OKN= oÉÅáêÅìä~íáçå=ï~íÉê=
WARNING
CHILLERS HAVE COMPONENTS, WHICH BECOME COLD ENOUGH TO CAUSE SERIOUS
INJURY. READ THE MANUFACTURER’S MANUALS BEFORE INSTALLING, OPERATING
OR MAINTAINING CHILLERS, AND ENSURE THAT ADEQUATE PROTECTIVE CLOTHING
IS WORN.
Recirculation water is used in systems where the user does not wish to consume water at a
high rate, or where it is wished to supply water to the system at a constant temperature.
Some systems contain aluminium in direct contact with coolant. These are the fçåÑ~Ä=RMM=
mäìë, the fçåÑ~Ä=PMM=mäìë, and mä~ëã~ä~Ä systems fitted with tables or turbomolecular pumps
with plain aluminium cooling channels. Certain aluminium chambers are also included.
However, the fçåÑ~Ä=PMM=mäìë does not contain aluminium in contact with water, unless it is
in the turbomolecular pump.
In these aluminium-containing systems, it is MANDATORY that the approved coolant is used
at all times. If the approved coolant and installation are not used, the system warranty will
not cover any damage caused by the coolant either directly or indirectly to the system or to
anything else.
It is the customer's responsibility to comply with the above requirement. If the customer has
any doubt as to whether their system falls into this category, they must contact OIPT for
further advice.
In systems that do not contain aluminium components, severe corrosion can still sometimes
occur: water will collect corrosive chemicals, and will be lacking in the oxygen that protects
stainless steel. We strongly recommend that the approved coolant is used. If an alternative
product or water is used, then it is entirely the customer’s responsibility to check for any signs
of corrosion or contaminated water. If the customer’s coolant has not been approved in
writing by OIPT, then the system warranty will not cover any damaged caused by the coolant
either directly or indirectly to the system or to anything else
1
.
The approved installation and coolant are shown in sub-section 2.1.1 and in Fig 2.1. Note that
the total water flow meter and the isolation valves are recommended for cooling monitoring
and ease of maintenance.
1
Customers should note, for example, that certain types of inhibitor will damage the pumps of heater/chiller units, and that the use
of municipal (drinking water) or deionized water can introduce harmful chemicals into the system. If plain water is used, distilled
water is by far the safest option.
Services Specifications
Printed: 6-Jan-06, 8:44 Page 5 of 22 Issue 16: April 05

lfmq=póëíÉãë lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== System Manual
DEDICATED
HEAT EXCHANGER
OR
CHILLER
PLASMALAB
OR
IONFAB
SYSTEM
ISOLATION
VALVES
10-MICRON METAL MESH FILTER
MAXIMUM PRESSURE DROP AT
RATED FLOW RATE 0.15 BAR (2.2 PSI)
TOTAL WATER FLOW
METER OF APPROPRIATE
RANGE
(FITTED WITH AN
EXTERNAL BYPASS
HAVING A CAPACITY OF
100% FLOW)
Fig 2.1: Recommended basic recirculation installation
OKNKN= j~åÇ~íçêó=péÉÅáÑáÅ~íáçåë=Ñçê=êÉÅáêÅìä~íÉÇ=ï~íÉê=ëóëíÉãë=
CAUTION
If clear (i.e. transparent) tubing is exposed to sunlight, algal growth can develop,
which can restrict coolant flow.
It is MANDATORY that clear tubing is not used in any part of the cooling system.
OIPT recommends the use of either black or dark green tubing.
CAUTION
It is the customer's responsibility not to exceed a pressure of 4.2 bar, or other
limit that has been set for the system. Exceeding this safe pressure may cause
irreparable damage to system components.
The water must be kept warm enough to prevent condensation on chamber surfaces and
outside system components. This applies to those parts of the system inside the clean room
and those parts in a service area. Condensation can damage components such as RF power
supplies, ferrofluidic seals and automatch units. Any damage so caused cannot be covered by
the system warranty.
Services Specifications
Issue 16: April 05 Page 6 of 22 Printed: 6-Jan-06, 8:44

System Manual lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== lfmq=póëíÉãë
Pressure: Adjustable 0.7 to 4.2 bar (10 to 60 psi). Chiller / heat exchanger
to be fitted with a bypass having a capacity of 100% of rated
flow.
Temperature range: See system installation data sheets.
Minimum flows: See system installation data sheets.
Cooling capacity: See system installation data sheets.
Coolant: Hexid A40 (OIPT Part No. G/WATER/SUN/007 for 15 litres)
2
. This
product is propylene glycol based, and is pre-diluted ready for
use. See sub-section 2.1 for the warranty impact of not using
this product.
Filtration: 10 micron metal mesh water filter. Maximum pressure drop
0.15 bar (2.2 psi) at rated flow. For example, filter element
Balston SMC-100-12-10.
OKO= qçí~ä=içëë=`ççäáåÖ=
Municipal (drinking) water may be used in total loss cooling of the system in situations
where clean water is freely available, but only if the water meets the specification in sub-
section 2.2.1. In case of any doubt, obtain the water specification from the water utility
company, and consult with OIPT. It is not practical to use total loss cooling where the water
temperature is critical.
PLASMALAB
OR
IONFAB
SYSTEM
ISOLATION
VALVES
FILTER
0.9 MICRON FULL FLOW
TOTAL WATER FLOW
METER OF APPROPRIATE
RANGE
DRAIN
Fig 2.2: Recommended basic total loss installation
2
To order replacement fluid, check the capacity of the model of heater/chiller to be fitted to the system. Also allow at least 5 litres
for the system and connecting lines, or more if necessary.
Services Specifications