ACT - Accuracy Check Tool / User Manual 09/2015 Edition
12
3 Preparatory Work
3.1 General Hardware Preparations
Ensure that the glass components are placed correctly in the divisions. The components should
be placed face downwards into the tray, see Figure 3-1.
Figure 3-1: Arrangement of the glass components; vacuum nozzle HANDI-VAC
NOTICE
Handling the glass components
Do not touch the underside of the glass surface where the QFP structure is applied
when you handle the glass components.
Use the "HANDI VAC HV-4-ESD" vacuum nozzle [00334547-xx] to suck the
components on the upside where name and serial number are imprinted.
Feeder for ceramic components (Cerampads)
Figure 3-2: Feeder for ceramic components (S-feeder in figure)
Key:
(1) Ceramic components (Cerampads) (2) Feeder
ACT - Accuracy Check Tool / User Manual 09/2015 Edition
13
NOTICE
Cleanliness of plate and glass components
To ensure proper measurement, make sure that the measuring plate and the
components are free of all contaminants (particularly the fiducials in the corners of the
glass components).
The glass components should be cleaned with glass cleaning agent (e.g. Viss universal
from Lever)
Clean the measuring plate and apply the double-sided adhesive foil as described below.
NOTICE
Applying the foil without air bubbles
Make sure that the foil is applied without air bubbles, particularly in the area around
the line markings on the measuring plate.
Applying the foil
Remove the narrow strip of protective paper from the foil (Figure 3-3).
Figure 3-3: Adhesive foil with protective paper
Key:
(1) Protective sheet on adhesive
foil
(2) Adhesive foil
ACT - Accuracy Check Tool / User Manual 09/2015 Edition
14
Apply the foil.
Start at the wide side of the measuring plate (Figure 3-4).
Smooth any air bubbles out towards the side.
Figure 3-4: ACT measuring plate with adhesive foil
Key:
(1) Protective sheet on
adhesive foil
(2) ACT measuring plate
Apply the foil as displayed in Figure 3-5.
Figure 3-5: ACT measuring plate with adhesive foil
Key:
(1) Protective sheet on
adhesive foil
(2) ACT measuring plate