Conguring the Trigger Settings
The trigger function allows you to start and stop measurements using specic signals.
When recording is started by specic signals, it is called “The instrument is triggered.”
In this chapter, the mark
represents the point the start trigger is activated.
In the descriptions of each trigger source, the mark
represents a point each trigger condition is
satised and a point each trigger is generated.
When the real-time save is set to [On], you cannot use the trigger function.
> [Trigger] > [Common]
Operation available on the [Trigger] screen
Conguring the logic trigger settings
(p. 114)
Logic-trigger satisfaction conditions
Trigger lter
Trigger pattern
Conguring the analog trigger settings
(p. 106)
Level trigger
Window trigger
Period trigger
Glitch trigger
Voltage drop trigger
Conguring the trigger setting
Trigger timing (p. 100)
Pre-Trigger (p. 102)
Trigger satisfaction condition (logical AND or OR
operation) among the trigger sources (p. 105)
Conguring the interval trigger settings
(p. 116)
Conguring the external trigger settings
(p. 118)
Forcible trigger (p. 118)
Trigger output (p. 213)
Conguring the trigger source settings
5 Conguring the Trigger Settings
5
Conguring the Trigger Settings
98
Trigger Setting Procedure
5.1 Trigger Setting Procedure
The instrument is triggered based on trigger logical-conditions (logical AND or OR operation)
among trigger sources except for the forcible trigger. (p. 105)
When triggered, the instrument outputs the TRIG OUT signal from the external control terminal.
(p. 213)
Congure the trigger function setting
Enables the trigger function. (p. 99)
Conguring the trigger timing setting
Specify the start timing of recordings
controlled by triggers.
(p. 100)
Conguring the pre-trigger settings
Specify a recording length that precedes
the start trigger point.
(p. 102)
Setting trigger logical-conditions (AND or
OR operation) among trigger sources
Setting trigger logical-conditions (AND or
OR operation) among the analog, logic,
external, and interval triggers to trigger the
instrument.
(p. 105)
Conguring the trigger type settings
Set trigger conditions for each trigger.
Analog trigger (p. 106)
Logic trigger (p. 114)
Interval trigger (p. 116)
External trigger (p. 118)
Forcible trigger (p. 118)
Choosing a measurement mode
Choose whether the instrument repeatedly
waits for a trigger after a measurement has
been stopped.
(p. 6)
Starting a measurement
Click the start icon to start a measurement. (The key
green LED lights up.)
When a trigger condition is satised, the instrument
starts to acquire data.
To stop the measurement, click the stop icon.
Click the icon once: Stops the measurement once the
instrument has acquired the recording-length data.
Click the icon twice: Immediately stops the recording.
99
Enabling the Trigger Function
5.2 Enabling the Trigger Function
> [Trigger] > [Common]
Click the [Trigger] button to set it to [On].
On
Enables the trigger function.
Off Disables the trigger function.
To copy settings to other channels
You can copy settings on the analog trigger setting screen.
Refer to “3.5 Copying Settings (Copy Function)” (p. 45).