
Remove the cause and return the axes to the home position.
If the error persists, exit the system and contact our service engineer or the agency.
E605330 Axis Driver: XMP error occurred on AWC.
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
153
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605331 “Axis Driver: AWC status error.”
[Cause]
The axis move was blocked during or at the start of the operation.
An initialization error or alarm may have been generated for the axis.
This error can also be displayed against errors for other axes.
Or, the XMP board may have failed.
[Treatment]
Press the emergency stop switch, and then return the axes to the home position again.
If the error persists, exit the system and contact our service engineer or the agency.
E605332 “Axis Driver: AWC failed motion start.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605333 “Axis Driver: AWC failed motion stop.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605510 “Axis Driver: Center motor parameter error.”
[Cause]
An internal data error occurred in the system.
The combination of the system software and machine being used may be incorrect.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605513 “Axis Driver: Center motor error. Servo power is OFF.”
[Cause]
The power OFF was detected during or at the start of the operation.
An event that causes the servo-OFF may have occurred such as; the detection by the limit
sensor/area sensor/feeder float detector, or pressing of the emergency stop switch. Or, other
axis may have failed and resulted in cutting of the power.
[Treatment]
Remove the cause and return the axes to the home position.
If the error persists, exit the system and contact our service engineer or the agency.
E605530 “Axis Driver: XMP error occurred on center motor.”
[Cause]
An internal error occurred in the system.
The XMP board or system HDD may have failed.
[Treatment]
Exit the system. Contact our service engineer or the agency.
E605531 “Axis Driver: Center motor status error.”
[Cause]
The axis move was blocked during or at the start of the operation.
An initialization error or alarm may have been generated for the axis.
This error can also be displayed against errors for other axes.
Or, the XMP board may have failed.
[Treatment]
Press the emergency stop switch, and then return the axes to the home position again.
If the error persists, exit the system and contact our service engineer or the agency.