EUKYX
1-36199-4100
4.6 Cutter Section
4.6.3 Parts replacement of the cutter unit
WARNING
Pay special attention to the cutter blade when maintaining.
•
You may cut your hand or nger with a cutter blade.
Make sure to wear gloves and maintain with caution.
•
Make sure to attach a cutter blade cover when
maintaining a cutter for protection.
Replacing fluorine Sheet
Cycle: 1 year (Required Time : 35 minutes)
(1) Remove the cutter from the main unit.
Refer to “4.6.1 Cutter Unit cleaning” for the procedure.
(2) Remove the slope from the cutter unit.
Refer to “4.6.1 Cutter Unit cleaning” for the procedure.
(3) Remove 5 hex socket head cap screws mounting the fluorine sheet assembly from the cutter
unit.
Hex socket head cap screws
Fluorine sheet assembly
Removed fluorine sheet assembly
F4B23A
Procedure

EUKYX
1-37199-4100
4.6 Cutter Section
(4) Remove 5 hex socket head cap screws from the fluorine sheet assembly
Fluorine sheet
Mounting plate
Take special care for hole
positions of each assembly.
Mounting bracket
Removed fluorine sheet assembly
Hex socket head cap screws
F4B23B
(5) Replace the fluorine sheet.
Assemble the fluorine sheet assemblies and return the cutter unit to the main body.
Make sure to assemble the fluorine sheet assemblies aligning the taps and the hole positions of
each assembly.
Note

EUKYX
1-38199-4100
4.6 Cutter Section
Replacing flat ring
Cycle: 1 year (Required Time : 50 minutes)
(1) Remove the cutter unit from the main unit.
Refer to “4.6.2 Cutter Unit Lubrication” for the procedure.
(2) Remove the upper slope from the cutter unit.
Refer to “4.6.2 Cutter Unit Lubrication” for the procedure.
(3) Remove the rear cover.
Refer to “4.6.2 Cutter Unit Lubrication” for the procedure.
(4) Remove 5 hex socket head cap screws mounting the fluorine sheet assembly from the cutter
unit. Remove the fluorine sheet assembly.
Refer to “Replacing fluorine sheet” in “4.6.3 Parts replacement of the cutter unit” for the
procedure.
(5) Remove 2 truss head screws mounting the polycarbonate plate and the cable.
Remove the polycarbonate plate.
Polycarbonate plate
Truss head screws
F4A81A
(6) Remove the hex socket head cap screws (3 each on left and right) mounting the side
brackets and the plates. Remove the side brackets and the plates.
Side bracket
Side bracket and plate
Hex socket head cap screws
Side plate
F4A81B
Procedure