2500_Users_Manual - 第146页

Operation 4-32 ProMa ster 25 00 U ser Ma nual Alignin g a Devic e to a DIP/SOIC/32-pi n PLCC Programm ing Mod ule When you process a device with a rectangular shaped body (DIP, SOIC, or 32-pin PLCC), the 2500 prompts you…

100%1 / 448
Operation
ProMaster 2500 User Manual 4-31
4. Press
START
. The beam rotates the device 90
°
and moves it over the
programming module. Push the beam down manually (on either side
of the beam limit bar), and ensure that the device moves into the
center of the programming module’s block.
When the alignment is correct, press and hold
D
on the 2500
keyboard to check the device’s insertion.
Repeat steps 1 through 4 until the device moves into the center of the
programming module and does not rub one set of contacts more than
another.
5. When the alignment is correct, press
START
to continue running the
Task.
Figure 4-25
Device too Close to the Front
Figure 4-26
Device too Close to the Back
If the device was too close to the FRONT, move the beam to the RIGHT.
1860-2
X4
X1
X3
FRONT
If the device was too close to the BACK, move the beam to the LEFT.
1861-2
X4
X1
X3
BACK
Operation
4-32 ProMaster 2500 User Manual
Aligning a Device to a DIP/SOIC/32-pin PLCC
Programming Module
When you process a device with a rectangular shaped body (DIP, SOIC,
or 32-pin PLCC), the 2500 prompts you to align the first device in the job
run. Follow these steps to align the beam to a device in the input track.
The track width should be adjusted for the device before you begin this
procedure (see page 4-12).
Note: This alignment procedure assumes that the devices are inserted in the
input track with
pin 1 oriented to the right
when you are standing in
front of the 2500.
1. Insert a tube of devices into the input track and close the hood.
2. Start the new Task. The beam moves over the first device and pauses.
The 2500 displays:
3. Press
and
to center the chuck over the device (left-to-right).
Pressing the arrow key once moves the beam a small step in that
direction.
Note: You can change beam alignment while a Task is running. When the beam
hesitates above the device, press
STOP
and use the front panel arrow keys.
4. Press and hold
D
on the 2500’s keyboard to lower the beam and
check the position of the chuck on the device. For a slower insertion,
lower the beam manually by pushing directly on the top of the beam
assembly, on either side of the beam’s limit bar (the limit bar is shown
in Figure 1-7). Adjust the left-to-right position if necessary. When the
beam is centered, press
START
.
5. The beam picks up the device, moves it over the programming
module, and pauses before inserting the device. The 2500 displays:
6. Align the device pins to the module’s contacts using the
and
keys to move the beam with the device. The left-most pins of the
device should line up with the left-most contacts on the module.
Press and hold
D
key to lower the device into the programming
module and check the alignment.
Note: Do not press the arrow keys while the device is on the programming
module. Doing so may cause the beam to break its vacuum hold and drop
the device.
PROGRAM/TEST ONLY
USE ARROW KEYS TO ALIGN BEAM WITH
DEVICE CENTER. PRESS [D] TO LOWER BEAM.
PRESS START TO CONTINUE.
PROGRAM/TEST ONLY
USE ARROW KEYS TO ALIGN DEVICE WITH
PROG. MODULE. PRESS [D] TO LOWER BEAM.
PRESS START TO CONTINUE.
Operation
ProMaster 2500 User Manual 4-33
If the device is not aligned properly, release the
D
key, use the arrow
keys to change the position, and lower the beam into the
programming module to check it again.
When you are satisfied with the left-to-right alignment, check the
front-to-back alignment. See if the device is pushing either the front
or the back contact set much more than the other.
7. If both front and back contact sets are being contacted equally, press
START
to begin processing devices.
If the device is not centered front-to-back,
note
whether it is too close to
the
front
or the
back
, and follow the procedure below to readjust its
position.
Adjust Front-to-back
Position
To realign so the device is inserted into the center of the programming
module (front-to-back axis), follow the steps below.
1. Exit the Task or Kit, restart it, and prepare to check the alignment
again.
2. The beam will pause above the device and the 2500 displays:
Figure 4-27
DIP Device Alignment
PROGRAM/TEST ONLY
USE ARROW KEYS TO ALIGN BEAM WITH
DEVICE CENTER. PRESS [D] TO LOWER BEAM.
PRESS START TO CONTINUE.
1945-2