SERVICE ACCESS
OIPT RECOMMENDS THAT AT LEAST 600mm
SERVICE ACCESS SPACE IS ALLOWED BETWEEN
ANY OBSTACLE (E.G. WALLS, PARTITIONS, ETC.)
AND SERVICEABLE ITEMS, E.G. THE POWER
DISTRIBUTION UNIT.
Fig 3: ICP 180 Layout (2-frame cluster upgradeable)
NOTE: For services panel details, see
Fig 4.
Installation Data (ICP 180)
Issue 5: June 05 Page 8 of 18 Printed: 22-Mar-06, 6:57

Installation Data= lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó mä~ëã~ä~ÄpóëíÉãNMM
PK= pÉêîáÅÉë=ÅçååÉÅíáçåë=
711
738.5
746
781
816
837
851
886
921
951
956
Distance
to floor
level
385 415 461.5
469
3/8“ SWAGELOK
3/8“ SWAGELOK
3/8“ SWAGELOK
1/4“ SWAGELOK
1/4“ SWAGELOK
1/4“ VCR
(Only used
if a gas ring
is fitted)
1/4“ VCR
TOP ELECTRODE
& AMU COOLING
OUTIN
OUT
IN
OUT
IN
OUTIN
RF GENERATOR COOLING
LOWER ELECTRODE
COOLING
TURBO COOLING
N2 IN
AIR IN
GAS RING
PROCESS
GAS
AIR OUT
CHAMBER BACKING
HELIUM IN
4 MM PUSH-FIT
LEGRIS
Distance to left-hand edge of frame
(viewed from rear of system)
LIQUID N2 IN
1/4” SWAGELOK
INSULATED
FITTING (CRYO
TABLE OPTION)
NITROGEN OUT
1/4” SWAGELOK
INSULATED
FITTING (CRYO
TABLE OPTION)
HELIUM SERVICE FOR
WAFER COOLING
(IF REQUIRED)
PLC SERVICE
ACCESS CONNECTOR
(THIS IS AN
ALTERNATIVE PC
COMMUNICATIONS
CONNECTION)
NOT
Fig 4: ICP 180 services panel
NOTE: The Automatic load lock services panel is similar to that shown above but only the
CHAMBER BACKING connector is used.
Installation Data (ICP 180)
Printed: 22-Mar-06, 6:57 Page 9 of 18 Issue 5: June 05

mä~ëã~ä~ÄpóëíÉãNMM= lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó Installation Data
POWER DISTRIBUTION
UNIT
(REMOVABLE COVER)
COOLING
FAN
SKT 7 SKT 10A
SKT 9
SKT 17SKT 16SKT 8
PUMP
PURGE
HEATER/
CHILLER
SKT 10
EARTH TERMINAL
SYSTEM POWER
CABLE GLAND
(CABLE LENGTH
= 4 METRES)
LOW VOLTAGE
FUSES
SKT 7
AUXILIARY
EMERGENCY OFF
SWITCH CONNECTOR
4-WAY BICC
PIN A EMO 1*
PIN B +24V
PIN C 0V
PIN D EMO 2*
(NOTE: EMO 1 & 2
MUST BE LINKED)
*EMO = EMERGENCY
OFF
2 AMP
FS1
+15V -15V +24V
FS2 FS3
2 AMP 2 AMP
1
2
3
4
5
6
PHASE 1
PHASE 2
PHASE 3
SPARE
SPARE
NEUTRAL
SUPPLY TO LOAD LOCK
ROTARY PUMP
SOCKET 10A
1
2
3
4
5
6
PHASE 1
PHASE 2
PHASE 3
SPARE
SPARE
NEUTRAL
SUPPLY TO ROTARY
PUMP 2
SOCKET 9
12
3 4
PHASE 1
PHASE 2
PHASE 3
SUPPLY TO ROTARY
PUMP 1
SOCKET 10
1
2 E
SUPPLY TO ROTARY
PUMP OIL FILTER
SOCKET 17
415V
Pin 1 Live
Pin 2 Neutral
Pin E Earth
208V
Pin 1 Phase 1
Pin 2 Phase 2
Pin E Phase 3
Fig 5: Electrical connections