2500_Users_Manual - 第195页

Preventi ve Mai nten ance ProM aster 25 00 User Manua l 5-41 Note: The beam traverse motor (tests 1 and 2) may appear to stall or fail. This could be caused by a dirty or dry lead screw. Refer to page 5-63 for the lead s…

100%1 / 448
Preventive Maintenance
5-40 ProMaster 2500 User Manual
The solenoids perform the following operations (confirm that LEDs on
the solenoids turn on for the duration of each test):
Blower
— Turns on low air pressure to the output tracks. There is no
air on the input track.
Cutoff
— Turns low air pressure to the beam on and off. If you are
running all tests, perform the cutoff test first when you enter the
diagnostic tests. At the start of the diagnostic tests, the beam should
be up. Pressing
3
directs air to the beam, allowing it to rise.
Note: After you run this test you must press
RESET
to exit the solenoid tests
and then reselect the solenoid tests. If you do not press
RESET
, the Beam
Up/Down and the High Pressure tests will not operate correctly.
Beam Up/Down
Directs low air pressure to raise the beam.
Pressing the
4
key allows the beam to lower.
Blow Off
Turns on low air pressure to the chuck tip to assist in
releasing a device. Feel for air blowing out of the chuck tip.
High Pressure
Turns on high pressure. The beam will lower
quicker than the Beam Up/Down test, which uses low air pressure.
Vacuum
Turns on the vacuum generator. Place your finger over
the end of the chuck and feel the vacuum.
Clamp
Activates the programming module holding clamps. They
should both advance and clamp the module.
To exit the solenoid test and return to the Diagnostic menu, press
E
and
then
ENTER
. To exit the solenoid test and return to the 2500’s main
menu, press
RESET
.
Motor Test
To select Motor Test from the Diagnostics menu, press
3
. The 2500
displays:
WARNING:To perform the motor tests, the hood has to be raised for
better visibility. To avoid electrical shock or mechanical
injury, these tests should be performed only by a service
technician trained on electromechanical equipment.
The beam moves quickly. To avoid injury, keep your
hands, hair, and loose clothing away from the path of the
beam while running these tests.
Refer to the motor test table and Figure 5-17. Press the number for the
motor you wish to test, and then press
START
to activate the motor.
Press
CAL
to reverse the rotation direction.
PRESS NUMBER TO TEST MOTOR, E TO EXIT
1 - BEAM FORWARD 4 - TUBE INPUT
2 - BEAM REVERSE 5 - TUBE OUTPUT
3 - BEAM ROTATION 6 - LABELER
Preventive Maintenance
ProMaster 2500 User Manual 5-41
Note: The beam traverse motor (tests 1 and 2) may appear to stall or fail. This
could be caused by a dirty or dry lead screw. Refer to page 5-63 for the lead
screw cleaning and lubricating procedure. If this does not resolve the
problem, adjust the motor speed by entering the STOP command
LOWER CASE
+
M
, described in Appendix B.
CAUTION: Be sure to open the hood during traverse motor test 1 to
prevent the beam from hitting the hood shock mount.
Figure 5-17
Location of the Motors
Test No. Motor Description
1Beam Forward
2Beam Reverse
3 Beam (Head) Rotate
4Tube Input (Orbital)
5Tube Output (Orbital)
6Label Drive
1942-2
(5) TUBE OUTPUT
(1 + 2) BEAM
TRAVERSE
(6) LABELER
(Hidden)
(4) TUBE INPUT
(3) BEAM ROTATE
MOTOR ENCODER
Preventive Maintenance
5-42 ProMaster 2500 User Manual
Cycle Parts
This option inserts devices into the programming module without
programming them. It differs from the diagnostic tests because it allows
you to assess the following system operations under operating conditions
without having to program devices:
Beam movement
Vacuum generator functionality
High and low air pressure
Applying labels to devices
To select Cycle Parts, press
4
from the Diagnostic menu. The 2500
displays:
To Cycle Only
If you do not want to label devices, press
N
at the
DO YOU WANT TO
LABEL?
prompt. The 2500 displays:
Use the arrow keys to select an existing device or describe a custom
device, and press
ENTER
. Only eight selections appear on the display at
once; use
and
to scroll to the selections not visible on the display.
The 2500 displays:
Use the 2500’s arrow keys to move the pin 1 icon to correspond to the
location of pin 1 on the device in the input tube, and press
ENTER
. The
2500 displays:
Use the arrow keys to select the pin 1 position of the device as it goes into
the receiving tube. Press
ENTER
when you are satisfied with your
selection.
The next display prompts you to enter the number of parts per tube. The
default value appears in parentheses.
DO YOU WANT TO LABEL? (Y OR N)
CYCLE PARTS
CUSTOM PLCC 20 PLCC 28 PLCC 32
PLCC 44 PLCC 52 PLCC 68 PLCC 84
MOVE TO SELECTION THEN PRESS ENTER
1 <------- TEMPORARY FILE
SELECT POSITION OF PIN 1
AS IT COMES OUT OF THE TUBE.
USE ARROW KEYS THEN PRESS ENTER.
<------- TEMPORARY FILE
SELECT POSITION OF PIN 1
AS IT WILL GO INTO THE TUBE.
1 USE ARROW KEYS THEN PRESS ENTER.
CYCLE PARTS
ENTER NUMBER OF PARTS PER TUBE (13): _