Equipment Manual lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== mä~ëã~ä~Ä=f`m=NUM=
WARNING
BEFORE PROCEEDING WITH ANY MAINTENANCE WORK, READ SECTION 1 - HEALTH AND SAFETY.
SKQ= `Ü~åÖáåÖ=íÜÉ=f`mNUM=ÇáÉäÉÅíêáÅ=íìÄÉ=
This sub-section applies to ICP180 second series builds (‘floating top’ version) with the following part
numbers:
100-6-54 ICP 180-FLOAT.TOP ASY(QUARTZ)
100-6-54/1 ICP180-FLOAT.TOP-ANOD.(QUARTZ)
100-6-55 ICP180-FL0.TOP ANG.VP(QUARTZ)
100-6-55/1 ICP180-FL.TOP-AN.VP-ANOD QU'TZ
100-6-56 ICP180 FLOAT TOP ASSY ALUMINA
100-6-56/1 ICP180 FLOAT TOP ASY ANOD ALUM
100-6-57 ICP180FLO.TOP ANG V/P(ALUMINA)
100-6-57/1 ICP180 FLO.TOP V/P ANOD ALUM.
Refer to assembly drawing MA00A23815
These instructions are given to assist personnel who are skilled in the maintenance of vacuum process
equipment. The task includes operating the vacuum equipment and mechanical assembly. There is some
lifting from the top of the equipment which may be easier with two persons.
Parts required:
MD00C18541 QUARTZ TUBE (ICP SOURCE)
Or
MD00C18542 ALUMINA TUBE (ICP SOURCE)
G/SEL/O-K/264 KALREZ O'RING BS 264 X2
G/GAS/FIT/614C ¼ INCH VCR GASKET X1
Safety
Risks associated with this procedure include:
Toxic gas. Pump the system long enough to reduce residual toxic gases to an acceptable
level. If halogen gases have been used in the chamber, run a plasma clean (1 hour 10
mTorr O2 50 sccm, 1000W ICP, 50 W bias or similar process) then pump for at least one
hour. Follow local practice concerning risk assessment and personal protective equipment.
OIPT practice is to use chemical filter masks when working on halogen process chambers.
Glass splinters. Quartz tubes may have sharp edges. Damaged tubes make leave glass
splinters. Take due care.
Implosion. There is a small risk of catastrophic tube failure, especially the first time it is
put under vacuum. Put the shielding in place before pumping the system. Always vent the
system before removing the covers.
RF radiation. The ICP180 coil is connected to a high power RF source which can cause
burns or interfere with other equipment. Make sure the RF generator is switched off
before removing any covers. After completing the following procedure, perform an RF
leakage test as detailed in Appendix A of this system manual.
ICP 180 Source
Printed: 18-Jan-06, 8:44 Page 19 of 26 Issue 4: January 06
mä~ëã~ä~Ä=f`m=NUM lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== Equipment Manual=
WARNING
BEFORE PROCEEDING WITH ANY MAINTENANCE WORK, READ SECTION 1 - HEALTH AND SAFETY.
=
Procedure
1) Clean and pump the process chamber as recommended above
2) Vent the chamber
3) If an endpoint camera is fitted, remove the camera
4) Remove the outer cover of the ICP180
5) Disconnect the ¼ inch VCR joint feeding gas into the top of the ICP180
6) Free the water cooling tubes or turn off the water and disconnect the tubes to allow the
ICP180 top vacuum lid to be removed
7) Undo the 6 bolts securing the top vacuum lid and remove the lid, exposing the dielectric tube
8) It will probably be necessary to open one or more of the inner RF shielding panels to gain
access to the cooling clamps at the top and bottom of the ICP dielectric tube. Slacken the
cooling clamps until it is possible to withdraw the old ICP tube.
Notes:
a) The tube may stick to the lower vacuum O-ring.
b) It should not be necessary to remove the electrostatic screen when changing
the tube.
9) Clean the lower and upper O-ring seal areas thoroughly: take care to remove any sharp
particles.
10) Inspect and if necessary replace the upper and lower O-ring seals
11) Inspect the new ICP tube for defects, especially chips or cracks at the vacuum sealing faces. Do
not install a damaged tube.
12) Install the tube.
13) Tighten the cooling clamps. Do not use excessive force. The lower clamp should be firm, and
the upper clamp should be much looser. This allows the electrostatic shield to move freely
upwards by thermal expansion. The electrostatic screen must not be bowed out too much.
Keep a clearance of at least 12 mm between the screen and the RF coil.
14) Reassemble the ICP 180 source
15) When the outer cover is in place, pump down and perform a leak check.
16) Perform an RF leakage test. See Appendix A of your Plasmalab system manual for details.
17) Dispose of the old tube in accordance with local regulations.
ICP 180 Source
Issue 4: January 06 Page 20 of 26 Printed: 18-Jan-06, 8:44
Equipment Manual lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== mä~ëã~ä~Ä=f`m=NUM=
WARNING
BEFORE PROCEEDING WITH ANY MAINTENANCE WORK, READ SECTION 1 - HEALTH AND SAFETY.
TK=qêçìÄäÉëÜççíáåÖ=
TKN= `ìëíçãÉê=pìééçêí=c~ÅáäáíáÉë=
Oxford Instruments Plasma Technology has global Customer Support Facilities to provide a co-
ordinated response to customer’s queries. All queries are recorded on our Support Database
and dealt with as quickly as possible. If we are not able to answer the query immediately, we
will contact you as soon as possible.
Before contacting a Customer Support Facility, please ensure that you have referred to the
appropriate section of your system manual, OEM manuals and electrical drawings.
Please direct all queries through your nearest support facility (see below) and have the
following details available.
System Type, e.g. mä~ëã~ä~ÄpóëíÉãNMM=etc.
Works Order No. – This can be found on the front cover of your system manual.
Contact Information – Your name, Company and how we can contact you.
Details of your query – nature of the problem, part numbers of spares required, etc.
You can contact us via any of the following:
China
Beijing
Tel: +86 106518 8160
Email:
pt.support@oichina.cn
Shanghai
Tel: +8621 6360 8530
Email:
pt.support@oichaina.cn
Germany
Tel: +49 6122 937 161
Email:
info@oxford.de
Japan
Tel: +81 3 5245 3591
Email:
oikkpt@oxinst.co.jp
Singapore
Tel: +65 6337 6848
Email: pt.support@oxford-instruments.com.sg
Taiwan
Tel: +886 3 5788696
Email:
pt.support@oxford-instruments.com.tw
UK
Customer Support Hotline: +44(0) 1934 837070
(0800 – 2000 UK local time)
Fax: +44(0) 1934 837071
Email:
pt.support@oxinst.co.uk
USA
Concord, MA
Tel: +1 978 9933 ext. 444
Toll Free: +1 800 447 4717 ext. 444
Fax: +1 978 369 8287
Email:
support.pt@ma.oxinst.com
ICP 180 Source
Printed: 18-Jan-06, 8:44 Page 21 of 26 Issue 4: January 06