Micron Maintenance Manual - 第78页

3 VACUUM FILTRATION UNITS 3.1 FILTER REPLACEMENT 78 MAINTENANCE MANUAL MICRON 11/2017 7. To separate the pre-filter unit, (3) from the carbon filter unit (4), pull the pre-filter upwards us- ing the pre-filter frame. 3 4…

100%1 / 92
3 VACUUM FILTRATION UNITS
3.1 FILTER REPLACEMENT
MAINTENANCE MANUAL MICRON 11/2017 77
5. The back of the carbon filter unit is secured beneath a filter securing bracket. To remove the
filter, lift the front and pull the filter out of the printer.
6. Turn the filter unit upside down.
3 VACUUM FILTRATION UNITS
3.1 FILTER REPLACEMENT
78 MAINTENANCE MANUAL MICRON 11/2017
7. To separate the pre-filter unit, (3) from the carbon filter unit (4), pull the pre-filter upwards us-
ing the pre-filter frame.
3
4
8. Replace the carbon\pre-filter as required and refit the pre-filter into the carbon filter unit.
3 VACUUM FILTRATION UNITS
3.1 FILTER REPLACEMENT
MAINTENANCE MANUAL MICRON 11/2017 79
9. Refit the carbon filter unit to the vacuum filtration unit.
Note
When refitting the carbon filter unit (5) to the base unit (7), ensure that the rear of the carbon
filter unit is located beneath the filter securing the bracket (6).
5
6
7
10. Secure the carbon filter unit using the filter retaining clip.
11. Refit and secure the air duct to the carbon filter unit.
12. Refit the panel removed in Step 2.